The pop-up window closes and the new values are displayed on the
control panel.
25.5 Timer settings
About this task
Up to six timer settings can be programmed into the timer memory.
Steps
1. Press Program.
2. Press Timer.
A pop-up window opens.
3. Press the timer that you want to adjust.
A new pop-up window opens.
4. Adjust the length of the timer.
The minimum value is 5 sec. and the maximum 20 min. The adjustment
step is 5 sec.
5. Press OK.
The pop-up window closes.
25.6 Bowl rinse and cup fill
25.6.1 Adjusting cup fill and bowl rinse flow rates
The flow rates of the bowl flush and cup fill can be adjusted with the two
black knobs located inside the dental unit.
25 Programming
User's manual