Chapter B - GRAPHIC USER INTERFACE
Planmeca Proline XC X-ray unit B-11
SERVICE SETTINGS
Technical Manual
2.3 Exposure counter
In the exposure counter mode it is possible to verify the number of exposures taken. There
are three counters in the system, one counting the normal panoramic exposures, one count-
ing the cephalostatic exposures (only in Planmeca Proline XC Pan/Ceph X-ray unit) and one
counting each exposure (also the ones made in the service settings mode). Exposures that
caused an error and were interrupted are not counted.
NOTE
The exposure counter can be cleared on the Factory preset window by touching the
Exposure counter button.
2.4 Signals display
The X-ray unit’s sensor signals are shown on the Signals display.
When the sensor is active the text ON appears beside the sensor name:
•
The cassette holder carriage start (panoramic) point limit sensor operation. (Only in Plan-
meca Proline XC Pan/Ceph X-ray unit)
•
The cassette holder carriage end (cephalometric) point limit sensor operation. (Only in
Planmeca Proline XC Pan/Ceph X-ray unit)
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......