OPERATING AND INSTALLATION INSTRUCTIONS LBE 250A / LBE 500A
PAGE 18
USER
GENERAL
SPECIALIS
T PERSONNEL
9.4.2 OPERATING TIME OF THE UVC TUBE
Please note!
The UVC tube has a guaranteed radiated power of at least
85 %
over a period of
8000 operating hours
. Never
reset the hours meters without inserting a new UVC tube! A reduced radiated power can cause an insufficient UVC water
disinfection
(see Chapter 8.3)
Instructions for the UVC tube replacement
(see Chapter 15.1)
9.4.3 OPERATING TIME OF THE WATER FILTER
For reasons of operational safety and for hygienic reasons, the water filter should be replaced when its maximally recommended
operating time is reached, however, once every 2 years at the latest.
Instructions for replacing the water filter
(see Chapter 15.3)