Piusi K24 Скачать руководство пользователя страница 1

ENGLISH

ENGLISH

ENGLISH

ENGLISH

ENGLISH

ENGLISH

K24

MANUALE D’USO, MANUTENZIONE
E CALIBRAZIONE
USE, MAINTENANCE AND 
CALIBRATION MANUAL

Bulletin    M0225 ITEN   rev. 0

Italiano

English

INDEX

DECLARATION OF CONFORMITY

GENERAL WARNINGS

SAFETY INSTRUCTIONS

 C.1 

SAFETY WARNINGS

 C.2 

FIRST AID RULES

 C.3 

GENERAL SAFETY RULES

  C.4 PACKAGING

 C.5 

PACKAGE CONTENTS/PRE-INSPECTION

BECOMING ACQUAINTED WITH K24

 

D.1 

COMPATIBLE  LIQUIDS

 

D.2 

DISPLAY LCD 

 

D.3 

DISPLAY POSITIONING (METER VERSION ONLY)

 

D2  

USERS BUTTONS

OPERATING MODES

INSTALLATION

DAILY USE

 

G.1 

DISPENSING IN NORMAL MODE

 

 

G.1.1 

PARTIAL RESET (NORMAL MODE)

 

 

G.1.2 

RESETTING THE RESET TOTAL

 

G.2 

DISPENSING WITH FLOW RATE MODE DISPLAY

 

 

G.2.1 

PARTIAL RESET (FLOW RATE MODE)

H  

CALIBRATION

  H1 DEFINITIONS

 

H2 

CALIBRATION MODE

 

 

H.2.1 

DISPLAY OF CURRENT CALIBRATION FACTOR AND 

   RESTORING 

FACTORY 

FACTOR.

 

 

H.2.2 

IN FIELD CALIBRATION

 

 

     H.2.2.1 

IN-FIELD CALIBRATION PROCEDURE

 

 

H.2.3 

DIRECT MODIFICATION OF K FACTOR

METER CONFIGURATION

L MAINTENANCE

M MALFUNCTIONS

N DISPOSAL

DATI TECNICI / TECHNICAL DATA

EXPLODED VIEWS AND OVERALL DIMENSIONS

DECLARATION OF CONFORMITY

The undersigned:

PIUSI S.p.AVia Pacinotti c.m. z.i.Rangavino

46029 Suzzara - Mantova -  Italy

HEREBY STATES

under its own responsibility, that the equipment described below:

Descriprion: METER

Model: K24

Serial number: refer to Lot Number shown on CE plate affixed to productYear of 

manufacture: refer to the year of production shown on the CE plate affixed to the 

productis in conformity with the legal provisions indicated in the directives : 

- Electromagnetic Compatibility Directive 2004/108/EC  

The documentation is at the disposal of the competent authority following motivated 

request at Piusi S.p.A. or following request sent to the email address: doc_tec@

piusi.comThe person authorised to compile the technical file and draw up the decla

-

ration is Otto Varini as legal representative.

     

 

Suzzara, 01/01/2010   

 

Otto Varini

 

 

 

 

 

legal representative.

GENERAL WARNINGS

Important 

precautions

To ensure operator safety and to protect the pump from po-

tential damage, workers must be fully acquainted with this 

instruction manual before performing any operation. 

Symbols used 

in the manual

The following symbols will be used throughout the manual 

to highlight safety information and precautions of particular 

importance:

ATTENTION

This symbol indicates safe working practices for opera-

tors and/or potentially exposed persons.

WARNING

This symbol indicates that there is risk of damage to the 

equipment and/or its components.

NOTE

This symbol indicates useful information.

Manual pres-

ervation

his manual should be complete and legible throughout. It 

should remain available to end users and specialist installation 

and maintenance technicians for consultation at any time.

Reproduction 

rights

All reproduction rights are reserved by Piusi S.p.A.  The text can-

not be reprinted without the written permission of Piusi S.p.A.

THIS MANUAL IS THE PROPERTY OF Piusi S.p.A.

ANY REPRODUCTION, EVEN PARTIAL, IS FORBIDDEN.

SAFETY INSTRUCTIONS

C.1 

SAFETY WARNINGS

Mains - prelim-

inary checks 

before instal-

lation

ATTENTION

You must avoid any contact between the electrical pow-

er supply and the fluid that needs to be FILTERED.

Maintenance 

control

Before any checks or maintenance work are carried out, dis-

connect the power source.

For your safety, 

review the ma-

jor warnings 

and cautions 

below before 

operating your 

meter

When metering flammable liquids, observe precautions 

against fire or explosion

When handling hazardous liquids, always follow the liquid 

manifacturer’s safety precautions

Always dispose of used cleaning solvents in a safe manner 

according to the solvent manifacturer’s instructions.

During meter removal, liquid may spill. Follow the liquid 

manifacturer’s safety precautions to clean up minor spills

Do not blow compressed air through the meter

Do not allow liquids to dry inside the meter

C.2 

FIRST AID RULES

Contact with 

the product

In the event of problems developing following EYE/SKIN 

CONTACT, INHALATION or INGESTION of the treated product, 

please refer  to the SAFETY DATA SHEET of the fluid handled

.

NOTE

Please refer to the safety data sheet for the product

SMOKING  

PROHIBITED

When operating the dispensing system and in particular dur-
ing refuelling, do not smoke and do not use open flame.

ATTENTION

When metering flammable liquids, observe precau-

tions against fire or explosion

When handling hazardous liquids, always follow the 

liquid manifacturer’s safety precautions.

Do not submerge the meter

C.3 

GENERAL SAFETY RULES

Essential pro-

tective equip-

ment charac-

teristics 

Wear protective equipment that is:

suited to the operations that need to be performed;

resistant to cleaning products.

Personal pro-

tective equip-

ment that 

must be worn

Wear the following personal protective equipment during 

handling and installation:

safety shoes

;

close-fitting clothing;

protective gloves;

safety goggles;

Protective 

equipment

instruction manual

WARNING

If handling hazardous liquids, always follow the Liquid 

Manifacturer’s Safety Precautions. Wear protective 

clothing such as goggles, gloves and respirator as 

instructed.

When metering flammable liquids, observe precau-

tions against fire or explosion. Do not meter in the pres-

ence of any source of ignition including running or hot 

engines, lighted cigarettes, or gas or electric heaters

C.4 PACKAGING

FOREWORD

K24 comes packed in a cardboard box with a label indicating the fol-

lowing data:

1 - contents of the pack-

age
2 - weight of the contents

3 - description of the 

product

C.5 

PACKAGE CONTENTS/PRE-INSPECTION

FOREWORD

To open the packaging, use a pair of scissors or a cutter, being careful 

not to damage the dispensing system or its components. 

NOTE

In the event that one or more of the components de-

scribed below are missing from inside the package, 

please contact Piusi inc technical support.

WARNING

Check that the data on the plate correspond to the 

desired specifications. In the event of any anomaly, 

contact the supplier immediately, indicating the nature 

of the defects. Do not use equipment which you suspect 

might not be safe.

BECOMING ACQUAINTED WITH K24

FOREWORD

Electronic digital meter featuring a turbine measurement system, de-

signed for precise measuring of low viscosity fluids.  

K24 is a bi-directional meter with LCD display and calibration 

buttons. The body is made of aluminum (conductive) and de-

signed for high flow 120 l/min. (32 GPM).

K24 is available in 3 versions:

1

a

METER – with LCD display and calibration buttons (SILVER LABEL)

b

METER – with LCD display and calibration buttons (RED LABEL)

2

PULSER – single-channel impulse, cannectable with a remote display.

D.1 

COMPATIBLE LIQUIDS

Turbine 

measurement 

system

The turbine is placed inside a hole through the body of k24,  fitted with 

M-f threaded inlet and outlet. 

The liquids compatible with k24 are at low viscosity, namely:

Diesel fuel

Kerosene

Gasoline

Gasoline blended alcol max 15%

ATTENTION

DO NOT USE WITH SUNDRIES LIQUIDS

Main components K24 Meter

Main components K24 Pulser

1

LCD display

1

Plate with technical data

2

RESET key

3

CAL key

4

Technical data plate

3

2 1

4

1

D.2 

DISPLAY LCD 

FOREWORD

The “LCD” of the METER features two numerical registers and various 

indications displayed to the user only when the applicable function 

so requires.

1

Partial register (5 figures with moving 

comma FROM 0.1 to 99999) indicating 

the volume dispensed since the reset 

button was last pressed

6

Indication of type of total, (TOTAL / 

Reset TOTAL);

2

Indication of battery charge

7

Indication of unit of measurement of 

Totals: 

L=Litres Gal=Gallons

3

Indication of calibration mode

8

Indication of Flow Rate mode

4

Totals register (6 figures with moving 

comma FROM 0.1 to 999999), that can 

indicate two types of Total:

4.1.  General Total that cannot be reset (TOTAL)

4.2.  Resettable total (Reset TOTAL)

9

Indication of unit of measurement of 

Partial:  Qts=Quarts  

Pts=Pints 

                L=Litres  

Gal=Gallons

5

Indication of total multiplication factor 

(x10  /  x100)

4

5

6

9

7

1

2

3

8

D.3 

DISPLAY POSITIONING (METER VERSION ONLY)

FOREWORD

The square shape of the k24 body allows the card to be rotated in its 

housing, thus ensuring great versatility in positioning

This allows easy  display readings in any position. The card housing 

is closed by a plastic cover sealed through a rubber protection act-

ing as a gasket as well. This can be easily removed unscrewing the 

4 screws that fix both the cover and the card (1).

ATTENTION

While fixing the K24 card, make sure the battery con-

tact cable is not placed above the circular housing of 

the bulb.

D4 

USERS BUTTONS

FOREWORD

The METER features two buttons (RESET and CAL) which individ-

ually perform two main functions and, together, other secondary 

functions.

MAIN 

FUNCTIONS 

PERFORMED

- for the RESET key, resetting the partial register and Reset Total 

- for the CAL key, entering instrument calibration mode

SECONDARY

FUNCTIONS

Used together, the two keys permit entering configuration mode 

where the desired unit of measurement can be set.

LEGEND

CALIBRATE MEANS PERFORMING ACTIONS ON THE METER KEYS.  

BELOW IS THE LEGEND OF THE SYMBOLS USED TO DESCRIBE 

THE ACTIONS TO BE PERFORMED

SHORT 

PRES-

SURE 

OF CAL 

KEY

CAL

LONG 

PRES-

SURE 

OF CAL 

KEY

CAL

CAL

CAL

SHORT 

PRES-

SURE 

OF 

RESET 

KEY

RESET

LONG 

PRES-

SURE 

OF 

RESET 

KEY

RESET

RESET

RESET

OPERATING MODES

OPERATING 

MODES

The user can choose between two different operating modes:

The meter features a non-volatile memory for storing the dispensing 

data, even in the event of a complete power break for long periods.

The measurement electronics and the LCD display are fitted in the top 

part of the K24 which remains isolated from the fluid-bath measure-

ment chamber and sealed from the outside by means of a cover.

1 - Normal 

Mode

Normal Mode: Mode with display of Partial and Total dis-

pensed quantities

2 - Flow rate 

Mode

Flow Rate Mode: Mode with display of Flow Rate, as well as 

Partial dispensed quantity.

INSTALLATION

FOREWORD

K24 features a threaded, perpendicular inlet and outlet (1” NPT 

or 

BSP

 male and female that can be combined together).  It has been 

designed to be easily installed in any position: fixed in-line or mobile 

on a dispensing nozzle. In order to improve the life of the turbine, it is 

recommended to fit a strainer before the meter itself

ATTENTION

For installations on system, position meter so that the 

battery housing can be easily reached.

Connections

To protect against the leakage, make sure all threads are 

sealed with two or three turns of thread tape or a sealing 

compound compatible with the liquid being metered

ATTENTION

Make sure the thread tape or sealing compound does 

interfere with flow

Make sure there are no leaks in the connections.

To seal leaks, remove and inspect the meter and re-

place the thread tape or sealant. Refer to the Trouble-

shooting Section 

DAILY USE

FOREWORD

The only operations that need to be done for daily use are partial and/

or resettable total register resetting.  The user should use only the dis-

pensing system of k24.  Occasionally the meter may need to be con-

figured or calibrated. To do so, please refer to the relevant chapters.

Below are the two typical normal operation displays.   One display page shows the par-

tial and reset total registers. The other shows the partial and general total. Switchover 

from resettable total to general total display is automatic and tied to phases and times 

that are in factory set and cannot be changed.

NOTA

6 digits are available for Totals, plus two icons  x 10 / 

x100. The increment sequence is the following:

0.0 -> 99999.9 -> 999999 -> 100000 x 10 -> 999999 x 

10 ->100000 x 100 -> 999999 x 100

G.1 

DISPENSING IN NORMAL MODE

FOREWORD

Normal mode is the standard dispensing. While the count is made, the 

partial and resettable total are displayed at the same time (reset total).

WARNING

Should one of the keys be accidentally pressed during 

dispensing, this will have no effect. 

STAND BY

A few seconds after dispensing has ended, on the lower register, the 

display switches from resettable total to general total: the word reset 

above the word total disappears, and the reset total is replaced by 

the general total.

This situation is called standby and remains stable until the user 

operates the k24 again.

G.1.1  PARTIAL RESET (NORMAL MODE)

The partial register can be reset by pressing the reset key when 

the meter is in standby, meaning when the display screen 

shows the word “TOTAL”.

After pressing the reset key, during reset, the display screen 

first of all shows all the lit-up digits and then all the digits that 

are not lit up. 

At the end of the process, a display page is first of all shown 

with the reset partial and the reset total

and, after a few moments, the reset total is replaced by the 

non resettableTotal.

G.1.2  RESETTING THE RESET TOTAL

VALID JUST FOR METER VERSION WITH SILVER LABEL

The reset total resetting operation can only be performed 

after resetting the partial register. The reset total can in 

fact be reset by pressing the reset  key at length while 

the display screen shows  reset total  as on the following 

display page:

Schematically, the steps to be taken are:

1 Wait for the display to show normal standby display page  

(with total only displayed)

2 Press the reset key quickly
3  The meter starts to reset the partial
4 While the display page showing the reset total is displayed

Press the reset key again for at least 1 second

5

The display screen again shows all the segments of the 

display followed by all the switched-off segments and 

finally shows the display page where the reset Reset 

Total is shown.

C.2 

DISPENSING WITH FLOW RATE MODE DISPLAY

VALID JUST FOR METER VERSION WITH SILVER LABEL
It is possible to dispense fluids, displaying at the same 

time::

1 the dispensed partial
2 the Flow Rate in [Partial Unit / minute] as shown on the 

following display page:

Procedure for entering this mode:
1 wait for the Remote Display to go to Standby, meaning 

the display screen shows Total only

2 quickly press the CAL key.

3 Start dispensing
The flow rate is updated every 0.7 seconds. Consequently, the display could be relatively 

unstable at lower flow rates. The higher the flow rate, the more stable the displayed 

value. 

IMPORTANT

The flow rate is measured with reference to the unit of 

measurement of the Partial. For this reason, in case of 

the unit of measurement of the Partial and Total being 

different, as in the example shown below, it should be 

remembered that the indicated flow rate relates to 

the unit of measurement of the partial. In the example 

shown, the flow rate is expressed in Qts/min.

The word “Gal” remaining alongside the flow rate 

refers to the register of the Totals (Reset or NON Reset)  

which are again displayed when exiting from the flow 

rate reading mode.  

To return to “Normal” mode, press the CAL key again.  If one of the two keys RESET or 

CAL is accidentally pressed during the count, this will have no effect.

IMPORTANT

Even though in this mode they are not displayed, both the 

Reset Total and the General Total (Total)  increase. Their 

value can be checked after dispensing has terminated, 

returning to “Normal” mode, by quickly pressing CAL.

G.2.1  PARTIAL RESET (FLOW RATE MODE)

To reset the Partial Register, finish dispensing and wait for 

the Remote Display to show a Flow Rate of 0.0 as indicated 

in the illustration

then quickly press RESET

H  

CALIBRATION

When operating close to extreme use or flow rate conditions (close to minimum or 

maximum acceptable values), an on-the-spot calibration may be required to suit 

the real conditions in which the K24 is required to operate.

H1 DEFINITIONS

CALIBRATION 

FACTOR OR 

“K FACTOR” 

Multiplication factor applied by the system to the electrical pulses 

received, to transform these into measured fluid units.

FACTORY K 

FACTOR

Factory-set default factor. It is equal to 1,000. This calibration factor 

ensures utmost precision in the following operating conditions:

Fluid 

           diesel

 Temperature:     20°C

Flow rate:              50 lit/min  (13 GPM)

Even after any changes have been made by the user, the factory k 

factor can be restored by means of a simple procedure.

USER K FACTOR: Customized calibration factor, meaning modified by calibration.

H2 

CALIBRATION MODE

Why calibrate? 1

Display the currently used calibration factor:

2

Return to factory calibration (Factory K Factor) after a pre-

vious calibration by the user

3

Change the calibration factor using one of the two previ-

ously indicated procedures

FOREWORD

Two procedures are available for changing the Calibration Factor:

1

In-Field Calibration, performed by means of a dispensing 

operation

2

Direct Calibration, performed by directly changing the 

calibration factor

In calibration mode, the partial and total dispensed quantities indicated on the display 

screen take on different meanings according to the calibration procedure phase. 

In calibration mode, the K24 cannot be used for normal dispensing operations.

In “Calibration” mode, the totals are not increased

ATTENTION

The K24 features a non-volatile memory that keeps 

the data concerning calibration and total dispensed 

quantity stored for an indefinite time, even in the 

case of a long power break; after changing the bat-

teries, calibration need not be repeated.

H.2.1  DISPLAY OF CURRENT CALIBRATION FACTOR 

 

AND RESTORING FACTORY FACTOR.

CAL

CAL

CAL

By pressing the CAL key while the appli-

ance is in Standby, the display page ap-

pears showing the current calibration fac-

tor used.   If no calibration has ever been 

performed, or the factory setting has 

been restored after previous calibrations, 

the following display page will appear:

The word “Fact”  abbreviation for “factory” 

shows that the factory calibration factor 

is being used

If, on the other hand, calibrations have 

been made by the user, the display page 

will appear showing the currently used 

calibration factor ( in our example 0,998) . 

The word “user” indicates a calibration fac-

tor set by the user is being used..

CAL

The flow chart alongside 

shows the switchover logic 

from one display page to 

another

In this condition, the Reset 

key permits switching 

from User factor to Factory 

factor.

To confirm the choice of 

calibration factor, quickly 

press CAL while “User” or 

“Fact” are displayed.

After the restart cycle, the 

K24 uses the calibration 

factor that has just been 

confirmed

CAL

CAL

CAL

CAL

RESET

RESET

TIME OUT

ATTENTION

When the Factory Factor is confirmed, the old User 

factor is deleted from the memory

H.2.2  IN FIELD CALIBRATION

FOREWORD

This procedure calls for the fluid to be dispensed into a gradu-

ated sample container in real operating conditions  ( flow rate,  

viscosity,  etc.) requiring maximum precision. 

ATTENTION

For correct K24 calibration, it is most important to:

When the Factory Factor is confirmed, the old User factor is deleted 

from the memory

use a precise Sample Container with a capacity of not less than 5 

litres, featuring an accurate graduated indicator.

ensure calibration dispensing is done at a constant flow rate equiva-

lent to that of normal use, until the container is full;

Not reduce the flow rate to reach the graduated area of the con-

tainer during the final dispensing stage (the correct method during 

the final stages of sample container filling consists in making short 

top-ups at normal operation flow rate) ;

after dispensing, wait a few minutes to make sure any air bubbles 

are eliminated from the sample container; only read the Real value 

at the end of this stage, during which the level in the container could 

drop.

Carefully follow the procedure indicated below.

H.2.2.1  

IN-FIELD CALIBRATION PROCEDURE

ACTION

DISPLAY

1

NONE

Meter in Standby

CAL

CAL

CAL

LONG CAL key keying

The Meter enters calibration mode, shows <<CAL>> and 

displays the calibration factor in use instead of partial. The 

words “Fact” and “USER” indicate which of the two factors 

(factory or user) is currently in use.

Important: This factor is that which the instrument also uses for 

field calibration measurement operations

RESET

RESET

RESET

LONG RESET key keying

The Meter shows “CAL” and the partial at zero. The Meter is 

ready to perform in-field calibration.

4

DISPENSING INTO SAMPLE CONTAINER

Without pressing any key, start dispensing into the sample 

container  

Dispensing can be interrupted and started again at will. 

Continue dispensing until the level of the fluid in the sample 

container has reached the graduated area. There is no need to 

reach a preset quantity. 

 

 

Indicated value            Real value

5  

RESET

SHORT RESET key keying

The Meter is informed that the calibration dispensing 

operation is finished.  

Make sure dispensing is correctly finished before performing 

this operation.  To calibrate the Meter, the value indicated by 

the partial totaliser (example 9.800) must be forced to the 

real value marked on the graduated sample container. In the 

bottom left part of the display an arrow appears (upwards and 

downwards), that shows the direction (increase or decrease) of 

the value change displayed when the following operations 6 

or 7 are performed.

6  

RESET

SHORT RESET key keying

The arrow changes direction. The operation can be repeated to 

alternate the direction of the arrow. 

7

CAL

CAL

CAL

CAL

SHORT/LONG CAL key keying

The indicated value changes in the direction indicated by the 

arrow

-  one unit for every short CAL key keying 

-  continually if the CAL key is kept pressed. The speed increase 

rises by keeping the key pressed.  If the desired value is 

exceeded, repeat the operations from point (6).

RESET

RESET

RESET

LONG RESET key keying

The Meter is informed that the calibration procedure is finished.

Before performing this operation, make sure the INDICATED 

value is the same as the REAL value.

.

 

  

 

  Indicated value            Real value

The Meter calculates the new USER K FACTOR ; this calculation 

could require a few seconds, depending on the correction to 

be made

ATTENTION:  If this operation is performed after action (5), 

without changing the indicated value, the USER K FACTOR would 

be the same as the FACTORY K FACTOR, thus it is ignored. 

9

NO OPERATION

At the end of the calculation, the new USER K FACTOR is shown 

for a few seconds, after which the restart cycle is repeated to 

finally achieve standby condition.

IMPORTANT: From now on, the indicated factor will become 

the calibration factor used by the Meter and will continue to 

remain such even after a battery change 

10

NO OPERATION

The Meter stores the new work calibration factor and is 

ready to begin dispensing, using the USER K FACTOR that 

has just been calculated..

H.2.3  DIRECT MODIFICATION OF K FACTOR

If normal Meter operation shows a mean percentage error, this can be corrected by ap-

plying to the currently used calibration factor a correction of the same percentage. In this 

case, the percentage correction of the USER  K FACTOR must be calculated by the opera-

tor in the following way

New cal. Factor    =    Old Cal Factor  *

  

(     )

100 - E%

100

Example:
Error percentage found:   E% - 0.9 %

CURRENT calibration factor:  1.000

New USER K FACTOR:          1.000 * [(100 – ( - 0.9))/100] = 1.000 * [(100 + 0.9)/100] = 1.009

If the Meter indicates less than the real dispensed value (negative error) the new calibra-

tion factor must be higher than the old one as shown in the example. The opposite ap-

plies if the Meter shows more than the real dispensed value (positive error).

ACTION

DISPLAY

1

NONE

METER  in Standby.

CAL

CAL

CAL

LONG CAL KEY KEYING

Meter enters calibration mode, shows  “CAL” and displays the 

calibration factor being used instead of the partial. The words 

“Fact” and “User” indicate which of the two factors (factory or 

user) is currently being used.

RESET

RESET

RESET

LONG RESET KEY KEYING

The Meter shows “CAL” and the zero partial total.

Meter is ready to perform in-field calibration by dispensing – 

see previous paragraph.

RESET

RESET

RESET

LONG RESET KEY KEYING

We now go on to Direct change of the calibration factor: 

the word “Direct” appears together with the Currently Used 

calibration factor.  In the bottom left part of the display, an 

arrow appears (upwards or downwards) defining the direction 

(increase or decrease) of change of the displayed value when 

subsequent operations 5 or 6 are performed.

RESET

SHORT RESET KEY KEYING

Changes the direction of the arrow. The operation can be 

repeated to alternate the direction of the arrow. 

CAL

CAL

CAL

CAL

SHORT/LONG CAL KEY KEYING

The indicated value changes in the direction indicated by the 

arrow

-  one unit for every short CAL key keying 

-  continually if the CAL key is kept pressed. The speed increase 

rises by keeping the key pressed.  If the desired value is 

exceeded, repeat the operations from point (5).

RESET

RESET

RESET

LONG RESET KEY KEYING

The Meter is informed that the calibration procedure is 

finished.

Before performing this operation, make sure the INDICATED 

value is that required.

8

NO OPERATION

At the end of the calculation, the new USER K FACTOR is shown 

for a few seconds, after which the restart cycle is repeated to 

finally achieve standby condition.

IMPORTANT: From now on, the indicated factor will become 

the calibration factor used by the Meter and will continue to 

remain such even after a battery change 

9

NO OPERATION

The Meter stores the new work calibration factor and is ready 

to begin dispensing, using the USER K FACTOR that has just 

been changed.

METER CONFIGURATION

The METER feature a menu with which the user can select the main measurement unit, 

Quarts (Qts), Pints (Pts), Litres (Lit), Gallons (Gal);

The combination of the unit of measurement of the Partial register and that of the 

Totals is predefined according to the following table:

Combination no.

Unit of Measurement

Partial Register

Unit of Measurement

Totals Register

1

Litres  (L)

Litres (L)

2

Gallons (Gal)

Gallons  (Gal)

3

Quarts (Qts)

Gallons (Gal)

4

Pints (Pts)

Gallons (Gal)

To choose between the 4 available combinations:

1

Wait for the METER to go to Standby

2  

CAL

RESET

+

Then press the CAL and RESET keys together. Keep these pressed 

until the word “UNIT” appears on the screen together with the unit of 

measurement set at that time (in this example Litres / Litres ) 

Every short press of the RESET key, the various combinations 

of the units of measurements are scrolled as shown below:

                                                           

Gal

Qts

Pts
Gal

RESET

RESET

RESET

CAL

CAL

CAL

By pressing the CAL key at length, the new settings will 

be stored, the METER will pass through the start cycle and 

will then be ready to dispense in the set units.

ATTENTION

The Reset Total and Total registers will be automati-

cally  changed to the new unit of measurement.  

NO new calibration is required after changing the 

Unit of Measurement.

L MAINTENANCE

BATTERY 

REPLACEMENT

Use  2x1.5 V alkaline batteries size AAA

WARNING

K24 should be installed in a position allowing the 

batteries to be replaced without removing it from 

the system.

BATTERIES

Check the batteries and terminals at least every 

year to ensure proper operation. It is strongly rec-

ommended that terminals be cleaned annually

K24 features two low-battery alarm levels:
1

When the battery charge falls below the first level on the LCD, 

the fixed battery symbol appears.  In this condition, K24 con-

tinues to operate correctly, but the fixed icon warns the user 

that it is ADVISABLE to change the batteries.

2

 If K24 operation continues without changing the batteries, 

the second battery alarm level will be reached which will 

prevent operation. In this condition the battery icon starts to 

flash and is the only one to remain visible on the LCD.

TO REMOVE BATTERY

ATTENTION

During meter removal, liquid may spill. Follow the 

liquid manifacturer’s safety precautions for clean 

up of minor spills.

1

Ensure all liquid is drained from the meter. This could 

include draining the hose, meter, nozzle or pipe

2

Wear protective clothing as necessary, loosen both ends 

of the meter. Use a wrench only on the meter’s flat metal 

surfaces

3

If the meter is not immediately installed again, cap the 

hose end or pipe to prevent spills

To change the 

batteries, with 

reference to the 

exploded diagram 

positions, proceed 

as follows

1

Press RESET to update all the totals

2

Loosen the 4 fixing screws of the lower cover

3

Remove the old batteries and disconnect the plug

4

Place the new batteries in the same position as the old ones 

(sure to put the battery in the correct way)

5

close the cover again, by positioning the rubber pro-

tection as a gasket

6

K24 will switch on automatically and normal opera-

tion can be resumed

The K24 will display the same Reset Total, the same Total and the same Partial indi-

cated before the batteries were changed. 

After changing the batteries, the meter does not need calibrating again.

ATTENTION

Never blow compressed air trought the meter. It 

could damage the rotor.

CLEANING

Only one operation is necessary to clean the k24.

After removing k24 from the plant where it was built 

in, any residual elements can be removed by washing 

or mechanically-handling.

If this operation does not restore a smooth rotation of 

the turbine, it will have to be replaced.

ATTENTION

Do not discard the old batteries in the environment. 

Refer to local disposal regulations.

Do not use compressed air onto the turbine in order 

to avoid its damage because of an excessive rotation

TO STORE

Follow the liquid manifacturer’s instructions for the 

disposal of contaminated cleaning solvents

K24 Front face replacement

1

Carefully remove the screws from the corners of the 

front panel, and then carefully lift the front cover up 

away from the main body of the meter.

2

 

Carefully remove the screws from the corners of the 

front panel, and then carefully lift the front cover up 

away from the main body of the meter.

3

When the new panel is fitted make sure the power 

adapter is fitted correctly with the location pin in the 

correct way

4

Carefully refit the display panel back onto the main 

body making sure the wire is tucked into the corner 

and replace the screws 

ELECTRONIC ALLUMINIUM 

TURBINE

METER - PULSER

M MALFUNCTIONS

Problem

Possible cause

Remedial Action

LCD: no indication

Bad battery contact

Check battery contacts

Not enough measure-

ment precision

Wrong K FACTOR

With reference to paragraph 

H, check the K FACTOR

The meter works below

minimum acceptable flow 

rate.

Increase the flow rate until an 

acceptable flow rate range has 

been achieved

Reduced or zero flow 

rate

TURBINE blocked 

Clean the TURBINE

The meter does not 

count, but the flow 

rate is correct

Incorrect installation of

gears after cleaning

Repeat the reassembly pro-

cedure

Possible electronic card 

problems

Contact your dealer

K24 is switched of

Battery discharged or installed 

in the wrong way

Check battery charge and/or 

check the battery position

N DISPOSAL

Foreword

If the system needs to be disposed, the parts which make it up 

must be delivered to companies that specialize in the recycling 

and disposal of industrial waste and, in particular:

Disposing of pack-

ing materials

The packaging consists of biodegradable cardboard which can be 

delivered to companies for normal recycling of cellulose.

Metal Parts Dis-

posal

Metal parts, whether paint-finished or in stainless steel, can be 

consigned to scrap metal collectors.

Disposal of electric 

and electronic 

components

These must be disposed of by companies that specialize in the 

disposal of electronic components, in accordance with the indica-

tions of directive 2002/96/CE (see text of directive below).

   Informa-

tion regard-

ing the en-

vironment 

for clients 

residing 

within the Euro-

pean Union

European Directive 2002/96/EC requires that all equipment 

marked with this symbol on the product and/or packaging not 

be disposed of together with non-differentiated urban waste. 

The symbol indicates that this product must not be disposed of 

together with normal household waste. It is the responsibility of 

the owner to dispose of these products as well as other electric or 

electronic equipment by means of the specific refuse collection 

structures indicated by the government or the local governing 

authorities.

Miscellaneous 

parts disposal

Other components, such as pipes, rubber gaskets, plastic parts 

and wires, must be disposed of by companies specialising in the 

disposal of industrial waste.

EXPLODED VIEWS AND OVERALL DIMENSIONS

 

/VISTE ESPLOSE ED INGOMBRI

PULSER

112.5

59.6

74.7

45.78

1"NPT

76.3

Отзывы: