Chapter E - DIMAX3 DIGITAL SYSTEM ADJUSTMENT
E-64 Planmeca Proline EC X-ray unit
CEPHALOSTAT ADJUSTMENT
Technical Manual
7.9 Second primary slot adjustment
Second primary slot in too low or high position
Loosen all the three screws located on the side of second primary slot with the 4mm allen
key.
Figure 115
Move the second primary slot in vertical direction until the slot is correctly positioned. Tighten
the loosened screws.
Second primary slot not vertical
Loosen the two screws located on the side of second primary slot.
Figure 116
Loosen these three screws.
Loosen these two screws.