Chapter 5
| Therapy
©2007-2015 Physio-Control, Inc.
LIFEPAK 15 Monitor/Defibrillator Operating Instructions
125
Electrodes or Therapy Cable Off
If therapy electrodes are not connected,
the
CONNECT ELECTRODES
prompts
occur until the patient is connected.
If the therapy cable is not connected to the
defibrillator, the
CONNECT CABLE
message
appears until the cable is connected.
Shock Counter
The shock counter (x) indicates how
many shocks have been delivered to the
patient. The shock counter resets to zero
whenever the defibrillator is turned off for
longer than 30 seconds.
CPR Time and Metronome
During use, CPR time shown on the
countdown timer will vary slightly due to
the metronome. When the CPR
metronome is active during use, CPR
times are adjusted to end CPR
compression “tocks” on a compression
cycle. As a result, the CPR countdown
timer shows CPR times that approximate
the seconds selected in Setup mode.
Even if the metronome is off or silent
during CPR time, the CPR time displayed
will vary slightly from the time set up in
Setup mode. This is because the
metronome keeps track of compression
“tocks” and ventilation prompts in the
background so that if the metronome is
activated, the CPR time ends with
compressions.
Содержание LIFEPAK 15
Страница 1: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 2: ......
Страница 3: ...MONITOR DEFIBRILLATOR Operating Instructions ...
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 20: ......
Страница 22: ...Front View 22 LIFEPAK 15 Monitor Defibrillator Operating Instructions Area 1 Figure 2 Area 1 Controls ...
Страница 44: ......
Страница 46: ......
Страница 112: ......
Страница 114: ......
Страница 148: ......
Страница 150: ......
Страница 160: ......
Страница 174: ......
Страница 190: ......
Страница 198: ......
Страница 200: ......
Страница 222: ......
Страница 244: ......
Страница 250: ......
Страница 252: ......
Страница 258: ......
Страница 262: ......
Страница 268: ......