Mounting the P-725 using the Mounting Holes in the Base Body
(Optional)
NOTICE
Warping of the P-725 when mounted on uneven surfaces!
Mounting the P-725 on an uneven surface could warp the P-725. Warping reduces the
accuracy.
Mount the P-725 onto a flat surface. The recommended
flatness of the surface is ≤10 µm.
For applications with large temperature changes:
Mount the P-725 onto surfaces that have the same or similar thermal expansion properties
as the P-725.
INFORMATION
Note that only the thread adapter on the objective side is required when mounting the
P-725 using the mounting holes in the base body.
The following figures serve as examples and can differ from your model.
Fig. 8: P-725.xxx (example view from above):
Mounting holes (black arrows) and locating holes
(hatched arrows) in the base body
Fig. 9: P-725.1CDE2 (example view from above):
Mounting holes (black arrows) and locating holes
(hatched arrows) in the base body
Requirements
You have read and understood the safety precautions (p. 6).
The P-725 is
not
connected to the electronics.
The flatness of the mounting surface is ≤10 µm.
Tools and accessories
Four M2.5 screws of suitable length (p. 34) for screwing the P-725 to the microscope
Optional: Two suitable locating pins (p. 34) for aligning the P-725
Thread adapter, objective side (refer to "Optional Accessories", p. 11).
Suitable screwdriver