Chapter B - GRAPHIC USER INTERFACE
Planmeca Proline XC X-ray unit B-7
USER PREFERENCE SETTINGS
Technical Manual
when looking directly at the patient.
Note that the patient’s right appear on the left of
the film.
The date will be format day/month/year. The date will appear on film with periods
between the day, month, and year (31.12.92).
•
In the ADA orientation the image appearing on the film is a view of the patient’s teeth seen
as if you were standing behind the patient.
Note that the patient’s right will appear on
right of the film.
The date will be format month/day/year. The date will appear on the film
with dashes between the month, day and year.
The Intensity of the Autoprint text can be adjusted with the
Intensity
arrow fields. There are
eight contrast levels, 1 will give you the lightest text and 8 the darkest.
1.7 Beam check (digital X-ray units)
This feature in the digital Planmeca Proline XC X-ray units allows the user to check that the X-
ray unit is properly collimated.
The panoramic beam check is described in section 2.6 “Beam check” on page E-23 and the
cephalometric beam check in section 7.10 “Beam check” on page E-87.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......