Chapter F - ADJUSTMENTS
Planmeca Compact i F-3
MECHANICAL ADJUSTMENTS
Technical Manual
1.3 Adjusting balance of the OP delivery arm
Remove the cable guide plate.
NOTE
When replacing the cable guide plate to its position, note that the distance
between the end of the guide and the end of the adjustment plate must be 0-
20mm (0-0.8 in.).
Adjusting the balance
The adjustment plate, which is connected to the gas spring, is held in position with the spring
support. To increase or decrease the tension of the gas spring, the position of the spring sup-
port compared to the adjustment plate must be altered. This is done with the aid of the adjust-
ing hook, which is pushed by a screwdriver (use a robust, round-shaped (not six-edged)
screwdriver with a diameter of 4.5…5.5 mm).
0...20 mm
Wireguide_4.2.eps
Adjustment plate
Cable
guide
plate
Adjustment plate
Gas spring
Spring support
Adjusting hook
y-spring
Содержание Compact I
Страница 1: ...technical manual 10007462_23 EN...
Страница 2: ......
Страница 20: ...Chapter A GENERAL TECHNICAL DATA A 12 Planmeca Compact i EMC INFORMATION Technical Manual...
Страница 78: ...Chapter C SERVICE MODE C 50 Planmeca Compact i SERVICE MODE DETAILED DESCRIPTION Technical Manual...
Страница 110: ...Chapter E PREVENTIVE MAINTENANCE E 10 Planmeca Compact i WATER MANAGEMENT SYSTEM WMS Technical Manual...
Страница 212: ...Chapter H FOOT CONTROL H 18 Planmeca Compact i ADJUSTMENTS Technical Manual...
Страница 213: ...Planmeca Compact i I 1 Chapter Technical Manual I DIAGRAMS...
Страница 214: ...Chapter I DIAGRAMS I 2 Planmeca Compact i Technical Manual...
Страница 216: ......
Страница 218: ......
Страница 220: ......
Страница 222: ......
Страница 224: ......
Страница 226: ......
Страница 228: ......
Страница 230: ......
Страница 232: ......
Страница 234: ......
Страница 236: ......
Страница 238: ......
Страница 240: ......
Страница 242: ......
Страница 243: ......