Phonak Roger iN Скачать руководство пользователя страница 3

 

 

Phonak Roger iN microphone installation guide          3 

Uninstalling Roger receiver from a hearing device 

 
Uninstallation steps:  

1.

 

Switch on the Roger iN microphone 

2.

 

Press and hold the install button for two seconds until the indicator light starts to blink green 

 

 

 

 

 

 

Roger Select iN   

 

 

Roger Table Mic II iN 

 

 

Roger Pen iN 

 

3.

 

Place the Roger iN microphone on the table 

4.

 

Switch on one hearing device (keep other hearing devices switched off) 

5.

 

Place the hearing device close to the Roger iN microphone (within 10 cm / 4 inches) 
 

 

 

6.

 

Wait approximately 5 seconds until the indicator light on the Roger iN microphone turns solid green 

7.

 

Repeat step 2 to 6 with the second hearing device if needed. 

 
In case the indicator light turns red after the uninstallation procedure, the uninstallation was not successful. This could be 
for the following reasons:  

-

 

The distance between the Roger iN microphone and the hearing device was too far 

-

 

The Roger iN microphone has already two Roger receivers installed 

-

 

The hearing device does not have Roger installed 

-

 

The hearing device is not RogerDirect compatible. 

V1.

00/

201

9

-0

5

 © 

Sono

va

 A

G

 A

ll r

ight

s r

es

er

ve

Отзывы: