background image

General

Accuracy

±(0.015% of Setting in Ohms + 0.05 Ohms)

Temperature Drift

± 0.05 Ohms/°C 

Operating Temperature Range

-25 to 60 °C (-10 to 140 °F)

Relative Humidity Range

10 % ≤RH ≤90 % (0 to 35 °C), Non-condensing
10 % ≤RH≤ 70 % (35 to 60 °C), Non-condensing

Size
       With Boot

4.96 x 2.73 x 1.79 inches, 126 x 69 x 45 mm (L x W x H)
5.67 x 3.06 x 2.05 inches, 144 x 78 x 52 mm (L x W x H)

Weight
      With Boot

8.4 ounces, 0.24 kg (including batteries)
11 ounces, 0.32 kg (including batteries)

Batteries

Four “AA” Alkaline 1.5V (LR6)

Battery Life

50 Hours

Optional NiMh Rechargeable battery kit

120 VAC for North America Only; charger,  four NiMh 

batteries, AC & DC cords [Part # 020-0103]

Low Battery

Low battery indication with nominal 1 hour of operation 

left

Protection against misconnection

Over-voltage protection to 60 V dc (rated for 30 seconds)

Display

High contrast graphic liquid crystal display. LED backlighting 

for use in low lit areas.

PIECAL 510B/511B Specifications

(Unless otherwise indicated all specifications are rated from a nominal 23 °C,

70 % RH for 1 year from calibration)

Source

Accuracy 
   From 1 to 10.2 mA External Excitation Current

   Below 1 mA of External Excitation Current

±(0.015% of S 0.05 Ohms)

±(0.015% of S                              + 0.05 Ohms)

Resistance Ranges

0.00 to 410.00, 410.1 to 4001.0 Ohms

Allowable Excitation Current Range

<410 Ohms: 10.2 mA max; steady or pulsed/intermittent
410 to 4001 Ohms: 1 mA max; steady or pulsed/intermittent

Pulsed Excitation Current Compatibility

DC to 0.01 second pulse width

 0.01 mV

mA Excitation Current

Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176

TestEquipmentDepot.com 

Содержание PIECAL 510B

Страница 1: ...atinum 100 alpha 3850 3902 3926 1000 alpha 3850 3750 Ohm Copper 10 50 Ohm Nickel 100 and 120 Ohm Easily set any value quickly to within 0 1 with the adjustable digital potentiometer DIAL plus store any three temperatures for instant recall with the EZ CHECK switch Compatible with all process instruments Connect directly to the RTD inputs of smart transmitters PLCs DCS and multichannel recorders an...

Страница 2: ...ANGING BATTERIES Low battery is indicated by BAT on the display Approximately one to four hours of typical operation remain before the PIECAL 510B 511B will automatically turn off To change the batteries remove the optional rubber boot remove the battery door from the back of the unit by sliding the door downward This allows access to the battery compartment Replace with four 4 AA 1 5V batteries b...

Страница 3: ...Menu will automatically exit after 15 seconds of inactivity countdown timer is displayed TEMP UNITS pressing the knob will toggle between C and F RTD 510 pressing the knob will toggle between the factory configured RTD Pt Cu Ni and Ohms 511 pressing the knob will cycle through the ten different RTD types Pt Cu Ni at different base resistances and alpha values and Ohms AUTO OFF If AUTO OFF is ON th...

Страница 4: ...high SPAN output temperature by moving the EZ CHECK switch to the HI position and turn the e EZ Dial knob until the desired temperature is on the display Press and hold the EZ Dial knob until STORED appears to store the value Release the EZ Dial knob 2 Store your low ZERO output temperature by moving the EZ CHECK switch to the LO position and turn the e EZ Dial knob until the desired temperature i...

Страница 5: ... 100 0 to 700 0 700 0 to 850 0 0 2 0 3 0 4 328 0 to 212 0 212 0 to 1292 0 1292 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm DIN IEC JIS 1989 1 3850 0 00385 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to 1562 0 0 4 0 6 0 7 Pt 1000 Ohm Hy Cal HVAC 1 3750 0 00375 200 0 to 200 0 200 0 to 600 0 600 0 to 850 0 0 2 0 3 0 4 328 0 to 392 0 392 0 to 1112 0 1112 0 to...

Страница 6: ...ration left Protection against misconnection Over voltage protection to 60 V dc rated for 30 seconds Display High contrast graphic liquid crystal display LED backlighting for use in low lit areas PIECAL 510B 511B Specifications Unless otherwise indicated all specifications are rated from a nominal 23 C 70 RH for 1 year from calibration Source Accuracy From 1 to 10 2 mA External Excitation Current ...

Страница 7: ... given under our warranty No responsibility is accepted for damage loss or other expense incurred through sale or use of our equipment Under no condition shall Practical Instrument Electronics Inc be liable for any special incidental or consequential damage Additional Information This product is calibrated on equipment traceable to NIST and includes a Certificate of Calibration Test Data is availa...

Страница 8: ...formation PIECAL 510B Choose one of the ordering codes below PIECAL 510B Pt 100 1 a 1 3850 PIECAL 510B Pt 100 2 a 1 3902 PIECAL 510B Pt 100 3 a 1 3916 PIECAL 510B Pt 100 4 a 1 3926 PIECAL 510B Pt 1000 5 a 1 3850 PIECAL 510B Pt 1000 6 a 1 375 PIECAL 510B Cu 10 PIECAL 510B Cu 50 PIECAL 510B Ni 110 PIECAL 510B Ni 120 PIECAL 511B Includes all ten RTD Types listed above More Than a Simple Boot The opti...

Отзывы: