Attach the quick connector instrument hoses (1) to the instrument console by
pushing them firmly into position and turning the locking ring clockwise.
Attach the syringe instrument into position next to the control panel i.e. to the
leftmost position.
CAUTION
Remember to verify that all the instrument hoses are properly connected and
the locking rings properly tightened to avoid any leakage.
Instrument console with balanced instrument arms: Route the instrument
hoses through the hose guide by carefully bending the hose guide hook and
passing the hose over the roller. Connect the instruments to the
corresponding hoses and place them into their positions over the instrument
console.
Instrument console with hanging-tube instruments: Connect the instruments
to the corresponding hoses and place them into their instrument holders.
Adjust the holders to the desired angles.
16.4 Installing apex locator assembly
NOTE
The apex locator is not available for Planmeca Compact i Classic.
16.4.1 Instrument console with balanced instrument arms
Steps
1. Check that the two DIN912 M3x4 screws on the apex locator housing
assembly are loosened before installing.
The screws are tightened in step 3.
16 Installing instruments
84 Planmeca Compact i3
Installation manual