Chapter C - SERVICE MODE
Planmeca Compact i C-39
SERVICE MODE DETAILED DESCRIPTION
Technical Manual
n.102 Disable dry spray mode when using Plus Package (0, 1, 2)
This service mode is for disabling dry spray mode from the instruments when using Plus
Package. 0 = dry spray mode enabled, 1 = dry spray mode disabled, Plus package signal
tones, 2 = dry spray mode disabled, normal signal tones.
The factory default is 0 (dry spray mode enabled).
n 103 Show the dental unit type (nn)
First digit shows type of Main PCB:
2 = Planmeca Compact i
Second digit shows IMUX version:
0 = no IMUX
1 = IMUX PCB version E or older
2 = IMUX PCB version F-G (1.08.x)
3 = IMUX PCB version F-G (1.09.x or 1.10.x)
4 = IMUX PCB version H or newer
Third digit shows dental unit type:
0 = Prostyle Compact
1 = Planmeca Compact e
2 = Planmeca Compact i
Fourth digit shows patient chair type:
0 = no patient chair
1 = Prostyle Compact chair without legrest or legrest up
2 = Prostyle chair with legrest down
3 = WE’S lifter
4 = Planmeca Compact e chair
5 = Planmeca Compact i chair with fixed legrest; or automatic legrest up
6 = Planmeca Compact i chair with 510 automatic legrest; in automatic state
7 = Planmeca Compact i chair with 460 automatic legrest; in automatic state
8 = Planmeca Compact e with swivelled position recognition and fixed legrest
9 = Planmeca Compact e with swivelled position recognition and automatic legrest
A = Planmeca Compact e swivelled to 90° with automatic legrest
NOTE
The Planmeca Compact i chair with automatic legrest can be either number 5
or 6/7; it is number 5 if the legrest is locked up, and it is number 6/7 when the
legrest is moving synchronously with the backrest.
n.104 Instrument multiplexer additional water valve control (0, 1,
2, A)
This service mode is for enabling the instrument multiplexer additional water valve control.
0 = disabled
1 = reserved for factory use
2 = enabled (default)
0 - 2 are for IMUX PCB version G
A = automatic
A is for IMUX PCB version H
n.105 Enable/disable instrument places ( IIIII signal )
In this service mode you can enable/disable instrument places. The different setting combina-
tion can be scrolled and selected by using the “A”- and “B”-keys. The display shows a short
line on the place where the instrument place is disabled.
Note, that the syringe cannot be
disabled.
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......