PIECAL 134 

4-20 Milliamp Loop Calibrator 

Operating Instructions

 

 

 

82 E. Main Street Suite 3.14 

 Webster, NY 14580 

Tel: 585-872-9350 

 Fax: 585-872-2638 

[email protected] 

 http://www.piecal.com/ 

Practical Instrument Electronics, Inc.   Copyright 

 2011.   All rights reserved. 

134-9002 Rev B   7/13/11 

1-4 

 

 

Basic Keypad Operations 

j

 

EZ-Check™ Switch  

 
SOURCE mode: Slide the switch to select from Hi and low range pre-

set values and the mid ranger (Dial) is selectable. Dial the mid range 

value and it will store the value with-in 5 seconds automatically.  
 
READ mode: Slide the switch to recall minimum and maximum 

readings. Press the EZ-Dial™ Knob to clear the stored values. 
 
 

 

 

k

 

SOURCE/OFF/READ Switch

 

Slide the SOURCE/OFF/READ Switch to SOURCE to 

output a mA signal and to do 2 - wire transmitter 

simulation. Use the READ position to read mA signal and 

power & measure 2 – wire transmitter. 
 

ƒ

 

EZ-Dial™ Knob 

Turn the knob to change display in 0.01mA increments. 

Push and turn for faster dialing. Push without turning to 

clear EZ-Check™ HI/LO points in READ mode. 
 
Press twice to select options:  
 
In Source mode select –  
% or mA 
2-Wire Transmitter Simulate 
% or mA  
low power (15V) or High power (24V) 
 
In Read mode select -   
% or mA 
Power and Measure 2-Wire Transmitter 
% or mA 
low power (15V) or High power (24V) 
 
 

HART® Protocol  

An internal jumper enables the Power & Measure 2 – wire 

transmitter mode to be compatible with HART® 

communicators and transmitters. 
 
 

 

EZ-Dial™ Knob 

Adjust the output up and down with the EZ-Dial

 knob. The increment is 0.01 mA (or 0.1 % if display units are % of 4-20 mA.) 

Press while turning to adjust 10X faster – 0.10 mA (or 1.00 %.) 

 

Quick Reference Bar Graph 

The Quick Reference Bar Graph indicates the input and output level to the PIECAL 134 in % of 4-20 mA with 1.0% resolution. If the 

input or output signal is outside the normal operating range of the PIECAL 134 the Quick Reference Bar Graph in source mode will 

flash, in Read mode display over range when above 24.5mA. 

 

Error Conditions 

Bar Graph will flash when any error conditions exist. 

 

HART® Protocol 

Remove the back of the case and remove the jumper that is located in position J6 on the PC board. By doing so it places a 250Ω 

resister in series with the output of the PIECAL 134. This internal resister eliminates the need to add an external load resister when 

communicating with a HART® transmitter. This reduces the typical drive capability to 950Ω. 