Planmeca PlanMill 40 S User Manual
PlanMill 40 S Safety
6
PlanMill 40 S Safety
Warning:
Failure to adhere to all safety warnings may result in personal injury, equipment damage, or data loss.
The mill weighs approximately 170 lb. (77 kg). When lifting or moving the mill, follow the guidelines in
page 4.
Always ensure the mill is on a surface capable of supporting its weight of approximately 170 lb. (77 kg).
Use caution when handling the mill’s cutting tools.
Before milling a restoration, ensure the mill has adequate milling fluid. Adequate milling fluid is required to prevent overheating
and possible damage to the restoration blank and cutting tools.
Do not open the mill’s lid while a milling operation is in progress. Doing so will result in the loss of data and restoration block; the
milling process will have to be repeated.
This equipment is intended to be grounded. Connect the products to earth grounded outlet only. Power strips or extension cords
should not be used.
If a power cord (other than the one provided with the equipment) is used for connecting to the power source, ensure it meets all
of the following requirements:
• Detachable power supply cord
• Type SJT
• 18 AWG 3 conductor
• Rated 10 A or better
• For products outside of the United States and Canada, the power cord must be marked “HAR” or with a suitable agency
marking from the country of intended use. The attachment plug and appliance coupler must be marked with a suitable
agency marking from the country of intended use.
Do not use the products for any purpose other than its intended and labeled use.
To prevent electrical shock, do not open any sealed or user restricted access panels or connectors.
A detachable non-locking type power cord has been provided with this equipment as the disconnect device. Do not block access
to the power cord. In case of emergency, remove power from the device by unplugging the cord at either end.
Do not block any of the product’s cooling vents located on each side and lower rear panel. Doing so may result in overheating and
damage to the products and will void the product warranty.
When placing components, adhere to all clearances stated in
“Minimum Clearance” on page 27.
Do not make any unauthorized repairs or modifications to the system software or hardware. This includes installing unauthorized
software on the computer system or altering or bypassing any safety switches or mechanisms.
Do not install or operate the products in an environment where an explosion hazard exists, e.g., high oxygen area.
Comply with all applicable regulations when disposing of waste materials.
Do not attach any equipment or devices to the mill unless their use has been specifically authorized by E4D Technologies or
Planmeca.
The wireless components in the products may be interfered with by other equipment, even if the other equipment is fully
compliant with CISPR (International Special Committee on Radio Interference) emission requirements.
When possible, electrical equipment should not be used when adjacent to other electrical equipment. If adjacent use is necessary,
the equipment should be observed to verify normal operation in the configuration in which it will be used.
When connecting the components, use only the cables supplied with the products. Failure to do so may result in increased
electromagnetic emissions or reduced immunity to external electromagnetic emissions.
Ensure your mill is properly maintained through periodic maintenance.
If you suspect equipment malfunction or failure, discontinue using the mill and contact Planmeca Customer Support immediately.
Do not attempt to make any repairs on the mill.
Read and comply with all safety, warning, and instructional labels on the mill.