![Piusi K24 Скачать руководство пользователя страница 1](http://html.mh-extra.com/html/piusi/k24/k24_use-maintenance-and-installation-manual_1560355001.webp)
ENGLISH
ENGLISH
ENGLISH
ENGLISH
ENGLISH
ENGLISH
K24
MANUALE D’USO, MANUTENZIONE
E CALIBRAZIONE
USE, MAINTENANCE AND
CALIBRATION MANUAL
Bulletin M0225 ITEN rev. 0
Italiano
English
INDEX
A
DECLARATION OF CONFORMITY
B
GENERAL WARNINGS
C
SAFETY INSTRUCTIONS
C.1
SAFETY WARNINGS
C.2
FIRST AID RULES
C.3
GENERAL SAFETY RULES
C.4 PACKAGING
C.5
PACKAGE CONTENTS/PRE-INSPECTION
D
BECOMING ACQUAINTED WITH K24
D.1
COMPATIBLE LIQUIDS
D.2
DISPLAY LCD
D.3
DISPLAY POSITIONING (METER VERSION ONLY)
D2
USERS BUTTONS
E
OPERATING MODES
F
INSTALLATION
G
DAILY USE
G.1
DISPENSING IN NORMAL MODE
G.1.1
PARTIAL RESET (NORMAL MODE)
G.1.2
RESETTING THE RESET TOTAL
G.2
DISPENSING WITH FLOW RATE MODE DISPLAY
G.2.1
PARTIAL RESET (FLOW RATE MODE)
H
CALIBRATION
H1 DEFINITIONS
H2
CALIBRATION MODE
H.2.1
DISPLAY OF CURRENT CALIBRATION FACTOR AND
RESTORING
FACTORY
FACTOR.
H.2.2
IN FIELD CALIBRATION
H.2.2.1
IN-FIELD CALIBRATION PROCEDURE
H.2.3
DIRECT MODIFICATION OF K FACTOR
I
METER CONFIGURATION
L MAINTENANCE
M MALFUNCTIONS
N DISPOSAL
O
DATI TECNICI / TECHNICAL DATA
P
EXPLODED VIEWS AND OVERALL DIMENSIONS
A
DECLARATION OF CONFORMITY
The undersigned:
PIUSI S.p.AVia Pacinotti c.m. z.i.Rangavino
46029 Suzzara - Mantova - Italy
HEREBY STATES
under its own responsibility, that the equipment described below:
Descriprion: METER
Model: K24
Serial number: refer to Lot Number shown on CE plate affixed to productYear of
manufacture: refer to the year of production shown on the CE plate affixed to the
productis in conformity with the legal provisions indicated in the directives :
- Electromagnetic Compatibility Directive 2004/108/EC
The documentation is at the disposal of the competent authority following motivated
request at Piusi S.p.A. or following request sent to the email address: doc_tec@
piusi.comThe person authorised to compile the technical file and draw up the decla
-
ration is Otto Varini as legal representative.
Suzzara, 01/01/2010
Otto Varini
legal representative.
B
GENERAL WARNINGS
Important
precautions
To ensure operator safety and to protect the pump from po-
tential damage, workers must be fully acquainted with this
instruction manual before performing any operation.
Symbols used
in the manual
The following symbols will be used throughout the manual
to highlight safety information and precautions of particular
importance:
ATTENTION
This symbol indicates safe working practices for opera-
tors and/or potentially exposed persons.
WARNING
This symbol indicates that there is risk of damage to the
equipment and/or its components.
NOTE
This symbol indicates useful information.
Manual pres-
ervation
his manual should be complete and legible throughout. It
should remain available to end users and specialist installation
and maintenance technicians for consultation at any time.
Reproduction
rights
All reproduction rights are reserved by Piusi S.p.A. The text can-
not be reprinted without the written permission of Piusi S.p.A.
THIS MANUAL IS THE PROPERTY OF Piusi S.p.A.
ANY REPRODUCTION, EVEN PARTIAL, IS FORBIDDEN.
C
SAFETY INSTRUCTIONS
C.1
SAFETY WARNINGS
Mains - prelim-
inary checks
before instal-
lation
ATTENTION
You must avoid any contact between the electrical pow-
er supply and the fluid that needs to be FILTERED.
Maintenance
control
Before any checks or maintenance work are carried out, dis-
connect the power source.
For your safety,
review the ma-
jor warnings
and cautions
below before
operating your
meter
When metering flammable liquids, observe precautions
against fire or explosion
When handling hazardous liquids, always follow the liquid
manifacturer’s safety precautions
Always dispose of used cleaning solvents in a safe manner
according to the solvent manifacturer’s instructions.
During meter removal, liquid may spill. Follow the liquid
manifacturer’s safety precautions to clean up minor spills
Do not blow compressed air through the meter
Do not allow liquids to dry inside the meter
C.2
FIRST AID RULES
Contact with
the product
In the event of problems developing following EYE/SKIN
CONTACT, INHALATION or INGESTION of the treated product,
please refer to the SAFETY DATA SHEET of the fluid handled
.
NOTE
Please refer to the safety data sheet for the product
SMOKING
PROHIBITED
When operating the dispensing system and in particular dur-
ing refuelling, do not smoke and do not use open flame.
ATTENTION
When metering flammable liquids, observe precau-
tions against fire or explosion
When handling hazardous liquids, always follow the
liquid manifacturer’s safety precautions.
Do not submerge the meter
C.3
GENERAL SAFETY RULES
Essential pro-
tective equip-
ment charac-
teristics
Wear protective equipment that is:
suited to the operations that need to be performed;
resistant to cleaning products.
Personal pro-
tective equip-
ment that
must be worn
Wear the following personal protective equipment during
handling and installation:
safety shoes
;
close-fitting clothing;
protective gloves;
safety goggles;
Protective
equipment
instruction manual
WARNING
If handling hazardous liquids, always follow the Liquid
Manifacturer’s Safety Precautions. Wear protective
clothing such as goggles, gloves and respirator as
instructed.
When metering flammable liquids, observe precau-
tions against fire or explosion. Do not meter in the pres-
ence of any source of ignition including running or hot
engines, lighted cigarettes, or gas or electric heaters
C.4 PACKAGING
FOREWORD
K24 comes packed in a cardboard box with a label indicating the fol-
lowing data:
1 - contents of the pack-
age
2 - weight of the contents
3 - description of the
product
C.5
PACKAGE CONTENTS/PRE-INSPECTION
FOREWORD
To open the packaging, use a pair of scissors or a cutter, being careful
not to damage the dispensing system or its components.
NOTE
In the event that one or more of the components de-
scribed below are missing from inside the package,
please contact Piusi inc technical support.
WARNING
Check that the data on the plate correspond to the
desired specifications. In the event of any anomaly,
contact the supplier immediately, indicating the nature
of the defects. Do not use equipment which you suspect
might not be safe.
D
BECOMING ACQUAINTED WITH K24
FOREWORD
Electronic digital meter featuring a turbine measurement system, de-
signed for precise measuring of low viscosity fluids.
K24 is a bi-directional meter with LCD display and calibration
buttons. The body is made of aluminum (conductive) and de-
signed for high flow 120 l/min. (32 GPM).
K24 is available in 3 versions:
1
a
METER – with LCD display and calibration buttons (SILVER LABEL)
b
METER – with LCD display and calibration buttons (RED LABEL)
2
PULSER – single-channel impulse, cannectable with a remote display.
D.1
COMPATIBLE LIQUIDS
Turbine
measurement
system
The turbine is placed inside a hole through the body of k24, fitted with
M-f threaded inlet and outlet.
The liquids compatible with k24 are at low viscosity, namely:
Diesel fuel
Kerosene
Gasoline
Gasoline blended alcol max 15%
ATTENTION
DO NOT USE WITH SUNDRIES LIQUIDS
Main components K24 Meter
Main components K24 Pulser
1
LCD display
1
Plate with technical data
2
RESET key
3
CAL key
4
Technical data plate
3
2 1
4
1
D.2
DISPLAY LCD
FOREWORD
The “LCD” of the METER features two numerical registers and various
indications displayed to the user only when the applicable function
so requires.
1
Partial register (5 figures with moving
comma FROM 0.1 to 99999) indicating
the volume dispensed since the reset
button was last pressed
6
Indication of type of total, (TOTAL /
Reset TOTAL);
2
Indication of battery charge
7
Indication of unit of measurement of
Totals:
L=Litres Gal=Gallons
3
Indication of calibration mode
8
Indication of Flow Rate mode
4
Totals register (6 figures with moving
comma FROM 0.1 to 999999), that can
indicate two types of Total:
4.1. General Total that cannot be reset (TOTAL)
4.2. Resettable total (Reset TOTAL)
9
Indication of unit of measurement of
Partial: Qts=Quarts
Pts=Pints
L=Litres
Gal=Gallons
5
Indication of total multiplication factor
(x10 / x100)
4
5
6
9
7
1
2
3
8
D.3
DISPLAY POSITIONING (METER VERSION ONLY)
FOREWORD
The square shape of the k24 body allows the card to be rotated in its
housing, thus ensuring great versatility in positioning
This allows easy display readings in any position. The card housing
is closed by a plastic cover sealed through a rubber protection act-
ing as a gasket as well. This can be easily removed unscrewing the
4 screws that fix both the cover and the card (1).
ATTENTION
While fixing the K24 card, make sure the battery con-
tact cable is not placed above the circular housing of
the bulb.
D4
USERS BUTTONS
FOREWORD
The METER features two buttons (RESET and CAL) which individ-
ually perform two main functions and, together, other secondary
functions.
MAIN
FUNCTIONS
PERFORMED
- for the RESET key, resetting the partial register and Reset Total
- for the CAL key, entering instrument calibration mode
SECONDARY
FUNCTIONS
Used together, the two keys permit entering configuration mode
where the desired unit of measurement can be set.
LEGEND
CALIBRATE MEANS PERFORMING ACTIONS ON THE METER KEYS.
BELOW IS THE LEGEND OF THE SYMBOLS USED TO DESCRIBE
THE ACTIONS TO BE PERFORMED
SHORT
PRES-
SURE
OF CAL
KEY
CAL
LONG
PRES-
SURE
OF CAL
KEY
CAL
CAL
CAL
SHORT
PRES-
SURE
OF
RESET
KEY
RESET
LONG
PRES-
SURE
OF
RESET
KEY
RESET
RESET
RESET
E
OPERATING MODES
OPERATING
MODES
The user can choose between two different operating modes:
The meter features a non-volatile memory for storing the dispensing
data, even in the event of a complete power break for long periods.
The measurement electronics and the LCD display are fitted in the top
part of the K24 which remains isolated from the fluid-bath measure-
ment chamber and sealed from the outside by means of a cover.
1 - Normal
Mode
Normal Mode: Mode with display of Partial and Total dis-
pensed quantities
2 - Flow rate
Mode
Flow Rate Mode: Mode with display of Flow Rate, as well as
Partial dispensed quantity.
F
INSTALLATION
FOREWORD
K24 features a threaded, perpendicular inlet and outlet (1” NPT
or
BSP
male and female that can be combined together). It has been
designed to be easily installed in any position: fixed in-line or mobile
on a dispensing nozzle. In order to improve the life of the turbine, it is
recommended to fit a strainer before the meter itself
ATTENTION
For installations on system, position meter so that the
battery housing can be easily reached.
Connections
To protect against the leakage, make sure all threads are
sealed with two or three turns of thread tape or a sealing
compound compatible with the liquid being metered
ATTENTION
Make sure the thread tape or sealing compound does
interfere with flow
Make sure there are no leaks in the connections.
To seal leaks, remove and inspect the meter and re-
place the thread tape or sealant. Refer to the Trouble-
shooting Section
G
DAILY USE
FOREWORD
The only operations that need to be done for daily use are partial and/
or resettable total register resetting. The user should use only the dis-
pensing system of k24. Occasionally the meter may need to be con-
figured or calibrated. To do so, please refer to the relevant chapters.
Below are the two typical normal operation displays. One display page shows the par-
tial and reset total registers. The other shows the partial and general total. Switchover
from resettable total to general total display is automatic and tied to phases and times
that are in factory set and cannot be changed.
NOTA
6 digits are available for Totals, plus two icons x 10 /
x100. The increment sequence is the following:
0.0 -> 99999.9 -> 999999 -> 100000 x 10 -> 999999 x
10 ->100000 x 100 -> 999999 x 100
G.1
DISPENSING IN NORMAL MODE
FOREWORD
Normal mode is the standard dispensing. While the count is made, the
partial and resettable total are displayed at the same time (reset total).
WARNING
Should one of the keys be accidentally pressed during
dispensing, this will have no effect.
STAND BY
A few seconds after dispensing has ended, on the lower register, the
display switches from resettable total to general total: the word reset
above the word total disappears, and the reset total is replaced by
the general total.
This situation is called standby and remains stable until the user
operates the k24 again.
G.1.1 PARTIAL RESET (NORMAL MODE)
The partial register can be reset by pressing the reset key when
the meter is in standby, meaning when the display screen
shows the word “TOTAL”.
After pressing the reset key, during reset, the display screen
first of all shows all the lit-up digits and then all the digits that
are not lit up.
At the end of the process, a display page is first of all shown
with the reset partial and the reset total
and, after a few moments, the reset total is replaced by the
non resettableTotal.
G.1.2 RESETTING THE RESET TOTAL
VALID JUST FOR METER VERSION WITH SILVER LABEL
The reset total resetting operation can only be performed
after resetting the partial register. The reset total can in
fact be reset by pressing the reset key at length while
the display screen shows reset total as on the following
display page:
Schematically, the steps to be taken are:
1 Wait for the display to show normal standby display page
(with total only displayed)
2 Press the reset key quickly
3 The meter starts to reset the partial
4 While the display page showing the reset total is displayed
Press the reset key again for at least 1 second
5
The display screen again shows all the segments of the
display followed by all the switched-off segments and
finally shows the display page where the reset Reset
Total is shown.
C.2
DISPENSING WITH FLOW RATE MODE DISPLAY
VALID JUST FOR METER VERSION WITH SILVER LABEL
It is possible to dispense fluids, displaying at the same
time::
1 the dispensed partial
2 the Flow Rate in [Partial Unit / minute] as shown on the
following display page:
Procedure for entering this mode:
1 wait for the Remote Display to go to Standby, meaning
the display screen shows Total only
2 quickly press the CAL key.
3 Start dispensing
The flow rate is updated every 0.7 seconds. Consequently, the display could be relatively
unstable at lower flow rates. The higher the flow rate, the more stable the displayed
value.
IMPORTANT
The flow rate is measured with reference to the unit of
measurement of the Partial. For this reason, in case of
the unit of measurement of the Partial and Total being
different, as in the example shown below, it should be
remembered that the indicated flow rate relates to
the unit of measurement of the partial. In the example
shown, the flow rate is expressed in Qts/min.
The word “Gal” remaining alongside the flow rate
refers to the register of the Totals (Reset or NON Reset)
which are again displayed when exiting from the flow
rate reading mode.
To return to “Normal” mode, press the CAL key again. If one of the two keys RESET or
CAL is accidentally pressed during the count, this will have no effect.
IMPORTANT
Even though in this mode they are not displayed, both the
Reset Total and the General Total (Total) increase. Their
value can be checked after dispensing has terminated,
returning to “Normal” mode, by quickly pressing CAL.
G.2.1 PARTIAL RESET (FLOW RATE MODE)
To reset the Partial Register, finish dispensing and wait for
the Remote Display to show a Flow Rate of 0.0 as indicated
in the illustration
then quickly press RESET
H
CALIBRATION
When operating close to extreme use or flow rate conditions (close to minimum or
maximum acceptable values), an on-the-spot calibration may be required to suit
the real conditions in which the K24 is required to operate.
H1 DEFINITIONS
CALIBRATION
FACTOR OR
“K FACTOR”
Multiplication factor applied by the system to the electrical pulses
received, to transform these into measured fluid units.
FACTORY K
FACTOR
Factory-set default factor. It is equal to 1,000. This calibration factor
ensures utmost precision in the following operating conditions:
Fluid
diesel
Temperature: 20°C
Flow rate: 50 lit/min (13 GPM)
Even after any changes have been made by the user, the factory k
factor can be restored by means of a simple procedure.
USER K FACTOR: Customized calibration factor, meaning modified by calibration.
H2
CALIBRATION MODE
Why calibrate? 1
Display the currently used calibration factor:
2
Return to factory calibration (Factory K Factor) after a pre-
vious calibration by the user
3
Change the calibration factor using one of the two previ-
ously indicated procedures
FOREWORD
Two procedures are available for changing the Calibration Factor:
1
In-Field Calibration, performed by means of a dispensing
operation
2
Direct Calibration, performed by directly changing the
calibration factor
In calibration mode, the partial and total dispensed quantities indicated on the display
screen take on different meanings according to the calibration procedure phase.
In calibration mode, the K24 cannot be used for normal dispensing operations.
In “Calibration” mode, the totals are not increased
ATTENTION
The K24 features a non-volatile memory that keeps
the data concerning calibration and total dispensed
quantity stored for an indefinite time, even in the
case of a long power break; after changing the bat-
teries, calibration need not be repeated.
H.2.1 DISPLAY OF CURRENT CALIBRATION FACTOR
AND RESTORING FACTORY FACTOR.
CAL
CAL
CAL
By pressing the CAL key while the appli-
ance is in Standby, the display page ap-
pears showing the current calibration fac-
tor used. If no calibration has ever been
performed, or the factory setting has
been restored after previous calibrations,
the following display page will appear:
The word “Fact” abbreviation for “factory”
shows that the factory calibration factor
is being used
If, on the other hand, calibrations have
been made by the user, the display page
will appear showing the currently used
calibration factor ( in our example 0,998) .
The word “user” indicates a calibration fac-
tor set by the user is being used..
CAL
The flow chart alongside
shows the switchover logic
from one display page to
another
In this condition, the Reset
key permits switching
from User factor to Factory
factor.
To confirm the choice of
calibration factor, quickly
press CAL while “User” or
“Fact” are displayed.
After the restart cycle, the
K24 uses the calibration
factor that has just been
confirmed
CAL
CAL
CAL
CAL
RESET
RESET
TIME OUT
ATTENTION
When the Factory Factor is confirmed, the old User
factor is deleted from the memory
H.2.2 IN FIELD CALIBRATION
FOREWORD
This procedure calls for the fluid to be dispensed into a gradu-
ated sample container in real operating conditions ( flow rate,
viscosity, etc.) requiring maximum precision.
ATTENTION
For correct K24 calibration, it is most important to:
1 When the Factory Factor is confirmed, the old User factor is deleted
from the memory
2 use a precise Sample Container with a capacity of not less than 5
litres, featuring an accurate graduated indicator.
3 ensure calibration dispensing is done at a constant flow rate equiva-
lent to that of normal use, until the container is full;
4 Not reduce the flow rate to reach the graduated area of the con-
tainer during the final dispensing stage (the correct method during
the final stages of sample container filling consists in making short
top-ups at normal operation flow rate) ;
5 after dispensing, wait a few minutes to make sure any air bubbles
are eliminated from the sample container; only read the Real value
at the end of this stage, during which the level in the container could
drop.
6 Carefully follow the procedure indicated below.
H.2.2.1
IN-FIELD CALIBRATION PROCEDURE
ACTION
DISPLAY
1
NONE
Meter in Standby
2
CAL
CAL
CAL
LONG CAL key keying
The Meter enters calibration mode, shows <<CAL>> and
displays the calibration factor in use instead of partial. The
words “Fact” and “USER” indicate which of the two factors
(factory or user) is currently in use.
Important: This factor is that which the instrument also uses for
field calibration measurement operations
3
RESET
RESET
RESET
LONG RESET key keying
The Meter shows “CAL” and the partial at zero. The Meter is
ready to perform in-field calibration.
4
DISPENSING INTO SAMPLE CONTAINER
Without pressing any key, start dispensing into the sample
container
Dispensing can be interrupted and started again at will.
Continue dispensing until the level of the fluid in the sample
container has reached the graduated area. There is no need to
reach a preset quantity.
Indicated value Real value
5
RESET
SHORT RESET key keying
The Meter is informed that the calibration dispensing
operation is finished.
Make sure dispensing is correctly finished before performing
this operation. To calibrate the Meter, the value indicated by
the partial totaliser (example 9.800) must be forced to the
real value marked on the graduated sample container. In the
bottom left part of the display an arrow appears (upwards and
downwards), that shows the direction (increase or decrease) of
the value change displayed when the following operations 6
or 7 are performed.
6
RESET
SHORT RESET key keying
The arrow changes direction. The operation can be repeated to
alternate the direction of the arrow.
7
CAL
CAL
CAL
CAL
SHORT/LONG CAL key keying
The indicated value changes in the direction indicated by the
arrow
- one unit for every short CAL key keying
- continually if the CAL key is kept pressed. The speed increase
rises by keeping the key pressed. If the desired value is
exceeded, repeat the operations from point (6).
8
RESET
RESET
RESET
LONG RESET key keying
The Meter is informed that the calibration procedure is finished.
Before performing this operation, make sure the INDICATED
value is the same as the REAL value.
.
Indicated value Real value
The Meter calculates the new USER K FACTOR ; this calculation
could require a few seconds, depending on the correction to
be made
ATTENTION: If this operation is performed after action (5),
without changing the indicated value, the USER K FACTOR would
be the same as the FACTORY K FACTOR, thus it is ignored.
9
NO OPERATION
At the end of the calculation, the new USER K FACTOR is shown
for a few seconds, after which the restart cycle is repeated to
finally achieve standby condition.
IMPORTANT: From now on, the indicated factor will become
the calibration factor used by the Meter and will continue to
remain such even after a battery change
10
NO OPERATION
The Meter stores the new work calibration factor and is
ready to begin dispensing, using the USER K FACTOR that
has just been calculated..
H.2.3 DIRECT MODIFICATION OF K FACTOR
If normal Meter operation shows a mean percentage error, this can be corrected by ap-
plying to the currently used calibration factor a correction of the same percentage. In this
case, the percentage correction of the USER K FACTOR must be calculated by the opera-
tor in the following way
New cal. Factor = Old Cal Factor *
( )
100 - E%
100
Example:
Error percentage found: E% - 0.9 %
CURRENT calibration factor: 1.000
New USER K FACTOR: 1.000 * [(100 – ( - 0.9))/100] = 1.000 * [(100 + 0.9)/100] = 1.009
If the Meter indicates less than the real dispensed value (negative error) the new calibra-
tion factor must be higher than the old one as shown in the example. The opposite ap-
plies if the Meter shows more than the real dispensed value (positive error).
ACTION
DISPLAY
1
NONE
METER in Standby.
2
CAL
CAL
CAL
LONG CAL KEY KEYING
Meter enters calibration mode, shows “CAL” and displays the
calibration factor being used instead of the partial. The words
“Fact” and “User” indicate which of the two factors (factory or
user) is currently being used.
3
RESET
RESET
RESET
LONG RESET KEY KEYING
The Meter shows “CAL” and the zero partial total.
Meter is ready to perform in-field calibration by dispensing –
see previous paragraph.
4
RESET
RESET
RESET
LONG RESET KEY KEYING
We now go on to Direct change of the calibration factor:
the word “Direct” appears together with the Currently Used
calibration factor. In the bottom left part of the display, an
arrow appears (upwards or downwards) defining the direction
(increase or decrease) of change of the displayed value when
subsequent operations 5 or 6 are performed.
5
RESET
SHORT RESET KEY KEYING
Changes the direction of the arrow. The operation can be
repeated to alternate the direction of the arrow.
6
CAL
CAL
CAL
CAL
SHORT/LONG CAL KEY KEYING
The indicated value changes in the direction indicated by the
arrow
- one unit for every short CAL key keying
- continually if the CAL key is kept pressed. The speed increase
rises by keeping the key pressed. If the desired value is
exceeded, repeat the operations from point (5).
7
RESET
RESET
RESET
LONG RESET KEY KEYING
The Meter is informed that the calibration procedure is
finished.
Before performing this operation, make sure the INDICATED
value is that required.
8
NO OPERATION
At the end of the calculation, the new USER K FACTOR is shown
for a few seconds, after which the restart cycle is repeated to
finally achieve standby condition.
IMPORTANT: From now on, the indicated factor will become
the calibration factor used by the Meter and will continue to
remain such even after a battery change
9
NO OPERATION
The Meter stores the new work calibration factor and is ready
to begin dispensing, using the USER K FACTOR that has just
been changed.
I
METER CONFIGURATION
The METER feature a menu with which the user can select the main measurement unit,
Quarts (Qts), Pints (Pts), Litres (Lit), Gallons (Gal);
The combination of the unit of measurement of the Partial register and that of the
Totals is predefined according to the following table:
Combination no.
Unit of Measurement
Partial Register
Unit of Measurement
Totals Register
1
Litres (L)
Litres (L)
2
Gallons (Gal)
Gallons (Gal)
3
Quarts (Qts)
Gallons (Gal)
4
Pints (Pts)
Gallons (Gal)
To choose between the 4 available combinations:
1
Wait for the METER to go to Standby
2
CAL
RESET
+
Then press the CAL and RESET keys together. Keep these pressed
until the word “UNIT” appears on the screen together with the unit of
measurement set at that time (in this example Litres / Litres )
3
Every short press of the RESET key, the various combinations
of the units of measurements are scrolled as shown below:
Gal
Qts
Pts
Gal
RESET
RESET
RESET
4
CAL
CAL
CAL
By pressing the CAL key at length, the new settings will
be stored, the METER will pass through the start cycle and
will then be ready to dispense in the set units.
ATTENTION
The Reset Total and Total registers will be automati-
cally changed to the new unit of measurement.
NO new calibration is required after changing the
Unit of Measurement.
L MAINTENANCE
BATTERY
REPLACEMENT
Use 2x1.5 V alkaline batteries size AAA
WARNING
K24 should be installed in a position allowing the
batteries to be replaced without removing it from
the system.
BATTERIES
Check the batteries and terminals at least every
year to ensure proper operation. It is strongly rec-
ommended that terminals be cleaned annually
K24 features two low-battery alarm levels:
1
When the battery charge falls below the first level on the LCD,
the fixed battery symbol appears. In this condition, K24 con-
tinues to operate correctly, but the fixed icon warns the user
that it is ADVISABLE to change the batteries.
2
If K24 operation continues without changing the batteries,
the second battery alarm level will be reached which will
prevent operation. In this condition the battery icon starts to
flash and is the only one to remain visible on the LCD.
TO REMOVE BATTERY
ATTENTION
During meter removal, liquid may spill. Follow the
liquid manifacturer’s safety precautions for clean
up of minor spills.
1
Ensure all liquid is drained from the meter. This could
include draining the hose, meter, nozzle or pipe
2
Wear protective clothing as necessary, loosen both ends
of the meter. Use a wrench only on the meter’s flat metal
surfaces
3
If the meter is not immediately installed again, cap the
hose end or pipe to prevent spills
To change the
batteries, with
reference to the
exploded diagram
positions, proceed
as follows
1
Press RESET to update all the totals
2
Loosen the 4 fixing screws of the lower cover
3
Remove the old batteries and disconnect the plug
4
Place the new batteries in the same position as the old ones
(sure to put the battery in the correct way)
5
close the cover again, by positioning the rubber pro-
tection as a gasket
6
K24 will switch on automatically and normal opera-
tion can be resumed
The K24 will display the same Reset Total, the same Total and the same Partial indi-
cated before the batteries were changed.
After changing the batteries, the meter does not need calibrating again.
ATTENTION
Never blow compressed air trought the meter. It
could damage the rotor.
CLEANING
Only one operation is necessary to clean the k24.
After removing k24 from the plant where it was built
in, any residual elements can be removed by washing
or mechanically-handling.
If this operation does not restore a smooth rotation of
the turbine, it will have to be replaced.
ATTENTION
Do not discard the old batteries in the environment.
Refer to local disposal regulations.
Do not use compressed air onto the turbine in order
to avoid its damage because of an excessive rotation
TO STORE
Follow the liquid manifacturer’s instructions for the
disposal of contaminated cleaning solvents
K24 Front face replacement
1
Carefully remove the screws from the corners of the
front panel, and then carefully lift the front cover up
away from the main body of the meter.
2
Carefully remove the screws from the corners of the
front panel, and then carefully lift the front cover up
away from the main body of the meter.
3
When the new panel is fitted make sure the power
adapter is fitted correctly with the location pin in the
correct way
4
Carefully refit the display panel back onto the main
body making sure the wire is tucked into the corner
and replace the screws
ELECTRONIC ALLUMINIUM
TURBINE
METER - PULSER
M MALFUNCTIONS
Problem
Possible cause
Remedial Action
LCD: no indication
Bad battery contact
Check battery contacts
Not enough measure-
ment precision
Wrong K FACTOR
With reference to paragraph
H, check the K FACTOR
The meter works below
minimum acceptable flow
rate.
Increase the flow rate until an
acceptable flow rate range has
been achieved
Reduced or zero flow
rate
TURBINE blocked
Clean the TURBINE
The meter does not
count, but the flow
rate is correct
Incorrect installation of
gears after cleaning
Repeat the reassembly pro-
cedure
Possible electronic card
problems
Contact your dealer
K24 is switched of
Battery discharged or installed
in the wrong way
Check battery charge and/or
check the battery position
N DISPOSAL
Foreword
If the system needs to be disposed, the parts which make it up
must be delivered to companies that specialize in the recycling
and disposal of industrial waste and, in particular:
Disposing of pack-
ing materials
The packaging consists of biodegradable cardboard which can be
delivered to companies for normal recycling of cellulose.
Metal Parts Dis-
posal
Metal parts, whether paint-finished or in stainless steel, can be
consigned to scrap metal collectors.
Disposal of electric
and electronic
components
These must be disposed of by companies that specialize in the
disposal of electronic components, in accordance with the indica-
tions of directive 2002/96/CE (see text of directive below).
Informa-
tion regard-
ing the en-
vironment
for clients
residing
within the Euro-
pean Union
European Directive 2002/96/EC requires that all equipment
marked with this symbol on the product and/or packaging not
be disposed of together with non-differentiated urban waste.
The symbol indicates that this product must not be disposed of
together with normal household waste. It is the responsibility of
the owner to dispose of these products as well as other electric or
electronic equipment by means of the specific refuse collection
structures indicated by the government or the local governing
authorities.
Miscellaneous
parts disposal
Other components, such as pipes, rubber gaskets, plastic parts
and wires, must be disposed of by companies specialising in the
disposal of industrial waste.
P
EXPLODED VIEWS AND OVERALL DIMENSIONS
/VISTE ESPLOSE ED INGOMBRI
PULSER
112.5
59.6
74.7
45.78
1"NPT
76.3