Chapter F - PREVENTIVE MAINTENANCE
F-6 Planmeca Proline XC X-ray unit
MECHANICAL CHECKS
Technical Manual
Exposure time measurement
A calibrated oscilloscope is needed for invasive exposure time measurement. Connect oscil-
loscope channel 1 to the Generator CPU board’s connector P18:10 and oscilloscope ground
to the Generator CPU board’s connector P4:2 (GND). The exposure time can be defined from
the oscilloscope screen as the time interval during which the tube potential exceeds 70% of
the peak tube potential. The exposure time must be within ±10% of the value displayed in the
user interface.
Feedback signal offset measurement
The KVM-feedback signals have a small offset voltage that is used for internal self-testing of
the equipment. In some cases, it can be useful to measure these offsets for troubleshooting
purposes etc. The offset of all feedback signals should be 277 ± 30 mV with respect to the
unit ground potential. The offsets should be measured in idle state (before exposure).
3 MECHANICAL CHECKS
Refer to the chapter “ADJUSTMENT & CALIBRATION” on page D-1.
Содержание Proline XC
Страница 1: ...technical manual 10010942_10 EN...
Страница 2: ......
Страница 8: ...TABLE OF CONTENTS 6 Planmeca Proline XC with Dimax3 Technical manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline XC X ray unit EMC INFORMATION Technical Manual...
Страница 42: ...Chapter B GRAPHIC USER INTERFACE B 18 Planmeca Proline XC X ray unit SERVICE SETTINGS Technical Manual...
Страница 236: ...Chapter H PARTS REPLACEMENT REPAIR H 14 Planmeca Proline XC X ray unit REPLACING THE SOFTWARE Technical Manual...
Страница 253: ...Planmeca Proline XC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 254: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline XC X ray unit Technical Manual...
Страница 256: ......
Страница 257: ......