29.5.2 Instrument flushing holder
The instrument flushing holder can be lifted away from its position.
Disinfect the flushing holder once a month in a washer-disinfector at 93°C or
autoclave it at 134°C.
Note that the flushing holder cover can not be autoclaved. Remove the cover
before autoclaving the instrument flushing holder.
29.5.3 Suction tube cleaning holder and separation tank
Clean the suction tube cleaning holder and the separation tank behind it
once a month.