Chapter B - KEYBOARD FUNCTIONS & MODES
Planmeca Proline EC X-ray unit B-3
NORMAL MODE KEYBOARD FUNCTIONS
Technical Manual
2.3 Automatic return
Press the CTL-key and then press and hold for 3 seconds the return- key to change the state
of the automatic return function. The indicator light of the return key indicates whether the
function is on or off. With the automatic return on the equipment will automatically return to
cassette loading position after panoramic exposures.
2.4 Temple rest motor inactivation
The temple rest motor can be inactivated. Press the CTL-key and press and hold down for 4
seconds the temple support key. The text HLd 0 will appear on the display. Remove the sup-
ports, the unit can be used without them.
WARNING SOUND
TONE ADJUST-
MENT
“EXPOSURE
WARNING SIG-
NAL ADJUST-
MENT” on page D-
26
PARAMETER HIS-
TORY DISPLAY
“Parameter his-
tory display” on
page B-4
AUTOPRINT SET-
TINGS
“Autoprint settings”
on page B-5
USER SETTING
KEY SEQUENCE
FOR DETAILS,
REFER TO
P
READY
PRET
CTL
READY
PRET
CTL
CTL
hold for
READY
PRET
3 seconds
AUTO
AUTO
CTL
hold for 3 seconds
CTL
hold for
4 seconds
CTL
Содержание Proline EC ceph
Страница 2: ......
Страница 8: ...TOC 6 Planmeca Proline EC Panoramic X ray TABLE OF CONTENTS Technical Manual...
Страница 24: ...Chapter A GENERAL TECHNICAL DATA A 16 Planmeca Proline EC X ray unit EMC INFORMATION Technical Manual...
Страница 164: ...Chapter E DIMAX3 DIGITAL SYSTEM ADJUSTMENT E 70 Planmeca Proline EC X ray unit IMAGE AREA SHIFTING Technical Manual...
Страница 182: ...Chapter G TROUBLESHOOTING G 12 Planmeca Proline EC X ray unit PROBLEMS IN RADIOGRAPHS Technical Manual...
Страница 188: ...Chapter H PARTS REPLACEMENT REPAIR H 6 Planmeca Proline EC X ray unit CHANGING THE TUBEHEAD Technical Manual...
Страница 205: ...Planmeca Proline EC X ray unit J 1 Chapter Technical Manual J WIRING DIAGRAM...
Страница 206: ...Chapter J WIRING DIAGRAM J 2 Planmeca Proline EC X ray unit Technical Manual...
Страница 208: ......
Страница 209: ......