Recipe for calibration solutions: 
To prepare 1 litre of calibration solution you will need: 
 

 

950 ml H

2

0 deionised 

 

3500 mg potassium chloride 

 

50 ml strengthening solution 

 
Take 100 ml of the prepared solution and add 900 ml of dis-
tilled  water  to  it.  You  now  have  900  ml  calibration  solution 
with 3500 mg/l and 1000 ml calibration solution with 350 mg/l 
available. 
Hints: 

To  ensure  measurement  accuracy,  the  measuring 
electrode should  be  immersed  in the  respective  so-
lutions  for  60  seconds  at  constant  temperature, 
buffer  solution  and  stirring  speed  before  calibration 
and measurement. 

Before  changing  samples,  the  electrode  should  be 
completely cleaned  and  the  residual water  carefully 
dried  to  avoid  cross-contamination  between  sam-
ples.  

 
 

5.4 Measured value recording  

For the ion measurement of the media to be examined, they 
must  first  be  mixed  with  the  strengthening  solution  (medi-
um:strengthening solution, factor 50:1). 
The  ion  electrode  is  completely  immersed  in  the  medium  to 
be measured at the measuring head. If a number of different 
test solutions are to be measured, it is advisable to rinse the 
ion  electrode  with  deionized  water  between  two  measure-
ments  and  carefully  shake  off  the  water  to  avoid  cross-
contamination of the samples.  
 
 

5.5 Maintenance and cleaning of the electrode 

After  use,  clean  the  electrode  with  deionized  water  and  dry 
carefully. Fit the protective cover, but make sure that the low-
er  end  of  the  electrode  does  not  rest  on  the  bottom  of  the 
protective cover. The electrode can be stored permanently in 
the activation solution. 
 

5.6 Replacing the battery 
 
Remove the battery 

 

Open  the  sensor  by  turning  the  screw 
cap  on  the  back  of  the  sensor  counter-
clockwise, e.g. with a coin. 
 
 
 
 
 

  

 

 Fig. 5 

 
Lever  the  battery  sensitively, 
e.g.  with  the  help  of  a  small 
screwdriver  or  a  small  pair  of 
scissors,  out  of  its  socket.  In-
sert  the  screwdriver  as  shown 
in Fig. 6. 

 

 

    

 

 

 

 

       

 

 

 

 

 

 

Fig. 6 

 
 
 
 

Insert new battery 

Slide the battery under the golden metal nose (Fig.7-1). Make 
sure that the battery is completely under the metal nose and 
completely pushed to the upper edge. 

Push  the  battery  into  the  socket 
by  pressing  lightly  on  the  oppo-
site side.  
The  battery  slips  under  the  two 
plastic  lugs  (Fig.  7-2),  which  is 
also noticed by a short "click". 
 
 

              
 

    

 

 

    

Fig. 7 

 
 

 

Before  closing,  make  sure  that 
the seal in the lid is not bent and 
lies neatly on the edge of the lid. 
Then  tighten  the  cover  clock-
wise. 

 

 
 
Fig. 8 

6  TECHNICAL DATA  

Operating temperature range: 5 - 40°C 
Relative humidity < 80%. 
 

 

Measuring range 
Resolution 
Accuracy* 

0.

4… 3900 mg/l 

1 mg/l 

±10% 

Max. data transfer rate 
Battery type 
Max. wireless range (open field)  
Dimensions (WxHxD) 

10 Hz 

CR2032 

30 m 

90 x 44 x 23 mm 

Weight (incl. electrode) 
 
*After calibration 

94 g 

7  SCOPE OF DELIVERY 

The scope of delivery includes: 

 

Cobra SMARTsense Nitrates Ion 

12916-00 

 

Nitrate Ion-selective electrode PK-2-004 

 

Instruction manual 

8  ACCESSORIES 

The following accessories are available: 

 

Button cell CR2032, 3 V 

07922-17 

 

Cobra SMARTlink 

12999-99 

 

Nitrate-ion compound electrode PK-2-004 

12916-10 

 

USB-Bluetooth-Adapter 

07936-00 

 

Software measureLAB 

14580-61 

 

Free measureApp available from supplier portals 