LIFEPAK 20/20e Defibrillator/Monitor
Troubleshooting
Service Code Table
(continued)
(Continued on next page)
Code
Service Code Description
Corrective Action
501d
RTC out of sync
1. Replace the coin cell battery
2. Replace the A01 System PCB
501e
System software execution error
Reload the device software*
501f
System software read error
Reload the device software*
5020
System software write error
Reload the device software*
5021
System software error
Reload the device software*
5022
System software exception code
Reload the device software*
5023
System software exception code
Reload the device software*
5024
System software exception code
Reload the device software*
5026
NVRAM low battery interrupt
Replace the coin cell battery
5027
System USB did not initialize
1. Clear error and perform PIP
2. Replace the A01 System PCB
5028
PP USB did not initialize
1. Clear error and perform PIP
2. Replace the A02 PP PCB