Chapter G - TROUBLESHOOTING
G-10 Planmeca Proline XC X-ray unit
DISTURBANCES IN CASSETTE MOVEMENT
Technical Manual
6.3 Trembles during movement
6.4 Abnormal noise from the mechanism
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Trembles during movement
Dirt in linear bearings or driving
surface.
Spiral gear damaged.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Abnormal noise from the mech-
anism
Cassette carriage touches sec-
ondary collimator or other con-
structions.
Inner secondary collimator
touches the outer secondary
collimator in connection with
the step motor resonance.
Cassette glide/adhesive strips
resonate.