Programmer's reference
37
Copyright © 2007 Pico Technology Limited. All rights reserved.
ps3425.en
3.2.8.17 ps3000_ready
short ps3000_ready (
short handle )
This function checks to see if the oscilloscope has finished the last data collection
operation.
Applicability
only. Does nothing if the oscilloscope is in
.
Arguments
handle,
the handle of the required device.
Returns
1
if ready. The oscilloscope has collected a complete block of data
or the auto trigger timeout has been reached.
0
if not ready. An invalid handle is passed, or the oscilloscope is
in streaming mode, or the scope is still collecting data in block
mode.
-1
if device not attached. The endpoint transfer fails, indicating
that the unit may well have been unplugged.
13
14