Chapter G - TROUBLESHOOTING
Planmeca Proline XC X-ray unit G-9
DISTURBANCES IN CASSETTE MOVEMENT
Technical Manual
5.5 Abnormal noise during rotation
6 DISTURBANCES IN CASSETTE MOVEMENT
6.1 Cassette carriage does not move
6.2 Cassette carriage stops suddenly
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Trembling from step motor gen-
erates resonance noise which
is characteristic to the mecha-
nism. Cables from the upper
part of the Z-carriage may also
generate noise. The noise
besides these is abnormal.
Attachment of the driving sur-
face to the groove.
Function of the cable guide.
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
Cassette motor is not running
Keyboard out of order
PCB defective
Motor defective or connector
loose
SYMPTOM
POSSIBLE REASON
CORRECTIVE ACTION
The motor stops running. If
error code appears, see error
code explanations.
Motor or PCB defective or
motor cable loose.
Motor stops/stalls
Gear assembly damaged.
One of the connectors loose.
Motor/drawing wheel glides
Linear bearing gets stuck.
Cassette carriage gets stuck to
secondary collimator or another
part of the mechanism.
Grease on driving wheel or
driving surface.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......