5 Installation
20
Version: 1.0.0
PZ103E
P-73x Nanopositioners
Figure 2: Mounting of the protective earth conductor (profile view)
1
Base body of the P-73x
2
Flat washer
3
Safety washer
4
Screw
5
Cable lug
6
Protective earth conductor
1. If necessary, fasten a suitable cable lug to the protective earth conductor.
2. Fasten the cable lug of the protective earth conductor using the M4 screw on
the protective earth connection of the P-73x as shown in the profile view.
3. Tighten the M4 screw with at least three rotations and a torque of 1.2 Nm to
1.5 Nm.
4. Make sure that the contact resistance at all connection points relevant for
mounting the protective earth conductor is <0.1
Ω at 25 A.