3. If you want to change the parameter values in the E-871.1A1N's volatile memory, you
have the following options:
a) Type the new parameter values into the corresponding input field in the
Active Value
column and press the enter key on the PC's keyboard or click the mouse button
outside of the input field.
➔
The modified parameter value is transferred to the E-871.1A1N's volatile memory.
b) Click
Load and Save Parameters -> Load all startup parameters of the axis / system
from controller.
➔
The values of all axis-related / system-related parameters are loaded from the
E-871.1A1N's nonvolatile memory.
c) Click
Load and Save Parameters > Load parameters from stage database... in the
expanded single axis window.
You can use
Load and Save Parameters > Reload parameters from stage
database…
to reload the currently loaded parameter set.
➔
A selected parameter set for the axis is loaded from the positioner database.
4. If you want to change parameter values in the E-871.1A1N's nonvolatile memory, you
have the following options:
a) Type the new parameter value into the corresponding input field in the list's
Startup
Value column and press the enter key on the PC's keyboard or click the mouse button
outside of the input field.
➔
The changed parameter value is transferred to the E-871.1A1N's volatile memory.
b) Click
Load and Save Parameters -> Save all currently active axis / system parameters
as startup parameters to controller.
You can skip parameters that do not have write access on the current command level.
➔
The values of all axis-related / system-related parameters are written from the
E-871.1A1N's volatile memory to the nonvolatile memory.
If a parameter value in the volatile memory (
Active Value column) is different from the
parameter value in the nonvolatile memory (
Startup Value column), the line in the list is
highlighted in color.
8.8.4
Creating or Changing Parameter Sets for Positioners
Overview
You can create and edit new parameter records in the PIStages3 database. This can be
required in the following cases, for example:
■
You want to operate a positioner with different servo control parameter settings than the
one from the default parameter set.
■
You want to adapt the soft limits of the positioner to your application.
■
You have a custom positioner.
Possibilities for creating and editing parameter sets in the PISTAGES3.DB database:
■
You can create a new positioner type easily by changing an existing parameter set in
PIMikroMove and saving it under a new name.
8 E-871.1A1N FUNCTIONS
PZ285EN
‒
1/30/2019
107
M O T I O N | P O S I T I O N I N G