10 TEMPOROMANDIBULAR JOINT (TMJ) EXPOSURE
52 Planmeca ProMax 2D & 3D s & 3D Classic
User’s Manual (2D)
10.5 Taking an exposure in three angle TMJ programs
NOTE
Make sure that you have selected the correct patient and
exposure mode in the Planmeca Romexis program.
1. Select the forward button.
Green lights flash on the touch screen and exposure
button when the X-ray system is getting ready for an
exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
2. Ask the patient to stay as still as possible. Explain to
the patient that this is a multiple exposure and that
the C-arm will move through three exposure cycles.
3. Move to a protected area.
4. Press and hold down the exposure button for the
duration of the exposure.
During exposure yellow radiation warning lights
illuminate on the exposure switch and on the touch
screen, and you hear a radiation warning tone.
Additionally, a radiation warning symbol is shown on
the touch screen.
NOTE
Maintain audio and visual contact with the patient and X-
ray unit during exposure. If the C-arm stops moving during
exposure, or moves in an erratic way, release the
exposure button immediately.
5. The image is shown on the computer screen.
•
Note that you must accept the image in the
Planmeca Romexis program. Refer to the
Romexis User’s Manual.
6. Guide the patient away from the X-ray unit.
Forward
button
ll6_2.eps
Green light
s
Fla
s
hing green =
Getting ready
Continuou
s
green =
Ready
_2eps
Yellow light
s
1/3