6 MAIN PARTS
User’s Manual (2D)
Planmeca ProMax 2D & 3D s & 3D Classic 13
6.6 Exposure switch
The exposure switch can be mounted on the wall, or it can
be hung from the hook provided on the stationary column
top if a protected area is within reach.
Green lights flash on the exposure button and on the
touch screen when the X-ray system is getting ready for
an exposure. The green lights stop flashing and stay on
continuously when the X-ray system is ready for an
exposure.
During exposure yellow radiation warning lights illuminate
on the exposure switch and on the touch screen. They
indicate that the X-ray unit is generating radiation.
6.7 Emergency stop button
The emergency stop button is located on the top of the
stationary column. Press the button to stop the X-ray unit
operating in an emergency. When the emergency stop
button is pressed down, all movements of the X-ray unit
are blocked and the unit will not generate radiation. The
up / down movement will stop within a distance of 10 mm
(0.4 in.).
A help message will appear on the touch screen. Guide
the patient away from the X-ray unit. Then release the
emergency stop button. The X-ray unit will automatically
restart.
6_2.eps
1) Fla
s
hing green =
Getting ready
2) Continuou
s
green =
Ready
3) Yellow = Radiation
Exposure switch
Exposure button