Technical specifications
11-3
Patient data
Table 11-3: Patient data ranges, resolutions, and accuracies during ventilation
Parameter
Range
Resolution
Accuracy
Patient data window
Breath phase/trigger indicator
Spont, Timed, Exhale
Color-coded display: Spont -
turquoise, Timed - orange, Ex-
hale - blue
N/A
PIP
0 to 50 cmH
2
O
1 cmH
2
O
± 2 cmH
2
O
Pt. Leak
0 to 200 L/min BTPS
1 L/min
N/A
Pt. Trig
0 to 100%
1%
± 10%
Rate
0 to 90 BPM
1 BPM
± 1 BPM
T
I
/T
TOT
0% to 91%
1%
± 5%
Tot.Leak
0 to 200 L/min BTPS
1 L/min
N/A
E
0 to 99.0 L/min BTPS
0.1 L/min
± 15% or 0.3 L/min (whichev-
er is greater)
V
T
0 to 3500 mL BTPS
1 mL
± 15% for volumes above
200 mL
Note: Accuracy specification
was measured with patient
circuit PN 582073
Waveform window
P waveform
0 to 50 cmH
2
O
Time axis: 1 second
N/A
waveform
-240 to 240 L/min BTPS
Time axis: 1 second
N/A
V waveform
0 to 3500 mL BTPS
Time axis: 1 second
N/A
Содержание Respironics V60
Страница 1: ...Respironics V60 V60 Plus Ventilator User Manual ...
Страница 2: ......
Страница 8: ...viii ...
Страница 28: ...2 10 Symbols ...
Страница 44: ...3 16 General information ...
Страница 60: ...4 16 Principles of operation ...
Страница 74: ...5 14 Setting up the ventilator for use ...
Страница 102: ...6 28 Operation ...
Страница 110: ...7 8 High flow therapy ...
Страница 114: ...8 4 Patient monitoring ...
Страница 132: ...9 18 Alarms messages and troubleshooting ...
Страница 142: ...10 10 Care and maintenance ...
Страница 162: ...A 10 First time installation ...
Страница 182: ...D 10 Regulatory compliance ...
Страница 198: ...E 16 Diagnostic mode ...
Страница 210: ...Index Index 6 ...
Страница 211: ......