Chapter G - TROUBLESHOOTING
Planmeca Proline XC X-ray unit G-11
TROUBLES IN ADJUSTMENT OF FOCAL TROUGH MOVEMENT
Technical Manual
7 TROUBLES IN ADJUSTMENT OF FOCAL TROUGH
MOVEMENT
7.1 Forwards/backwards movement does not function
7.2 Movement area abnormal
7.3 Erroneous position indication
The position indication is correct if the display shows 00 when the light meets the bite piece.
Erroneous position indication may depend on the fact that the display has not been adjusted
to zero after installing the reference Hall sensor or keyboard processor PCB or that the accu-
mulator is empty.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Motor is not running
Keyboard out of order.
PCB out of order or connector
loose.
Motor out of order or connector
loose.
Motor
stalls
Drawing screw or nut dam-
aged.
Obstruction in the mechanism.
Motor is running but mecha-
nism is immobile.
Cogged belt broken or loose.
Cogged belt wheel glides from
the axle attachment.
Thread of drawing nut dam-
aged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Movement area abnormal
Position information has disap-
peared.
Reference hall sensor defec-
tive.
Reference magnet incorrectly
installed.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......