Technical specifications
11-9
Alarm-related
specifications
Other specifications
Table 11-12: Alarm-related specifications
Parameter
Specification
Delay time from the onset of the
alarm condition to the point that
the representation of the alarm
condition leaves the signal out-
put port
< 10 ms from onset of the alarm condition to the
transmission of the signal
< 500 s until alarms received from the ventilator are
displayed on the IntelliVue patient monitor. This de-
lay is in addition to the alarm detection and process-
ing delays of the external device.
Remote alarm delay time
The time it takes the message to appear on the re-
mote alarm depends on the characteristics of the de-
vice.
Maximum time from the alarm
condition triggering event to the
generation of an alarm signal
All alarms except those listed below: < 10 s
• Patient disconnect alarm: 11 s
• Proximal pressure line disconnect alarm: 15 s
• Low Inspiratory Pressure (LIP) alarm:
user-configurable from 5 to 60 s
• Check vent: Blower temperature high: 10 minutes
Mean delay, or the sum of mean
delays, from the triggering event
to the generation of an alarm sig-
nal
All alarms except those listed below: < 5 s
• Cooling fan failure alarm: 10 s
• Blower temperature alarm: 10 s
• ADC wraparound: 6 s
Audio alarm loudness
*
* In accordance with 3rd Edition testing methods
Highest volume setting: Average sound pressure level
is approximately 76 dB(A)
Lowest volume setting: Average sound pressure level
is approximately 62 dB(A)
High-priority auditory alarm sig-
nal sound pressure level range,
measured per
ISO 3744:2010 ANSI/
Average sound power level is approximately 54 dB(A)
measured at the ventilator
Average sound pressure level is approximately 46
dB(A) measured 1 m from the ventilator
Table 11-13: Other specifications
Parameter/Characteristic
Specification
Expected ventilator service life
10 years
Flow delivery
150 L/min at 40 cmH
2
O at 1951 m (6400 ft) altitude
(10% degradation in flow at 2286 m (7500 ft))
Flow range
0 to 240 L/min BTPS
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......