LIFEPAK CR Plus/EXPRESS Defibrillator
Preventive Maintenance
CHARGE-PAK Battery Charger Maintenance
The CHARGE-PAK battery charger provides a trickle charge to the defibrillator’s
internal battery. To prevent damage to the internal battery, always keep a
functional battery charger in the defibrillator, including during storage and
shipping.
Replace the battery charger if:
■
The device has been attached to a patient during an event, and the battery
charger has not been replaced.
■
The device has been on for more than 30 cumulative minutes without a
patient connected to it, the battery charger has not been replaced, and the
device is turned off.
■
The battery charger has been in the device two years or more and the device
is turned off.
■
The
BATTERY
indicator appears in the Readiness Display.
Note:
It is important to store the defibrillator at an operating temperature
range of 0° to 50° C (32° to 122° F) when the battery is installed. If the
ambient temperature is outside this range, the defibrillator may erroneously
detect and report a problem.
(Continued on next page)