General information
3-15
Starting up the
ventilator
NOTE:
Upon power-on the ventilator automatically runs a test of the backup
audible alarm followed by the primary audible alarm. You should hear
a high-pitched tone, followed by a beep. If you do not hear all of these
sounds, discontinue use of the ventilator and have it serviced.
1. Power on the ventilator with the ON/Shutdown key.
2. Verify the ventilator operation, as described on page 5-8.
Shutting down the
ventilator
Shut down the ventilator as follows:
1. Press and release the ON/Shutdown key. The Shutdown window opens.
2. Select Ventilator Shutdown. The ventilator shuts down.
NOTE:
Improper shutdown may cause a Power has been restored message the
next time the ventilator is turned on.
NOTE:
If the screen is blank and the dialogue box cannot be displayed, shut
down the ventilator by pressing the ON/Shutdown key, then the Accept
button on the top-right front of the ventilator.
Training
Product training is available. Contact your local Philips sales representative or
Philips Customer Support for assistance. Call 1-800-225-0230 for ordering
and 1-800-722-9377 for service.
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......