background image

HemiTor Operating Manual 

 

12 

Piranha Color Systems 

 

Enter desired speed (RPM) for test                   

                               

 

Press 

OK

 

 

Place graduated cylinder beneath the filling nozzle, 

DO NOT let the nozzle touch the graduated cylinder.

 

                             

 

Enter target volume (in mL).  Press 

OK

 when done 

 
 

                              

 

Are Lines Primed

?  Press 

OK 

 

 
 
 

Содержание HemiTor

Страница 1: ...HemiTor Precision Pump System Piranha Color Systems 3531 Thomas Rd Santa Clara Ca 95054 www HemiTorPump com HemiTor Operating Manual ...

Страница 2: ...from peristaltic pumps Even at low RPM peristaltic pump operation can lead to undesirable foaming bubbles and cell shear The HemiTor pump system is a groundbreaking new product for the manufacturing and bioprocess lab that enables a completely integrated disposable flow path liquid handling system that is pulse free spallation free and single use Bags filters and tubing assemblies can now be self ...

Страница 3: ...g Started 4 Overview 5 Features 5 User Instructions 6 Scale Calibration 7 System Setup Weighing Point 7 Pump Actions Operating Modes 9 Pump RPM continuous mode 10 Flow Duration 13 Weight Fill 15 HemiTor Pump Cassette Installation 16 Technical References 17 ...

Страница 4: ...HemiTor Operating Manual 3 Piranha Color Systems 110VAC in RJ45 Ethernet Port USB USB C for programming only Factory use only Com port for Flow Sensor Com port for Scale Back panel Power Switch ...

Страница 5: ...HemiTor Operating Manual 4 Piranha Color Systems 6 6 7 8 9 Getting Started HemiTor Scale Functions 1 Start 2 Stop 3 OK 4 Clear 5 Print 6 Gross Weight Tare 7 Zero 8 Tare 9 Input Alpha Num var 2 3 4 5 1 ...

Страница 6: ... can be used for many different fluid transfer applications The basic functions are as follows Accurately Dispenses by weight and volume Continuous Flow Rate mode Timed Flow Rate Dispense mode Simple universal tubing size Fluid Flow Pulse Dampening Input and Output Unique software easy to update for different customer requirements Features The HemiTor Patented Single Use cassette features internal...

Страница 7: ...Systems User instructions Scale Calibration Turn on power switch on back of pump Wait for pump controller to start Screen will show system booting To set and calibrate scale make sure scale is connected Press Login Operating screen will appear ...

Страница 8: ...HemiTor Operating Manual 7 Piranha Color Systems Scroll down to System Setup Press OK Scroll down to Weighing Point Press OK Press Calib Press Modify ...

Страница 9: ...olor Systems Set Max weight Press this key alpha num sym once to set Max d enter 7000 0 Press OK Press Twice to set Units Set to g grams Press OK Set Scale Interval Press OK Scroll to 0 1g Press OK Set Dead Load at Press by Load Press OK ...

Страница 10: ...ss OK Follow on screen instructions place 5 kg weight on scale enter weight number e g 5kg press OK Remove weight and return to case Press Save The pump and scale have now been calibrated Operating Modes Turn on power Controller will boot up login screen will appear Press Login ...

Страница 11: ...ump Actions Pump RPM for manual pump mode Flow Duration for a measured rate for a specific time Weight Fill for dispensing to a specific weight target Pump RPM Press Set Pump speed RPM Press to set input to 123 Enter desired pump RPM Press OK when done ...

Страница 12: ...ering 0 will add 999 days to duration Enter a number then press again for abc enter time increment s m or d e g 1m15s Press OK Target Volume note a flow rate must be calibrated To Calibrate Flow Rate Scale must be connected to calibrate flow rate Press F1 key test will measure 15 second fill time Press OK ...

Страница 13: ...nha Color Systems Enter desired speed RPM for test Press OK Place graduated cylinder beneath the filling nozzle DO NOT let the nozzle touch the graduated cylinder Enter target volume in mL Press OK when done Are Lines Primed Press OK ...

Страница 14: ...Systems Pump will run for 15 seconds Pump will stop after 15 seconds Observe volume dispensed into graduated cylinder Enter Volume dispensed mL Press OK Flow Rate has now been calculated Target Volume feature can now be used Flow Duration Press ...

Страница 15: ...ow Rate Enter Duration Enter Start RPM Press Prime Pump will stop after priming Press Start Pump will run for the designated time period targeting desired flow rate Pump will automatically change its RPM to target desired flow rate over the desired duration starting at user defined Start RPM ...

Страница 16: ...press OK Enter Bulk RPM pump speed for filling press OK Enter Bulk to Fine how far from target for pump to slow from bulk rpm to fine rpm press OK Enter Fine RPM pump speed for delivering fine amounts slow fill Press OK Enter Target Overshoot to stop pump Check Flow Direction Press Start Pump will stop after dispensing User defined amount ...

Страница 17: ...HemiTor Operating Manual 16 Piranha Color Systems HemiTor Pump Cassette Installation ...

Страница 18: ...0 59 0 4 40 40 79 0 5 50 50 99 0 6 60 61 18 0 7 70 71 38 0 8 80 81 58 0 9 90 91 77 1 100 101 97 2 200 203 94 3 300 305 92 4 400 407 89 5 500 509 86 6 600 611 83 7 700 713 8 8 800 815 78 9 900 917 75 1N 1 Newton 1 kg m s2 1p 1 Pond Conversion Factors Units of force Newton SI unit Dyne kilogram force kilopond pound force poundal 1 N 1 kg m s2 105 dyn 0 10197 kp 0 22481 lbf 7 2330 pdl 1 dyn 10 5 N 1 ...

Отзывы: