background image

 

 

 

 

 

 

www.phywe.com, © All rights reserved 

 

12944-00 / 4021

 

3  FUNCTIONAL AND OPERATING ELEMENTS 

3.1 Operating elements 

The sensor has an on-button and two LEDs for indicating the 
Bluetooth and battery charge status. 
 

On-button 

To switch the sensor on and off in Bluetooth mode, the power 
button must be pressed for longer than 3s. If the sensor is to 
be connected via USB, it is not necessary to press the power 
button. 

Bluetooth-LED  

Flashing red every 2 seconds 

Not connected 

Flashing green every 2 seconds 

Connected to the ter-
minal device 

Flashing green every 4 seconds 

Running measurement 

 

Battery charge LED 

Flashing red every 2 seconds 

Low battery 

Illuminated red 

Active charging process 

Illuminated red 

Charging process 
completed 

 
3.2 Function elements 

There is a Luer-Lock connector on the front of the sensor, to 
which the supplied cuff must be connected. 
 

3.3 USB port 

The  battery,  which  is  permanently  installed  in  the  sensor,  is 
charged  via  the  type  C USB  port.

 

Furthermore,  communica-

tion with a computer takes place via this interface. 

4  NOTES ON OPERATION 

This    device  fulfils  all  of  the  technical  requirements  that  are 
compiled in current EC guidelines. The characteristics of this 
product qualify it for the CE mark. 
This instrument is only to be put into operation under special-
ist supervision in a controlled electromagnetic environment in 
research, educational and training facilities (schools, universi-
ties, institutes and laboratories). 
 The  individual  connecting  leads  are  each  not  to  be  longer 
than 2 m. 
The instrument can be so influenced by electrostatic charges 
and  other  electromagnetic  phenomena  (HF,  bursts,  indirect 
lightning  discharges)  that it  no  longer  works  within  the  given 
specifications. Carry out the following measures to reduce or 
eliminate  the  effect  of  such  disturbance:  Ensure  potential 
equalization at the PC (especially with Laptops). Use screen-
ing.  When  a  total  failure  of  the  instrument  occurs,  unplug  it 
and plug it back in again for a reset. 
 

5  HANDLING 

This section describes the start-up of the sensor and the re-
cording of measurement  data.  Please  read this section  thor-
oughly in order to avoid failures or operating errors. 

 
5.1  Charging process 

Use  a  USB-C  cable  to  connect  the  sensor  to  a  computer  or 
USB charger (not included). 
During  the  charging  process,  the  battery  charge  LED  lights 
up  red.  When  the  charging  process  is  complete,  the  battery 
charge  LED  lights  up  green.  The  charging  time  for  a  com-
pletely discharged battery is 3 hours maximum. 

 

Disconnect the charger at the latest four hours af-
ter the completion of the charging process. Other-
wise,  the  service  life  of  the  battery  may  be  nega-
tively affected. 

 
5.2  Putting on the blood pressure cuff 

 
The  enclosed  blood  pressure  cuff  must  be  connected  to  the 
sensor unit with the open hose end (1) 

 
 
 
 
 
 
 
 
 
 
 

 

  Fig. 2  

 

 

Fig. 3 

 
 
Slide  the  cuff  over  the 
upper arm until the lower 
edge  of  the  cuff  is  2-3 
cm  above  the  crook  of 
the arm. 
 
When  used  on  the  left 
arm, the air hose runs to 
the  device  in  the  middle 
of  the  crook  of  the  arm. 
This  automatically  marks 
the 4 cm long artery is placed 

 

 

Fig. 4 

icentrally above the pulse point. 
 
When  using  the  device  on  the  right  arm,  the  cuff  must  be 
turned to the left until the artery marking is positioned on the 
pulse  point.  The  air  hose  then  runs  along  the  inside  of  the 
upper arm. 
The  cuff  should  be  tight  enough  to  allow  2  fingers  to  fit  be-
tween the arm and the cuff. Now pull the free end of the cuff 
tight and close the Velcro fastener. 
 
Check that the measuring arrow on the cuff is within the "arm 
circumference mark" on the edge of the cuff. 
 
 
 
 
 
 
 

   

 

 

 

Fig. 5 

 
Place  the  arm  with  the  cuff  loosely  stretched  out  on  a  table 
surface and keep it still during the measurement. The palm of 
the hand points upwards. 
 
 
 
 

Отзывы: