User manual - ATR224 - DRR224 - 19
11
Table of configuration parameters
GROUP A -
A.in.1
- Analogue input 1
1
SEN.1
Sensor AI1
Analogue input configuration / sensor AI1 selection
tc. k
Tc-K
-260° C..1360° C. (
Default
)
tc. s
Tc-S
-40° C..1760° C
tc. r
Tc-R
-40° C..1760° C
tc. J
Tc-J
-200° C..1200° C
Tc. T
Tc-T
-260° C..400° C
tc. E
Tc-E
-260° C..980° C
tc. N
Tc-N
-260° C..1280° C
tc. b
Tc-B
100° C..1820° C
Pt100
Pt100
-200° C..600° C
ni100
Ni100
-60° C..180° C
ntc 1
NTC 10K β3435K
-40° C..125° C
Ptc
PTC 1K
-50° C..150° C
Pt500
Pt500
-200° C..600° C
Pt1k
Pt1000
-200° C..600° C
0-1
0..1 V
0-5
0..5 V
0-10
0..10 V
0-20
0..20 mA
4-20
4..20 mA
0-60
0..60 mV
Pot.
Potentiometer (set the value on parameter 6)
ni120
Ni120
-60 °C..240 °C
ntc 2
NTC 10K β3694K
-40 °C..150 °C
ntc 3
NTC 2252 β3976K
-40 °C..150 °C
2
d.P. 1
Decimal Point 1
Select number of displayed decimal points for AI1
0
Default
0.0
1 decimal
0.00
2 decimals
0.000
3 decimals
3
deGr.
Degree
°C
Celsius (
Default
)
°F
Fahrenheit
K
Kelvin
4
L.L.i.1
Lower Linear Input AI1
AI1 lower limit only for linear signals. Ex.: with input 4..20 mA this parameter takes value associated
to 4 mA. The value may be greater than the one entered on the next parameter.
-9999..+30000
[digit
Default
: 0.
5
u.L.i.1
Upper Linear Input AI1
AI1 upper limit only for linear signals Ex: with input 4..20 mA this parameter takes value associated
to 20 mA.The value may be lower than the one entered on the previous parameter.
-9999..+30000
[digit
Default
:1000
6
P.vA.1
Potentiometer Value AI1
Selects the value of the potentiometer connected on AI1
1..150 kohm
.
Default
: 10kohm