Answers to the Tutorial for Troubleshooting the Instrument

218

  

Troubleshooting the Instrument

Answers to the Tutorial for Troubleshooting the 
Instrument

1) c. (See page 197 for more details.)

2) a. (See page 208 for more details.)

3) e. (See pages 5-2 to page 175 for more details.)

4) c. (See pages 197 for more details.)

Содержание M3000A

Страница 1: ...I E N T M O N I T O R I N G S E R V I C E G U I D E M e a s u r e m e n t s a n d M o n i t o r i n g M3 M4 Monitors M3046A Measurement Server M3001A and M3000A Measurement Server Extensions M3015A and M3016A ...

Страница 2: ...Printed in Germany 03 05 Sixth Edition M3046 9300F Part Number M3046 9300F 4512 610 07711 S ...

Страница 3: ...300F Reordering Number 4512 610 07711 Printed in Germany March 2005 Seventh Edition M3046A M2 M3 M4 Monitors M3000A M3001A Measurement Servers M3015A M3016A Extensions to the Measurement Servers Service Guide ...

Страница 4: ...arranties or merchantability and fitness for a particular purpose Philips Medical Systems shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing performance or use of this material 1995 2005 Koninklijke Philips Electronics N V All Rights Reserved Reproduction in whole or in part is prohibited without the prior written consent of...

Страница 5: ...ous page are not considered revised The documentation printing date and part number indicate its current edition The printing date changes when a new edition is printed Minor corrections and updates which are incorporated at reprint do not cause the date to change The document part number changes when extensive technical changes are incorporated First Edition July 1997 Second Edition February 1999...

Страница 6: ...epairs are carried out by persons authorized by Philips and the electrical installation of the relevant room complies with national standards and the instrument is used in accordance with the instructions for use To ensure safety use only those Philips parts and accessories specified for use with the Monitor If non Philips parts are used Philips Medical Systems is not liable for any damage that th...

Страница 7: ...hould Use this Book If you are a biomedical engineer or a technician responsible for troubleshooting repairing and maintaining Philips s patient monitoring systems this book is designed for you If you are new to Philips s product line or monitoring systems you may find this book helpful as an orientation to the equipment If you have already worked on the systems and now want further details on how...

Страница 8: ...vi ...

Страница 9: ...e M3016A Measurement Server Extension 10 A Quick Description of the Main Screen 10 Theories of Operation and Functional Descriptions 11 Section 1 Monitor Description 11 Monitor Theory of Operation 12 Display and User Interface Software Module 13 Alarm Manager Software Module 13 Admit Discharge Transfer ADT Software Module 13 Trend Software Module 13 Events Software Module 13 Printer Manager 14 Rec...

Страница 10: ...al Oxygen Saturation and Pleth SpO2 PLETH Measurement 33 Description 33 Measurements 33 SpO2 PLETH Features 33 Block Diagram of the SpO2 PLETH Circuit 34 Theory of Operation 34 Temperature and Invasive Blood Pressure Temp Press measurement 37 Description 37 Measurements 37 Temp Press Features 37 Block Diagram 38 Theory of Operation 38 Temperature and Invasive Pressure Software 39 Block Diagram of ...

Страница 11: ...ty 64 Patient Safety 64 Patient Leakage Current 64 Preparing to Install the Monitor 64 Power Source Requirements 64 Protecting against Electric Shock 64 65 Equipotential Grounding 65 Combining Equipment 65 Environment 66 Explanation of symbols used 67 Unpacking the Monitor 70 Installing the Monitor 70 Connecting the Measurement Server 71 with the Measurement Server directly on the Monitor 71 with ...

Страница 12: ... A11 99 Warnings Cautions and Safety Precautions Relating to Wall Mount Installation 100 Installing the Wall Rail M3080A A13 101 Installing the Tilt Swivel Mount M3080A A14 101 Screwed Directly to a Wall 102 Mounted to the GCX Wall Channel M3080A A15 103 Attached to a Universal Pole Clamp 104 Attached to the Monitor 105 Installing the GCX Wall Channel M3080A A15 106 Installing the Universal Pole C...

Страница 13: ... for Maintaining the Instrument 135 Answers to the Tutorial for Maintaining the Instrument 136 4 Testing the Instrument 137 Objectives 137 Concepts 137 Test Reporting 138 Recommendations for Test Frequency 139 Test Map 140 Testing Checklist 141 Serial Numbers 142 Passwords 142 Visual Test 142 Power On Test 142 Functionality Assurance Tests 143 Performance Assurance Test 143 Quick System Check 143 ...

Страница 14: ...solating and Solving Instrument Problems 177 INOP Messages 177 Isolating the Defective Component 183 Part 3 Using Support Functions 197 The Status Log and Error Codes 197 List of Error Codes 199 Testing Wireless Network Connectivity 204 Antenna Check 204 Using Service Mode 206 Service Mode Hardware Tests 206 How To Access the Monitor and Server Revision Screen 209 Troubleshooting the Installed Ins...

Страница 15: ...assis 231 Refitting the Battery 231 Disassembly Procedures for the M3015A Measurement Server Extension 232 Removing the Front Cover 232 Removing the Extension Bottom Cover 233 Removing the CO2 Scrubber 235 Removing the Pump 236 Refit Procedures for the M3015A Measurement Server Extension 237 Refitting the CO2 Scrubber 237 Refitting the Pump 238 Refitting the Extension Bottom Cover 238 Refitting th...

Страница 16: ...5A Measurement Server Extension Parts 266 M3015A Parts 266 M3015A Exchange Parts 268 M3016A Measurement Server Extension Parts 270 M3016A Parts 270 M3016A Measurement Server Extension Exchange Parts 271 Support Related Parts 272 Cables and MSL Through Wall Installation Parts 272 MMS Measurement Server Link MSL Cable Clamp Kit 273 Wireless Assembly Exchange Parts 273 Battery and Battery Related Par...

Страница 17: ...rement Server Hardware Electrocardiogram Respiration ECG Resp Measurement Non invasive Blood Pressure NBP Measurement Arterial Oxygen Saturation and Pleth SpO2 PLETH Measurement Temperature and Invasive Blood Pressure Temp Press measurement Section 3 Measurement Server Extensions Description and Features M3015A Measurement Server Extension Theory of Operation Functional Description of the M3015A M...

Страница 18: ...ware The M3001A Multi Measurement Server is compatible only with the Release E M3046A monitors and is incompatible with any earlier releases The Hemodynamic Extension M3012A is for use together with the Multi Measurement Server M3001A This combination supports the transfer of patient data between IntelliVue M80xxA and M3 M4 M3046A Revision E monitors The M3012A s Pressure Temperature channel works...

Страница 19: ...apter allows use of a vehicle power supply for the instrument and the M8043A Battery Charger and Conditioner allows the recharging and conditioning of batteries of the instrument M3001A C06 M3001A C12 M3001A C18 M3012Ab M3015Ac M3015A C06c M3016A A01c M3016A A02c Key supported not supported a Using a standard 5 electrode lead set You can connect a 10 electrode lead set for example if a patient is ...

Страница 20: ...t Server and Measurement Server Extensions are shown in the following diagram Functional descriptions of these components are to be found later in this chapter Measurement Server M3000A and M3001A Monitor M3046A Measurement Server Extensions M3016A M3015A M3000A has a gray bezel M3001A has a white bezel ...

Страница 21: ...Off Standby AC Power LED Green when AC Power is Connected Battery LED Green Yellow Blinking Red Battery full 95 Battery charging Battery empty On Off Standby LED Green when Monitor is on Silence Reset Key Indicator Infrared Printer Port Battery Compartment AC Power Connector Equipotential Grounding Post ECG Out TouchStrips 100 to 240Vac 50 60Hz Marker In 12V Menu Highlight Up Key Menu Highlight Do...

Страница 22: ...N Software Update Nurse Call Relay Connector 36V Connector to Catches for attaching the Measurement Server the Measurement Server 48V Connector 5V Locking Mechanism for the Measurement Server Connector for an additional display VGA Interface 3 3V Protective earth connector point for additional display ...

Страница 23: ...on of the Measurement Server Overview of the Measurement Server Measurement Connectors for the M3000A C06 M3001A C06 and C18 Measurement Servers Measurement Connectors for the Standard M3000A and M3001A Measurement Server 9 6 7 8 Note Press and Temp cannot be used at the same time 1 2 3 4 5 1 2 3 ...

Страница 24: ...ero key initiates a zero procedure for the connected pressure transducer when pressed and held for a second 3 Red NBP connector 8 Alarm Silence Reset 4 5 Combined pressure red and temperature brown connector connect either invasive pressure transducer or temperature probe M3000A C06 M3001A C06 and C18 only Press and temp cannot be used at the same time You might have a version of the Measurement S...

Страница 25: ...easurement Server Extensions M3015A M3016A Measurement Connectors for the M3015A Measurement Server Extension Connectors to Monitor Measurement Connectors Measurement Server Catches for attaching the Measurement Server SIDESTREAM CO2 Temp Press Note gas outlet exhaust gas inlet Press and Temp M3015A C06 only cannot be used at the same time on the same Extension ...

Страница 26: ...nd Temp cannot be used at the same time on the same Extension 2 Option A01 only SCHLACK ANDREAS Adult 23 11 Alarms Suspended HR 90 65 SpO 2 100 90 ABP sys 150 100 120 70 91 97 70 II 1 mV Pleth ABP 120 0 NBP 60min 17 15 mean 120 90 120 Resp 135 101 12 Wave Wave Label Patient Name Time Patient Size Alarm Message Numeric Alarm Limits Numeric Label Alarm Off Symbol SmartKey Label Icon measurement Quic...

Страница 27: ...al Description of the M3015A Measurement Server Extension Hardware Sidestream CO2 Measurement M3016A Measurement Server Extension Theory of Operation Functional Description of the M3016A Measurement Server Extension Hardware Mainstream CO2 Measurement Section 1 Monitor Description The M3046A Patient Monitor is a small size lightweight monitor with a TouchBar human interface The monitor has a color...

Страница 28: ...a network LAN link Data from the Monitor can be output to a printer via an infrared serial link or via the LAN connector to a central print server The Monitor can communicate with an Philips Information Center via the LAN Connector wired network or via the Wireless LAN Assembly wireless network when the appropriate options are present Each of these modules is described in the following sections Co...

Страница 29: ...rates alarm event triggers for any user defined trigger conditions Admit Discharge Transfer ADT Software Module This module maintains the patient s demographics and controls the upload of trend data from the Measurement Server and the Measurement Server Extension It allows the user to Admit a new patient Transfer a patient to another Monitor Discharge a patient Trend Software Module This module ma...

Страница 30: ...ect between a real time local recording or a delayed recording A real time recording is started by pressing the Local Record SmartKey When a recording is running you can stop it by pressing the key again A delayed recording can be started by pressing the Local Delayed SmartKey Pressing the key again will extend the recording Tabular Trend Report The recorder manager takes raw data from the trend m...

Страница 31: ...eving the printer recorder status The IrDA Serial Communication manager feeds the printer recorder status time out paper out and so on back to the printer or recorder manager which in turn notifies the user of any errors in the print record process Communication Software Module The communication module maintains a data link between the Monitor the Measurement Server and the Measurement Server Exte...

Страница 32: ...C Power PIC Battery PIC Display Adapter Board Inverter Board LCD Display Backlight 48 Pin DC DC Converter 48V current limiter Battery charger Display ECG Out SRL to Measurement CPU System 360 Flash SRAM DRAM HIF 83C552 TouchStrip Keys Sound LEDs Alarm Relay Battery 34 Wire 40 Wire Processor Battery 6 Pin 2x2 Pin 31 Pin Flat Cable Flat Cable cable Flex cable Connector 6 Pin Connector 5 Pin Connecto...

Страница 33: ...erface board The Bezel Assembly connects to the System Board via a 34 wire flat ribbon cable Speaker Connected to the System Board with a 2 wire cable The loudspeaker provides the audible output for alarms and audible feedback when the user presses a manual control AC Power Supply Connected to the Connector Board to power the Instrument and or charge the battery depending on the operating mode Sma...

Страница 34: ...a Nurse Call to be connected to the device Battery Controller The battery controller is the interface to the Smart battery Signals from the battery inform the battery controller of the most effective charging current with which to load the battery Two LEDs are mounted on the Battery Controller Battery LED This is illuminated green if the battery is fully charged and yellow if the battery is chargi...

Страница 35: ...e the Monitor and provides the software drivers for the RF Board The RF Board transforms wired LAN signals into 2 4 GHz signals for transmission The modulation technique FHSS frequency hopping spread spectrum ensures optimum transmission performance The 2 4 GHz band is available worldwide for industrial scientific and medical purposes and is called the ISM band Section 2 Measurement Server Descrip...

Страница 36: ... Lead ECG when used with IntelliVue M3001A C18 As standard package plus Press Temp for M3 M4 Conventional 12 Lead ECG when used with IntelliVue M3000A Measurement Server Standard Package The Measurement Server M3000A standard package includes Measurements of ECG Resp NBP and SpO2 Signal and alarm processing M3000A C06 Measurement Server Extended Measurements Package The Measurement Server M3000A C...

Страница 37: ...ent Server Extension and controlling software Digitized patient information transmitted over the link bar may be waves for ECG pressure respiration pleth numeric information for heart rate pulse pressure values SpO2 and respiration rate or alert information for alarms and assorted status information For Service Procedures there is a special Service Link Bar which is used in place of the standard L...

Страница 38: ... the data to the Monitor via the Server to Monitor link bar The following block diagram shows the main functional components of the Measurement Server The main functional areas are summarized below CPU Board Consisting of a 68360 Controller the Memory System Flash ROM SRAM DRAM ASIC RTC etc the NBP A D Converters and a connector link to a Monitor or an Extension From Patient CPU System CPU Flash R...

Страница 39: ...ed with adult neonatal or pediatric patients in ICU and OR environments Measurements The ECG Resp measurement produces continuous real time waves for both cardiac and pulmonary activity It also generates numerics for the average heart rate HR derived from the ECG and for the respiration rate RR You can use either standard or EASI lead placements with the M3046A Release E together with the M3001A U...

Страница 40: ... QRS complexes are detected automatically In non paced mode no pace pulses are expected and no pace pulse rejection occurs In paced mode pace pulses of channel 1 and 2 are annotated with a small dash on the screen Resp Modes In auto mode the Monitor measures respiration and adjusts the detection level automatically In manual mode the user sets the detection level for measuring respiration This sta...

Страница 41: ...es and lead cables via the input connector Input Protection Network The Input Protection Network and ESU filter eliminate extraneous signals This protects the rest of the circuitry from defibrillator voltages high frequency interference signals and electrostatic discharges ECG ASIC The signals are processed by the ECG Application Specific Integrated Circuit ECG ASIC which has an input amplifier wi...

Страница 42: ...tended ASIC configurations such as gain A D sampling rate and DAC operation Acquires the digitized ECG and Resp signals Digitized ECG signals are used to calculate 3 ECG leads which are then passed on to the Pace Pulse detecting software and the wave processing and filtering software Performs ECG wave filtering The waves are digitally filtered and scaled then passed to the software that communicat...

Страница 43: ...osed on the ECG wave to be displayed ECG waves are communicated to the display unit via the operating system Cardiotach This software derives the averaged value of the heart rate Pace pulses and undershoots are removed from the signal The multi channel QRS detector generates beat identification tags for the cardiotach The detection algorithm weights noisy ECG channels less than channels with clear...

Страница 44: ...he respiration wave The filtered respiration waveform is communicated to the display unit via the operating system Breath Detector This software detects the respiration rate using a trigger threshold that is dynamically adjusted according to the amplitude of the wave The time between breaths is measured and the respiration rate is calculated This is communicated directly to the Resp alarming and v...

Страница 45: ...he measurements as follows Manual For each request one measurement of systolic diastolic and mean pressures is taken Auto Repeated measurements of the three values are taken at timed intervals specified by the user Stat Measurements of the three values are taken immediately and repeatedly over a period of five minutes This method uses a faster measurement procedure but produces a less accurate rea...

Страница 46: ...lated an INOP is generated and the valve opens Mode First Inflation Subsequent Inflations Above Systolic Pressure Stat Mode Adult 165 25 15 Pediatric 130 20 15 Neonatal 100 15 15 Mode Systolic Diastolic Mean Adult 30 270 10 245 20 255 Pediatric 30 180 10 150 20 160 Neonatal 30 130 10 100 20 120 Mode Maximum Measurement Time Maximum Time Pressure Over pressure Maximum Adult 180 seconds 180 seconds ...

Страница 47: ...ion to the Instrument 31 Block Diagram for NBP Components The following components carry out the major signal processing functions within the A D PRE AMP PRE AMP Microcontroller DEFLATION SYSTEM PRESSURE SENSOR PRESSURE SENSOR PRESSURE PUMP CUFF A D ASIC ...

Страница 48: ...tion the cuff is inflated by the pressure pump to a set pressure which is determined by the patient size Thereafter the cuff is inflated by the pressure pump to a cuff pressure above the patient s systolic pressure Depending on the measurement method used inflation occurs once or repeatedly When the cuff pressure is greater than the systolic pressure the artery is occluded the pressure sensor then...

Страница 49: ...ation below Safety To ensure the safety of the patient the patient applied part is isolated from ground by opto couplers and a transformer The circuit is also encapsulated in plastic PLETH Wave The circuit automatically and continuously adjusts the size of the wave which represents the quality of the SpO2 measurement signals Manual wave adjustments in this mode are not possible If the signal quali...

Страница 50: ...ier The photo amplifier is an active input current to voltage converter The input signal is filtered by a low pass filter to eliminate higher frequencies generated for example by electro surgery units Then the input current from the photo diode of the sensor is converted to a voltage Clipping Detector A comparator detects clipping of the photo amplifier signal caused by for example ambient light T...

Страница 51: ... complete circuitry starting at the photo amplifier stage Just before the processing of the patient signal begins the test signal is switched on to check correct functioning of the circuitry RCode Measurement Circuit This circuit measures the coding resistor of the transducer digitizes it and sends it to the SpO2 CPU Each transducer has coding resistors in the connector so that it can be identifie...

Страница 52: ...ed to the alarming software The Average Calculation software receives the SpO2 value for each beat from the SpO2 algorithm and calculates the average over the time interval configured by the user This value is communicated to the Monitor via the operating system and to the Alarming software The SpO2 Alarming software gets the average SpO2 value from the average calculation software and compares it...

Страница 53: ... rate and numeric readings for the systolic diastolic and mean blood pressure values Temp Press Features This illustration shows the user controls and connector for the Temp Press Press Wave Blood pressure is depicted as a pressure wave with the numerics for systolic diastolic and mean pressure values The blood pressure shows the cycles of contraction and release within the heart and the resultant...

Страница 54: ...s follows Excitation Voltage This supplies 5VDC to a connected transducer If a short circuit is detected by this circuit the CPU switches the voltage source off to reduce power consumption Input Protection Network This provides protection for the rest of the circuit against defibrillator voltage electrostatic discharge and any electromagnetic interference The signal is passed on to the analog to d...

Страница 55: ... CPU controls the A D conversion and accesses the digitized data for further processing Central Processing Unit CPU The CPU controls the ADC and receives the digitized pressure or temperature data from the ADC The CPU calculates the temperature values and scales the raw pressure waveform using stored pressure calibration data The CPU stores pressure calibration data and user settings into a non vo...

Страница 56: ...CPU Pressure wave Temperature numeric INOP and error messages Status messages The following data is received from the CPU Control messages All messages in both directions are secured by checksums ADC Controller and Data Acquisition The data exchange between the A D converter and the CPU is based on a serial communication link The calibration information is periodically refreshed to ensure proper o...

Страница 57: ...ed and must be within specified limits This is only done when no pressure transducer is present Invasive Pressure Software Module The Invasive Blood Pressure Module derives three numerical pressure values systolic diastolic and mean a blood pressure waveform and a pulse rate The module is designed to be used with DC pressure transducers 5 µV V mmHg only Alarm limits can be set for pressure and pul...

Страница 58: ...lse values are averaged and converted to the user selected measurement unit mmHg or kPa Pressure Alarming This component generates the appropriate alarm when the averaged pressure or pulse values exceed the user selected alarm limits Technical alarms are generated when the Acquisition Control component detects errors or failures Pressure Pulse User Controls This component controls the user selecte...

Страница 59: ...failures to the Alarming Module Average Calculation This component averages the raw measured temperature values over an interval of 1 second The averaged values are converted to the user selected unit Temperature Alarming This component generates high low alarms if an alarm limit is exceeded Additionally a technical alarm is generated if no temperature can be measured Temperature User Controls Thi...

Страница 60: ...ounted on a bed or a roller stand M3015A Measurement Server Extension The Measurement Server Extension for sidestream CO2 measurement M3015A includes Measurements of sidestream CO2 and Press Temp Signal and alarm processing M3015A C06 only M3016A Measurement Server Extension The Measurement Server Extension for mainstream CO2 measurement M3016A includes Measurements of mainstream CO2 and Press Tem...

Страница 61: ...nt information transmitted over the link bar may be waves e g for ECG pressure respiration etc numeric information for heart rate pulse pressure values and respiration rate or alert information for alarms and assorted status information M3015A Measurement Server Extension Theory of Operation The application specific software for the second pressure temp and CO2 runs on the main CPU in the Measurem...

Страница 62: ...orrectly both the Server and the Extension must have compatible software revisions WARNING Never use a Measurement Extension with a Measurement Server which contains Release A software A XX XX Since the software required to process data from the Extension is absent the additional measurements will not be displayed To view the software revision press the Setup key and select Revisions Pairing the S...

Страница 63: ...DC DC DC Converter Board connecting to the Floating Non floating isolation area on the Front End Board Consisting also of a multiplexer for Front End Link communication to the Measurement Server PRESS TEMP Front End Board consisting of the PRESS TEMP Front End and the Floating Non floating Isolation area all feeding signals to the DC DC Board Power Supply PRESS TEMP Microstream CO2 BOARD MUX 5V 15...

Страница 64: ...nute End Tidal CO2 EtCO2 The highest partial pressure of CO2 measured during one expi ration Inspired Minimum CO2 ImCO2 The lowest partial pressure of CO2 during inspira tion Instantaneous CO2 The CO2 measurement at any instant Ventilation The movement of air in and out of the lungs by inspiration and expiration Measurements The sidestream CO2 measurement produces respiratory CO2 gas readings in a...

Страница 65: ...f 10 mmHg 1kPa Measurement Mode In the measurement mode the Extension continuously measures CO2 concentration calculates medical parameters and exchanges information with the Server via FELP N2O Correction This can be turned on or off If N2O correction is off only oxygen cor rection to CO2 is made In this case it is assumed that the gas mixture consists mainly of O2 and N2 respiratory intensive ca...

Страница 66: ... is assumed that the gas mixture consists mainly of O2 and N2O operating room condi tions O2 Correction There is a fixed correction of 45 O2 Humidity Correction This correction is selectable between Body Temperature Pressure Saturated BTPS and Standard Pressure Temperature Dry STPD The Extension meas ures STPD and uses this correction formula BTPS 0 94 x STPD Max Hold There are three possible sele...

Страница 67: ...used to compensate temperature drift of the CO2 reading Signals from the temperature sensor in the detector are amplified and then passed through an Analog to Digital converter Exciter and Infrared Source The exciter generates a high frequency high voltage signal to ignite the infra red source and to generate the infrared radiation needed to measure the CO2 concentration in the measurement cell Ga...

Страница 68: ...ls The controller reads the digitized values from the analog section and calculates the CO2 wave and numerics based on the measured main reference pressure and temperature signals The controller also controls the CO2 frontend based on the control information received from the Measurement Server the Optical Code Recognition and the measured cell pressure Gas Inlet with Optical Code Recognition The ...

Страница 69: ...ation specific software on the main CPU of the Measurement Server The Temp Press features available on the Extension are identical to those available on the Server The Temp Press selections specific to the Extension are T2 and P2 If you press the Zero key on the Server all invasive pressure channels in use are zeroed To set independent zero and labels for P2 however use the softkeys on the Monitor...

Страница 70: ...ctly both the Server and the Extension must have compatible software revisions WARNING Never use a Measurement Extension with a Measurement Server which contains Release A software A XX XX Since the software required to process data from the Extension is absent the additional measurements will not be displayed To view the software revision press the Setup key and select Revisions Pairing the Serve...

Страница 71: ...r Board DC DC Converter Board connecting to the Floating Non floating isolation area on the Front End Board Consisting also of a multiplexer for Front End Link communication to the Measurement Server Power Supply PRESS TEMP Mainstream CO2 MUX 36 60V 28Vpp 2 Isoblock Optocoupler and Power transformer 36 60 Volt Power Sync RxD TxD FEL Addresses CO2 Transducer Pressure or Temperature Transducer Opto ...

Страница 72: ...of CO2 measured during one expira tion Inspired Minimum CO2 ImCO2 Lowest partial pressure of CO2 during inspiration Instantaneous CO2 The CO2 measurement at any instant Ventilation The movement of air in and out of the lungs by inspiration and expiration Measurements The mainstream CO2 measurement produces respiratory CO2 gas readings in a real time CO2 waveform together with numerics for End Tida...

Страница 73: ...e from 20 mmHg 2kPa to 100 mmHg 13 kPa in steps of 10 mmHg 1kPa Measurement Mode In the measurement mode the Extension continuously measures CO2 concentration and calculates medical parameters N2O Correction This can be turned on or off If N2O correction is off only oxygen cor rection to CO2 is made In this case it is assumed that the gas mixture consists mainly of O2 and N2 respiratory intensive ...

Страница 74: ...re PH20 47mmHg Max Hold There are three possible selections Off The EtCO2 numerics display the breath to breath value 10 seconds The EtCO2 numerics display the highest value within a moving window over a 10 second period 20 seconds The EtCO2 numerics display the highest value within a moving window over a 20 second period Safety To ensure the safety of the patient the patient applied part is isola...

Страница 75: ...ment COUNTERS MICRO PROCESSOR OPTO COUPLER EEPROM DUAL SL0PE A D AZ PULSE MONOFLOP MOTOR PHASE COMPARATOR POSITION PULSE AMPLIFIER A D HEATER CONTROL AMPLIFIER MOTOR DRIVE BIAS REGULATION CO INPUT AMPLIFIER 2 IR SOURCE IR DETECTOR DRIVE COIL SENSE COIL Temp SENSOR HEATER GAS CO MODULE 2 TRANSDUCER M1460A ...

Страница 76: ...Comparator Counters Motor Drive Drive Coil The rotations of the chopper wheel in the transducer are regulated to 40 rotations per second The sense coil together with the motor phase comparator and the counters are used to measure the actual number of rotations per second The microprocessor reads the values out of the counters and adjusts the motor drive output accordingly Infrared Detector CO2 Inp...

Страница 77: ...es the M3046 Monitor support it a A wireless interface functions in the same way as a normal serial interface except that a modulated infrared beam is used to exchange data and status information instead of a wire connection Wireless printing is not presently supported on the M3046 Monitor b A wireless interface functions in the same way as a normal serial interface except that a modulated infrare...

Страница 78: ...the Introduction to the Instrument 62 Introduction to the Instrument Answers to the Tutorial for the Introduction to the Instrument 1 a See page 2 for more details 2 b See page 15 for more details 3 c See page 21 for more details ...

Страница 79: ...s chapter Concepts The following section contains information that you need to understand before attempting an installation of an M3046A Monitor an M3000A M3001A Measurement Server and where present M3015A M3016A Measurement Server Extensions Instrument Grounding The detachable three wire power cable grounds the Instrument to the power line ground when plugged into an appropriate three wire recept...

Страница 80: ...ng or infecting personnel the service environment or other equipment make sure that equipment which has been used before has been appropriately disinfected and decontaminated Power Source Requirements See Electrical Specifications in the Specifications chapter of the Instructions for Use Protecting against Electric Shock The M3046A Monitor is classified as Class I Equipment with an internal power ...

Страница 81: ...quipotential grounding post on the side of the Instrument and the other end is connected to one point of the equipotential grounding system Examinations in or on the heart or brain should only be carried out in medically used rooms incorporating an equipotential grounding system according to national standards Combining Equipment All combinations of medical equipment with non medical equipment mus...

Страница 82: ...tallation allow sufficient room at the front for operation and sufficient room at the rear for servicing with the cabinet access door open Approximately 15 minutes after switch on the Monitor operates within specifications at the ambient temperatures shown in the tables given in Monitor Environmental Specifications and Measurement Server Environmental Specifications in the Specifications chapter o...

Страница 83: ... sent out from the Monitor to a defibrillator or other external device such as an intra aortic balloon pump On the M3015A Measurement Server Extension Gas Outlet Exhaust On the Monitor Data Output to serial recorder Serial interface socket for connection to recorder M3046A J16 only Alternating Current Equipotential Grounding Post see Equipotential Grounding earlier in this chapter Battery Compartm...

Страница 84: ...ve 93 42 EEC of 14 June 1993 Medical Device Directive and Council Directive 1999 5 EC of 9 March 1999 Radio and Telecommunications Terminal Equipment Directive This device complies with FCC part 15 of the FCC rules Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that...

Страница 85: ...t Server Extension The M3001A Multi Measurement Server and M3015A M3016A Measurement Server Extension comply with the requirements of the Council Directive 93 42 EEC of 14 June 1993 Medical Device Directive Prod No M30XXA SN XXXXXXXXXX Opt XXX XXX XXXXXX D 71034 Boeblingen Germany M30XXA Made in Germany 0366 NRTL C R 2003 05 Shows date of manufacture 0366 ...

Страница 86: ...en damaged in transit keep the packing material for inspection and contact your Philips Medical Systems representative immediately Do not use the Monitor if the casing has been damaged If the Monitor is damaged make sure that the screen is not leaking There is no known danger from the fluid of irritation to skin or eyes or by inhalation The median lethal dose if taken orally is 2 0g kg There are n...

Страница 87: ...the Measurement Server with the Measurement Server directly on the Monitor You can connect the Measurement Server to the Monitor by mounting it directly on the Monitor Step 1 Make sure that your Monitor has a link bar If your Monitor does not have a link bar a Position the link bar as shown in the diagram above Make sure that the guide is in the slot under the plug which connects to the Measuremen...

Страница 88: ...ack of the Monitor If it is not tight against the back of the Monitor slip it away from the link bar until it is Step 3 Slip the Measurement Server forward until it clicks into place To remove the Measurement Server from the Monitor move the latch in the middle at the top of the Monitor toward the front of the Monitor and slide the Measurement Server away from the link bar Latch ...

Страница 89: ...ble to the plug on the link bar or directly to the Monitor as follows To remove the link bar a Turn the latch guard away from the edge of the latch b Unlatch the link bar from the back of the Monitor c Slide the link bar away from the Monitor Step 2 Attach the socket end of the cable to the Monitor Step 3 Attach the other end of the cable to the Measurement Server or if present the M3015A M3016A M...

Страница 90: ...Server Extension by sliding it into the grooves on the Measurement Server Extension and clicking it into place Attach the combination of Measurement Server Extension and Measurement Server to the Monitor exactly as described above for the Measurement Server by sliding the Extension into the grooves on the Monitor and clicking it into place Attaching the Monitor to a Mount Step 1 Make sure the fron...

Страница 91: ...t is Step 3 Slip the Measurement Server forward until it clicks into place Detaching the Measurement Server from a Mount Step 1 Press and hold the latch in the middle at the top of the mount away from the Measurement Server Step 2 Slide the Measurement Server off the mount in the direction of the measurement connectors Positioning the Measurement Server on a Clamp Mount For convenience if you have...

Страница 92: ... Call Relay for the notification of alarm conditions The relay output cannot be checked by the Monitor and the Monitor cannot notify the user of any failure of the relay See the specifications for the Nurse Call Relay in the Monitor and Measurement Server Specifications chapter of the Instructions for Use and the documentation for the device you are connecting Modification for Nurse Call Alarm Rel...

Страница 93: ...Very carefully remove all copper or plastic shavings from the board Step 3 Turn board over Step 4 Build the new connection to the Open On Alarm contact shown below Using a piece of wire and solder connect the three 3 points on the bottom of the circuit board as illustrated in the following photographs 1 Cut connector path 2 Remove Shavings 3 Turn board over 4 Build new connection ...

Страница 94: ...n section of the monitor s Instructions for Use for information on assigning a label to the monitor Configuring the Radio Frequency of the M3 M4 Monitor The configuration tool for this procedure is contained on the IntelliVue Information Center Application Software CD ROM that corresponds to your revision of the monitor and Information Center and must be copied to the PC used for the configuration...

Страница 95: ...the configuring PC Step 3 Open Windows Explorer as follows a Click on Start in the lower left of the Windows Main Menu to open the Windows Workstation menu b Click Programs to display the Programs menu c Click Windows Explorer to display folders in the PCs CD ROM drive Find the Viridia directory on the CD ROM shown in the figure below and click on it to display the tools directory Step 4 Copy the ...

Страница 96: ...tory by clicking on the ConfigFiles folder in the tools menu on the configuring PC The files on the right of the previous figure will be displayed Step 2 Click on the file labelled WB_WirelessM3_1 2b15 CFG to select it NOTE The selected file must be made writeable because settings used in the configuration process are saved to this file so they can be reused when the configuration tool is used aga...

Страница 97: ...his file set as Read only the following error message will be displayed Clicking OK will exit the tool and the Read only attribute must be removed to clear this condition Running the Configuration Tool Once the tool has been copied and made writeable it can be run The first step is to select which device Access Point Switch or M3 M4 Monitor to configure Step 5 Run the configuration tools as follow...

Страница 98: ... for the device being configured Access Point Network Switch or Wireless Bedside Step 6 Select the Serial Port to be used COM1 or COM2 in the Select Serial Port field by clicking in the circle preceding the appropriate port Step 7 Select the appropriate Device to Configure in the drop down list in the Configuration Tool window of the configuring PC See previous figure Step 8 Select the file tools ...

Страница 99: ...tep 11 Connect the phono plug end of the 9 Pin D female 1 8 in male Stereo Phono cable into the phono plug connector on the Wireless Adapter as shown above and the 9 pin D end of the cable into the 9 Pin D Serial Port connector on the configuring PC Step 12 Turn ON the M3 M4 Monitor and insure that it passes its self test Step 13 Select the WB_WirelessM3_1 2b15 cfg item in the drop down list in th...

Страница 100: ...uration process takes about 1 minute During configuration status messages will be displayed in the field at the bottom of the Wireless Bedside Parameters window as the tool resets the configuration to factory default values sets the configuration parameters and then resets the Wireless Adapter When the tool has successfully completed the configuration the Wireless Bedside Configuration is complete...

Страница 101: ...llowing the specification of a different filename b enter the new filename in the field provided c click Yes to create a logfile with the new filename If you do not want to create a log file click No to close this window NOTE If a filename is entered for which a configuration file already exists that file will be overwritten by the new file Configuration of this Wireless Adapter is now complete St...

Страница 102: ... COM port is inaccessible or not working one of the following windows will be displayed Check the serial cable and its connections carefully to assure that a good interconnection is being made Ensure that no other applications in the configuring PC have control of its Serial Port e g Hyperterminal Close all other applications Ensure that the proper COM port is selected in the Configuration Tool wi...

Страница 103: ...tware revisions D and later allow the NBP Measurement Characteristic to be changed WARNING Consult technical support before changing the NBP measurement characteristic in Adult or Pediatric mode Changing the NBP measurement characteristic will significantly change the NBP readings especially in the elevated pressure range The setting of NBP Measurement Characteristic is not transferred as part of ...

Страница 104: ... comply with the American National Standard for Electronic or Automated Sphygmomanometers ANSI AAMI SP 10 1992 in relation to mean error and standard deviation when compared to auscultatory measurements in a representative patient population For the auscultatory reference the 5th Korotkoff sound was used to determine the diastolic pressure 3 Factory Default NBP measurement characteristic of M3 M4 ...

Страница 105: ...Release C compared to readings obtained using the manual auscultatory method Korotkoff and direct arterial measurement The data for Figure 1 and Figure 2 was derived from M3 Clinical Evaluations and Auscultatory Measurement of Arterial Pressure During Anesthesia a Reassessment of Korotkoff Sounds E Pereira et al European Journal of Anesthesiology 1985 2 II 20 The setting of NBP Measurement Charact...

Страница 106: ...ons for the VGA interface may be used with your monitor When connected to the patient monitor these displays along with any other Class I non medical equipment must comply with IEC 60950 and with IEC 60601 1 1 when connected to the M3 M4 Monitor If the resulting enclosure leakage current of the display exceeds the value laid down by IEC 60601 1 60601 1 1 then certain action needs to be taken to en...

Страница 107: ...e signal line of the patient monitor is connected to equipment outside of the medically used room for example in another building with a different mains supply circuit A separation device for example an separation transformer included in the signal line prevents a hazard to the patient or user It should be placed as near as practicable to the medical electrical equipment The inclusion of a separat...

Страница 108: ...ion transformer must meet the following requirements Installation WARNING Do not touch the patient when connecting or disconnecting the cable to the Video Out connector CAUTION When an additional display is connected and an additional protective earth connection is needed to be made to the Monitor this protective earth connection may only be used if the leakage current does not exceed the limits o...

Страница 109: ...g a Printer If you are printing locally you can use either of the following accessories a PCL 3 or PCL 5 compatible printer with an infrared to parallel converter JetEye Printer wireless infrared connector which you can order as M3080A Option H05 or a PCL 3 or PCL 5 compatible with a built in infrared interface You can also use the LaserJet with the JetEye but then you will also need a Centronics ...

Страница 110: ... a parallel and an infrared interface To find out how to install the Wireless Infrared Printer Connector software refer to the accompanying documentation Install the device support tray supplied with kit and connector as follows Local To enable a local printer Remotea 1 To enable a printer connected through the network b Remotea 2 To enable a printer connected through the network b Remotea 3 To en...

Страница 111: ...ition to line up the infrared devices Connecting a Local Printer WARNING The printer and any other non medical equipment such as the infrared to parallel converter are not allowed to be used within the patient vicinity 1 5m 4 9 ft If you are using an infrared to parallel converter do the following Step 1 Connect the parallel port to the printer Step 2 Position the converter according to the follow...

Страница 112: ...nics XE 50p to your monitor via the serial recorder interface available as option M3046A J16 A dedicated cable is supplied with the recorder Details of how to set up use and care for the recorder are given in the Operator s Manual that accompanies it Additional information can be found in the Monitor s Instructions for Use Step 1 Connect the 9 pin female connector of the serial interface cable to ...

Страница 113: ...ext into the localized language Site Preparation Guidelines Although the Instrument Monitor and Measurement Server or where present Measurement Server Extension is primarily a portable Monitor some site preparation must be done before using it in certain locations Additionally good operating practices are also recommended Check that the operating environment is reasonably free from dust vibration ...

Страница 114: ...nt M3080A A10 The Table Mount is intended for any flat level surface To mount the plate securely you need 4 screws These are not provided because the type of surface dictates the type of screw Step 1 Unpack the Table Mount and check all parts are available Step 2 Select 4 screws appropriate to surface material Step 3 Insert the 4 screws through the screw holes provided in the plate Step 4 Tighten ...

Страница 115: ...k the Bed Hanger and check all parts are available Step 2 Rotate the arms to their open position Step 3 Hang the hanger on the rail or as appropriate Step 4 If necessary adjust the bed hanger to a vertical position by pulling out the slider at the rear of the hanger Step 5 If necessary fix the bed hanger by hooking the rubber strap around a vertical bar as shown in the insert The Monitor snap fits...

Страница 116: ...he appropriate fasteners and mounts In a new construction or in a remodelled construction where the wall can be completely or partially removed a 16 gauge steel stud should be mounted for the purpose of attaching the wall channel This design safety factor is based on four times 4x the maximum rated load GCX Wall Channel Always reposition the adjustable channel safety stop under the adapter plate w...

Страница 117: ...ll Rail and check all parts are available Step 2 Select 3 screws appropriate to surface material Step 3 Insert the 3 screws through the screw holes provided in the plate Step 4 Tighten the screws Step 5 Hang the bed hanger over the wall rail Installing the Tilt Swivel Mount M3080A A14 The Tilt Swivel mount can be used in the following situations Screwed directly to a wall Mounted to the GCX Wall C...

Страница 118: ...tep 1 Unpack the Tilt Swivel mount and check all parts are available Step 2 Remove the adapter plate this is only necessary for the GCX Wall Channel Step 3 Secure the rail to the Tilt Swivel mount with the 2 screws provided Step 4 Select 3 screws appropriate to surface material Step 5 Insert the 3 screws through the screw holes provided in the plate Step 6 Tighten the screws to secure the mount to...

Страница 119: ...ack the Tilt Swivel mount and check all parts are available Step 2 Make sure the wall channel end stop has been fitted Step 3 Secure the rail to the Tilt Swivel mount with the 2 screws provided Step 4 Slide the adapter plate down the wall channel Step 5 Adjust the assembly to the correct height with the adjustable channel stop Step 6 Use adjustment knobs to adjust to the required tilt swivel angle...

Страница 120: ...p 2 Remove the adapter plate this is only necessary for the GCX Wall Channel Step 3 Secure the rail to the Tilt Swivel mount with the 2 screws provided Step 4 Decide whether the tilt swivel mounting is to be secured for vertical positioning on a pole or for horizontal positioning on a rail Step 5 Secure the pole clamp to the tilt swivel mounting with the two screws provided Step 6 Tighten the scre...

Страница 121: ...el mount directly to the rear of the Monitor The Monitor can then be mounted on the GCX Wall Channel or can be used together with the Universal Pole Clamp Step 1 Unpack the Tilt Swivel mount and check all parts are available Step 2 Secure the tilt swivel mounting to the Monitor rear panel with the 2 screws provided Step 3 Tighten the screws ...

Страница 122: ... the GCX Wall Channel M3080A A15 106 Installing the Instrument Installing the GCX Wall Channel M3080A A15 This option is the GCX Wall Channel only To install it follow the documentation packaged with the channel ...

Страница 123: ...clamp and check all parts are available Step 2 Decide whether the clamp is to be secured for vertical positioning on a pole or for horizontal positioning on a rail The rear of the Monitor is provided with 4 screw holes to cover both positioning possibilities Step 3 Secure the clamp to the Monitor with the 2 screws provided Step 4 Tighten the screws Attach the clamp to the rail or pole as appropria...

Страница 124: ...unting plate It is important for the customer to decide whether the MSL cable clamp is to be used or not at installation time as once the decision is taken to fit the clamp the MSL cable should not be removed from the clamp You cannot fit the MSL cable clamp to earlier versions of the mounting plate Server Mounting Plate M3080A A01 This option includes a pack of 5 mounting plates To mount the plat...

Страница 125: ...de 3 and the positioning notch 4 to ensure you have the cable correctly positioned Step 4 Squeeze open the side clips 5 and ease the cable into the clamp so that the clips grip the gray mouldings 6 Step 5 Press the cable down firmly to ensure that it is fully seated in the clamp NOTE Once fixed do not remove the cable from the MSL clamp 3 2 1 4 5 6 Latch ...

Страница 126: ...face and drill suitable holes Step 7 Insert the 2 screws through the screw holes provided in the plate Step 8 Tighten the screws Step 9 Align the feet on the MMS with the locating shoes 7 on the mounting plate and slide the MMS onto the mounting plate The Server snap fits onto the plate To release the Server press and hold the latch in the middle at the top of the mount away from the Server and sl...

Страница 127: ...plate M3080A A02 starting on page 108 steps 2 to 5 Step 3 Secure the clamp to a pole or rail The Server snap fits onto the plate To release the Server press and hold the latch in the middle at the top of the mount away from the Server and slide it out To rotate the plate press and hold the lever on the clamp away from plate while rotating the plate Mounting Options for the Local Recorder Two mount...

Страница 128: ...re is the same for both connector boxes This means you must perform steps 3 to 8 of this procedure twice The connectors on each box are different so you must ensure that the correct box is placed at the correct location The symbol on the plastic angled cover indicates at which site you should install the box Symbol is connector box in and must be placed at the monitor site Symbol is connector box ...

Страница 129: ...schematic where each color corresponds to a number NOTE The Insertion Tool should be set to cutting mode on The Insertion Tool automatically removes the insulation at the connection site clamps the wire in the correct position and cuts off the excess wire Step 4 Use a small screwdriver to connect the two drain wires to the PCB see the wiring schematic above ...

Страница 130: ...ge to the wall mounting plate Step 7 Mount the plastic cover The plastic cover consists of two pieces Frame Angled cover Put the frame over the mounting insert and the PCB Place the angled cover on top and fasten with two screws Step 8 Connect the monitor and the measurement server to the wall installation Step 9 Perform the following tests as described in the Testing the Instrument section of thi...

Страница 131: ...manufacturer contact your local supplier for details The Monitor and Measurement Server Extension can be disassembled see Repairing the Instrument on page 219 There is no metal moulded into the plastic case and no metal sprays on the plastic All plastic parts with a weight greater than 10g 0 35 ounces are marked with the ISO code for identification All labelling on the product has been done by las...

Страница 132: ...onnecting a cable to a Video Out connector b It is important to read and become familiar with the Monitor and Measurement Server Specifications chapter in the Instructions for Use c When connected to an AC power supply either mains or vehicle the cabinet of the monitor must be grounded d If other patient connected equipment is used simultaneous to the Monitor the summation of leakage currents can ...

Страница 133: ...al for Installing the Instrument Installing the Instrument 117 Answers to the Tutorial for Installing the Instrument 1 b See page 100 for more details 2 e See entire chapter for more details 3 b See page 65 for more details ...

Страница 134: ...Answers to the Tutorial for Installing the Instrument 118 Installing the Instrument ...

Страница 135: ...ng procedures found in Testing the Instrument The tests included in Testing the Instrument are functionality assurance tests performance assurance tests preventive maintenance tests accuracy and performance procedures and patient safety checks used to verify that the Instrument is operating properly Concepts Light Maintenance Light maintenance tasks can be defined as those tasks required to mainta...

Страница 136: ...as needed See Testing the Instrument Preventive Maintenance for the M3000A M3001A Measurement Server Cleaning procedures Non invasive Blood Pressure Cali bration test Testing procedures Clean as needed Once every two years See Testing the Instrument Preventive Maintenance for the M3015A Measurement Server Extension Cleaning procedures Replace pump and CO2 scrubber if required units with serial num...

Страница 137: ...on and Perform ance tests 146 M3015A Replacement of the Pump and CO2 Scrubber units with serial number prefix DE020xxxxx only 125 Sidestream CO2 Performance Test 148 General Cleaning of the Instrument Cleaning Guidelines 125 Battery Handling Maintenance and Good Practices Checking the Battery Status 127 Charging the Battery 129 Conditioning a Battery 130 Instrument Tests for Functionality Assuranc...

Страница 138: ... NOTE Philips recommends replacement of the backlight tube at the latest after 20 000 hours approximately 3 years of use Inspect Cables and Cords Step 1 Examine the line power plug for damage Ensure that the prongs of the plug do not move in the casing Step 2 Inspect the line power cord for signs of damage If damaged replace the entire cord Ensure that any new line power cord or plug is fitted wit...

Страница 139: ...e Backlight Tube Assembly after 3 years 20 000 hours of continuous use or if there is noticeable decrease in the display illumination Display illumination decreases by about half after 3 years of continuous use The procedure for replacing the Backlight Tube Assembly is found in Repairing the Instrument NOTE After fitting new tubes go to Service Mode and reset the Backlight Tube counter to zero M30...

Страница 140: ...x DE020xxxxx only and reset the operating time counter Perform barometric pressure check and calibrate if necessary see Barometric Pressure Check and Calibration on page 149 Perform leakage check see Leakage Check on page 150 Perform Pump Check Perform flow check and calibrate if necessary see Flow Rate Check and Calibration on page 151 Perform Noise Check Perform CO2 calibration check check the d...

Страница 141: ...e Mode and reset the counter for the part s you have replaced The procedure for resetting the counters is found in Repairing the Instrument When the PumpOpTime has been reset an INOP will be generated CO2 OCCLUSION To clear this INOP you must perform a flow check and store the flow in Service Mode select Store Flow Replacement of the Infrared Lamp The infrared lamp can be exchanged only in units w...

Страница 142: ...agent Dilute ammonia Dilute formaldehyde 35 37 Hydrogen peroxide 3 Ethyl alcohol Isopropyl alcohol Dilute sodium hypochlorite laundry bleach NOTE Concentrations ranging from approximately 500 ppm sodium hypochlorite 1 100 dilution of household bleach to 5000 ppm sodium hypochlorite 1 10 dilution of household bleach are effective depending on the amount of organic material e g blood mucus present o...

Страница 143: ...n it is not being used regularly as the main source of power for the monitor Recharge the battery every 3 to 4 months The shelf life of a fully charged battery is about 6 months Once a battery has been activated keep it charged If it is not in use recharge it every 3 to 4 months Condition the battery approximately every 50 recharge cycles If the battery shows damage or signs of leakage replace it ...

Страница 144: ...battery is partly or fully loaded the Monitor starts operating If the new battery is empty If the battery voltage no load is below 11 5V a single red flash is emitted assuming that there is still enough power to emit this single flash If the battery voltage is higher than 11 5V no load the Monitor tries to restart If the bat tery voltage falls below 10 5 V during charging the Monitor registers the...

Страница 145: ... you need to rely on accurate indication of battery operating time access the Battery Status Window and check the TimeToEmpty field see Accessing the Battery Status Window on page 132 If the battery is empty the inner bar is gray If the Monitor cannot access the battery because of a battery or hardware malfunction the text string Malfunction is displayed on a black background inside the battery ga...

Страница 146: ...ttery Status What Causes the Conditioning Message on the Monitor In addition to the value for the full capacity the battery also stores a value for the Max Error The Max Error tracks the maximum possible deviation of the estimated charge of a battery from the actual charge If a battery is charged or discharged partially or if it is charged while the monitor is being used the accuracy of the refere...

Страница 147: ...at accompanies the charger for details on charger use You can leave batteries in the Battery Charger and Conditioner M8043A as no trickle charging takes place Conditioning a Battery Using the Monitor If you do not have access to a Battery Charger and Conditioner M8043A you can condition the batteries using the monitor We recommend two full charge and discharge cycles to ensure proper conditioning ...

Страница 148: ...the battery once a month See the instructions that accompany the charger for details on charger use Trickle charging only takes place for batteries stored in a monitor connected to AC mains You can leave batteries in the Battery Charger and Conditioner M8043A as no trickle charging takes place Accessing the Battery Status Window The Battery Status window shows additional information about the batt...

Страница 149: ...Battery Handling Maintenance and Good Practices Maintaining the Instrument 133 When the Time to Full shows 0 the battery charge LED on the front panel remains lit until battery calibration is complete ...

Страница 150: ...y LED blinks at a rate of 1 5 red flashes per second A severe INOP switches the main alarms suspended status back to on and generates an INOP alarm If this alarm is silenced it recurs after 3 minutes The previous two INOPs are cleared if AC power is connected while the following INOP stays no matter whether AC power is connected BATTERY MALFUNCTION This is a severe INOP which indicates a malfuncti...

Страница 151: ...charge cycles Conditioning a battery refers to the complete discharge of a charged battery by allowing it to die out while in non critical use The empty battery may then be recharged and put back into use The rechargeable battery used in the Instrument must be conditioned every 50 cycles to prepare it for further charge discharge cycles Question 2 According to this chapter which of the following s...

Страница 152: ...aintaining the Instrument 136 Maintaining the Instrument Answers to the Tutorial for Maintaining the Instrument 1 c See pages 119 and 127 for more details 2 d See page 122 for more details 3 a See pages 125 and 126 for more details ...

Страница 153: ...Instrument Concepts Functionality Assurance This refers to the combined Performance Assurance Test and Functionality Testing Procedures to be found in this chapter These tests verify correct Instrument function in general terms Preventive Maintenance Preventive Maintenance refers specifically to the service calibration tests required to make sure the Instrument measurement results are accurate Whe...

Страница 154: ...s chapter Where P Pass F Fail and X x are the measured values as defined in the tests described in this chapter Test What to record Visual V P or V F Power On PO P or PO F P NIBP PN P X1 X2 X3 X4 or PN F X1 X2 X3 X4 P CO2 PCO2 P X1 X2 X3 X4 X5 X6 X7 X8 or PCO2 F X1 X2 X3 X4 X5 X6 X7 X8 Safety S 1 P x1 x2 or S 1 F x1 x2 S 2 P x1 or S 2 F x1 S 3 P x1 or S 3 F x1 ...

Страница 155: ...en the monitor has been installed an assembly has been replaced to verify the operation any time the unit is suspect Preventive Maintenance Tests NBP Calibration Sidestream CO2 Calibration Once every two Once a year or following any Instrument repairs or the replacement of any Instrument parts applies to units with serial number prefix DE020xxxxx only Performance and Safety Tests Temperature Accur...

Страница 156: ...Safety 1 test blocks Installation of M3 M4 with a recorder connected to the serial interface Perform Visual Power On and Safety 1 test blocks Perform test print on recorder Repairs where the power supply in the M3 M4 is replaced Perform Power On and Safety 2 Test blocks Repairs where the monitor has been dropped Perform Power On and Safety 2 and 3 Test blocks All other M3046A repairs Hardware or S...

Страница 157: ...racy and Performance Procedures Temperature Accuracy 155 ECG Resp Performance 155 Invasive Pressure Performance Test 156 SpO2 Performance Test 157 Mainstream CO2 Performance Test 157 Nurse Call Relay Performance Test 157 ECG Sync Performance Test 158 Patient Safety Checks S 1 Part 1 System Enclosure Leakage Current NC normal condition 160 S 1 Part 2 System Enclosure Leakage current Single Fault op...

Страница 158: ... that the back of the extension is visible Passwords The following table show the passwords you need to enter to access the different modes Go to Operating Modes and use the TouchStrip to specify the password Press OK Visual Test Inspect the system for obvious signs of damage Also check external leads and accessories What to record on the service record V P or V F where P pass i e the system has n...

Страница 159: ...ssing the ON OFF button Step 2 After approximately one minute the Monitor should complete its boot up process Observe that there is a screen display and that no error codes are displayed Step 3 If any error codes are displayed or the screen remains black refer to Troubleshooting the Instrument The Instrument has a Self Test routine which generates and displays test waveforms and corresponding nume...

Страница 160: ...he Monitor and the Monitor is switched on For SpO2 mainstream CO2 Pressure or Temperature an appropriate transducer must be plugged into the Measurement Server or for mainstream CO2 the Measurement Server Extension in order to get a test reading For sidestream CO2 appropriate tubing must be plugged into the Measurement Server Extension in order to get a test reading Step 2 Press the Setup button S...

Страница 161: ...EDI NEO Simulated squarewave and numerics SpO2 Pletha SpO2 Pleth 777 9 100 Wave on display Numeric only Simulated wave Temperaturea 40 C 104 F Numeric only Pleth Pulse Rate 60 bpm in ADULT NEO mode NBP 120 80 90 ADULT 100 60 80 PEDI 80 50 60 NEO Simulated squarewave and numerics Sidestream CO2 b b Appropriate tubing must be plugged into the Server Extension in order to get the test reading When in...

Страница 162: ...o years and when the Instrument is repaired or when Instrument parts are replaced Connect the equipment as shown in the following Tools required Reference manometer includes hand pump and valve accuracy 0 2 of reading Expansion chamber volume 250 ml 10 Or alternatively an NBP cuff can be used When using a cuff take care not to move it or touch it during the procedure as this can cause the calibrat...

Страница 163: ...lowing prompt messages appears during this step check whether there is leakage in the setup NBP unable to calibrate cannot adjust pressure NBP unable to calibrate unstable signal Step 12 Press Confirm NOTE If the INOP NBP Equipment Malfunction message occurs in Monitoring Mode go back to Service Mode and repeat the calibration procedure NBP Leakage Test This test checks the integrity of the system...

Страница 164: ... Document the value displayed by the M3046A x4 Sidestream CO2 Performance Test NOTE 1 Allow 5 seconds between individual service procedures in order to ensure stable equipment conditions 2 When certain Monitor procedures are running for example AutoZero or purging service procedures are not possible and trying to start them will result in a message Service Operation Failed in the status line on th...

Страница 165: ...tion for the sidestream Extension consists of the following steps Barometric Pressure Check and Calibration if required Leakage Check Pump Check Flow Check and Calibration if required Noise Check CO2 Cal Check and Calibration if required CO2 Cal Verification using 2nd cal gas NOTE All steps should be performed in the same session Barometric Pressure Check and Calibration Check the barometric press...

Страница 166: ...ervice Mode and highlight CO2 Step 2 Connect a FilterLine to the sidestream CO2 input to start the pump running Step 3 Check the ambient and cell pressure shown in the status line on the screen The cell pressure should be approximately 20 mmHg lower than ambient pressure Step 4 Connect the flowmeter outlet to the FilterLine inlet using a flexible connecting tube Step 5 Block the M3015A outlet usin...

Страница 167: ...fingertip and observe the cell pressure on the M3046A display The cell pressure x4 should be more than 120 mmHg below the ambient pressure shown If the pressure difference is less than 120 mmHg the pump is not strong enough and should be replaced irrespective of the Pump OpTime Flow Rate Check and Calibration Check the flow rate in the sidestream CO2 Extension as follows Step 1 Connect the flowmet...

Страница 168: ...g with an ambient pressure of 736 mmHg Step 3 Open the valve on the flow regulator to allow 5 CO2 gas to flow into the Extension Allow the value to stabilize Step 4 Check that the value on the instrument measurement value on the Main Screen x7 matches the calculated mmHg value 2 6 mmHg If the value is outside the tolerance calibrate as described in Step 9 onwards Step 5 Disconnect the 5 calibratio...

Страница 169: ...in tolerances the M3015A must be exchanged Checking and Resetting Time Counters The time counters on the sidestream CO2 Extension must be checked before calibrating the Instrument As well when parts are replaced the appropriate counters must be reset to zero applies to units with serial number prefix DE020xxxxx only To check the counters do the following Step 1 Select the counter you want to check...

Страница 170: ...ailed X1 xx two digits X2 x x X3 x x X4 xxx X5 x x X6 x x X7 x x X8 x x Leakage Check parts 1 and 2 X2 Value of part 1 leakage check on Flowmeter X2 4 0 ml min X3 Value of part 2 leakage check on Flowmeter X3 4 0 ml min Pump Check X4 difference in pressure between cell pressure and ambient pressure displayed on the M3046A during occlu sion X4 120 mmHg Flow Check X5 difference between measured valu...

Страница 171: ...he temperature connector on the Measurement Server or Measurement Server Extension Step 2 Configure the patient simulator to 40 oC or alternatively 100 oF Step 3 The value should be 40 oC 0 2 oC or 100 oF 0 4 oF If the displayed value is not within the tolerances given above refer to Troubleshooting the Instrument ECG Resp Performance This test checks the performance of the ECG and respiration par...

Страница 172: ...ry two years Tools required Patient Simulator Step 1 Connect the Patient Simulator to the Pressure connector on the Measurement Server or the Measurement Server Extension Step 2 Set Patient Simulator to 0 pressure Step 3 Make a zero calibration Step 4 Configure the Patient Simulator as follows P static 200 mmHg Step 5 Wait for the display Step 6 The value should be 200 mmHg 5 mmHg If the value is ...

Страница 173: ...tructions for Use There is no requirement for additional tests to be performed by the Service Engineer or Biomedical Engineer Nurse Call Relay Performance Test This test checks the operation of the Nurse Call Relay The Nurse Call Relay test is recommended for customer sites where the nurse call is in use It should be performed once every two years and when the Instrument is repaired or when Instru...

Страница 174: ...ired or when Instrument parts are replaced Tools required Defibrillator with ECG Sync and Marker Output Patient simulator Step 1 Connect the patient simulator to the ECG connector on the Measurement server and the defibrillator to the ECG Sync Output on the M3046A Step 2 Set the Patient simulator to the following configuration HR 100 bpm ECG sinus rhythm Step 3 Switch the defibrillator to simulati...

Страница 175: ...rding to UL2601 1 Additional tests may be required according to local regulations Normally a Safety Analyzer is used to perform these procedures Popular testers include the DEMPSEY 232D or for use in Europe testers like the Rigel Metron or Gerb Follow the instructions of the Instrument manufacturer CAUTION If the Dempsey is used for an extended length of time it could be damaged by the high amp cu...

Страница 176: ...em Enclosure Leakage current Single Fault open earth Not present in Class 2 Can be multiple different connections to different equipment at same time PE Instrument under test Applied part S4 N L S1 S2 L N Signal parts in and or output MD Medical electrical system Signal parts in and or output Other Instrument C a n b e m u ltip le diffe re nt co nnec tio ns to different eq uipm e nt at sa m e tim ...

Страница 177: ...or S 1 F x1 x2 where P Pass F Fail and x1 x2 are the values defined in the two tests described above S 2 Protective Earth Continuity Expected test results With mains cable maximum impedance x 100 mOhms IEC 60601 1 and UL2601 1 Measures impedance of Protective Earth PE terminal to all exposed metal parts of Instrument under Test IUT which are for safety reasons connected to the Protective Earth PE ...

Страница 178: ...nt leakage current from applied Part to earth caused by external main voltage on applied Part with switch S5 open and closed Each polarity combination possible is tested using S2 and S6 This test is applicable for every measurement input ECG RESP Invasive Pressure CO2 SpO2 Temperature Reporting safety test S 3 in the Service record S 3 P x or S 3 F x where P Pass F Fail and x is the value defined ...

Страница 179: ... in product development It is not necessary for hospital personnel to report results c Preventive Maintenance can be defined as service calibration testing required to make sure the Instrument measurement results are accurate These specific tests are required for the NBP parameter and for the sidestream CO2 parameter d All of the above answers are correct Question 2 In the case of all testing proc...

Страница 180: ...torial for Testing the Instrument 164 Testing the Instrument Answers to the Tutorial for Testing the Instrument 1 d See page 137 for more details 2 a See entire chapter for more details 3 c See page 157 for more details ...

Страница 181: ...the obvious hints and tips Part 2 Isolating and Solving Instrument Problems This part provides the following Tables that list Instrument symptoms as well as causes and remedies for Instrument faults Techniques for troubleshooting the Instrument before you disassemble it Checks you can make to isolate problems down to a specific replaceable part for example the Power Supply You can often determine ...

Страница 182: ...er and if present the Measurement Server Extension inserted correctly Checks Before Opening the Instrument You can isolate many problems by observing indicators on the Instrument before it is necessary to open the Instrument Checks with the Instrument switched Off AC connected without battery AC Power LED is on green AC connected with battery AC Power LED is on green Battery LED is on green if ful...

Страница 183: ...with an audible tone Shortly after the LEDs switch off the display backlight switches on and the M3 logo dis plays Checks with the Instrument Switched On AC connected with Battery The LEDs behave the same as described in the Checks with the Instrument Switched on AC Connected without Battery above except that the Battery LED does the following The Battery LED is initially on then switches first to...

Страница 184: ...re displayed refer to error codes later in this chapter The brightness of the LCD display is adequate If not adjust it accordingly The green LED on the System Board is on see the following section LED Normal Operation Defect Condition On Off Standby Green The Monitor is on Off The Monitor is in standby mode The LED remains off Suspect 1 Monitor Bezel cable connec tion 2 System Board Go to the trou...

Страница 185: ...pply and associated cabling Yel low Main CPU status LED When on and blinking this LED indicates normal CPU operation This LED starts to blink 2 times per second after the red Error LED switches off then slows to blink 1 time per second after the red Error LED switches off when the system boot has finished approximately 2 seconds If permanently on or off indicates a hung CPU Reload software Exchang...

Страница 186: ...age Voltages dangerous to life are present in the Instrument when it is connected to the mains power supply Do not attempt any adjustment or make any contact with the monitor while AC power is connected The green LED on the Display Adapter Board can be seen if the chassis is removed from the plastic housing It is located on the left side speaker side looking through the lower hole of the metal dis...

Страница 187: ... the wireless assembly includes a CPU board and a radio card If a hardware failure is found during operation the red LED will blink repeatedly The Status LEDs are yellow directly after Power On during initialization During normal operation they are green If a problem occurs during operation the LEDs blink red in a repeating pattern Radio LED The radio LED blinks yellow when the wireless assembly i...

Страница 188: ...Go to the troubleshooting section Yel low Main CPU status LED When on and blinking this LED indicates normal CPU operation Starts to blink 2 times per second after the red Error LED switches off then slows down to blink 1 time per second after the red Error LED switches off when the system boot has finished approximately 2 sec onds If permanently on or off indicates a hung CPU Reload software Exch...

Страница 189: ...operates prop erly Make sure that the battery is adequately charged press charge indicator on battery at least one green LED should switch on Insert the battery into the Monitor AC power still connected the Battery LED should switch on green yellow or red If the Battery LED remains off or is red replace the battery with a known good one If the Battery LED still remains off either the LED is defect...

Страница 190: ...after approximately 2 seconds Red LED The red LED switches on for about 1 second after a power on RESET of the System Board then switches off If the boot process is not successful for any reason check for obvious problems Part 1 Troubleshooting Checklists What To do If the Display Remains Dark If the three LEDs on the System Board indicate that the first boot phase has finished green on red off ye...

Страница 191: ...ied to the Instrument Failure to adhere to this warning could cause serious injury or death Troubleshooting the ECG OUT If no marker pulse is displayed on the Monitor assuming the defibrillator and defibrillator cable are in working order check the following Disconnect the Server and Defib cable Switch the Monitor off then on again Observe the red LED in the ECG OUT section The Instrument Self Tes...

Страница 192: ... Marker In and ECG Wave Defib Monitor Measurement Server DEFIB RS485 LAN Link Bar Connector Isolat Conn MEASUREMENT Link Bar ECG_OUT CPU Communication System and Video Display Assembly Marker In Data Path Monitor System Board Server ECG Wave Data Path Link Bar Defib Cable Marker merged to ECG Wave here ECG Wave Marker Application Software SERVER ...

Страница 193: ...tion of Monitor Measurement Server and cable Switch off the monitor and contact your biomedical department Use standard equipment link bar cable Server or Extension The link bar connects the Server and where present the Extension to the Moni tor Update Software BAD SERVER LINK plus Measurement Server Revision not supported status message in red An M3000A Measurement Server with revision A software...

Страница 194: ... Board CHARGER MAL FUNCT Either the battery or the charging hardware within the Instrument is defective 1 Change battery 2 Change the System Board CHECK INPUT DEVICE The Monitor has detected 5 minutes or more of constant user interface operation or the user interface is faulty 1 Make sure nothing is pressing on the keys or the TouchStrip 2 Check that the Moni tor Bezel is mounted securely and for ...

Страница 195: ...ystem Board SPEAKER MAL FUNCTION Defective speaker or cables 1 Check cable connec tions 2 Replace speaker 3 Replace the System Board UNSUPPORTED LAN An unsupported device has been connected to LAN port of the Moni tor Disconnect the unsup ported device CO2 EQUIP MALF M3015A and M3016A Malfunction in the CO2 hardware or the M1460A transducer hardware Verify failure and replace the Extension with a ...

Страница 196: ...6A The CAL 1 calibration cycle is complete the CAL 2 calibration cycle must be started Place sensor on other cal cell and start the CAL 2 calibration cycle CO2 CAL RUN NING M3016A CO2 calibration is run ning Wait for CO2 calibra tion to complete CO2 CAL MODE M3016A Cal mode is set but the calibration has not been initiated Initiate the calibration or switch off cal mode CO2 CHANGE SCALE ETCO2 wave...

Страница 197: ... cannot be taken or 2 The outlet is occluded at Power On 3 This INOP also appears after the Pump Op Time is reset 1 and 2 Check the Fil terLine and exhaust tube then disconnect and reconnect the Fil terLine If the INOP is still displayed use a new FilterLine 3 If Pump Op Time has been reset perform a flow check and store the flow in Service Mode select Store Flow CO2 OVERRANGE M3015A The CO2 value...

Страница 198: ... removed by purging the Measurement Extension will go into Standby Mode and a CO2 OCCLUSION INOP will be displayed Check for an occlusion and remove If necessary replace the FilterLine NO CENTRAL MONITORING Central monitoring has been interrupted With a wireless net work Check that the Monitor has not been moved out of range of an access point and that no microwave oven or other non monitoring wir...

Страница 199: ... already present remove it and check the charge level press the button on the battery Refit the battery making sure to push it completely into position Some or all numerics or waves are not displayed Parameters are switched off No transducers are con nected Defective transducer Unsupported server Server defective Switch parameters on Connect the required transducers Replace the suspect trans ducer...

Страница 200: ... lamp on the front panel is on or off Check System Board LEDs see Troubleshoot ing the System Board LEDs Check the ribbon cable connection from the Sys tem Board to the Monitor Bezel Replace backlight tubes Replace display Monitor screen is blank but the Instrument appears to boot as it should alarm LEDs switch on off successful test sound and yel low LED on the System Board blinks Display cable m...

Страница 201: ...f a battery is in remove it and put in new data Replace the connector board and put in some patient data If this does not correct the fault replace the System Board Printing is not possible Defective printer Infrared connection is not correctly aligned Defective sender Try printing with another Monitor Line up the sender and receiver Try another Monitor Bezel and replace if necessary Check cable C...

Страница 202: ...et the time and date Replace the System Board Cannot repro gram the time and date correctly Defective System Board Replace the System Board Compatibility related problems Some measure ments are not supported by the Monitor prompt message A measurement extension M3015A or M3016A is connected to an M3046A M3 monitor No measurements CO2 2nd Press Temp from the Measurement Extension are available Use ...

Страница 203: ...Press Temp from the Measurement Extension are available Or A Measurement Extension M3015A or M3016A is connected to a standard M3001A Measurement Server noninvasive meas urements only and an M3046A Monitor with software Rev E No meas urements CO2 2nd Press Temp from the Measure ment Extension are availa ble Use extensions only with Extended Measurement Servers C06 for M3000A C06 and C18 for M3001A...

Страница 204: ...l LEDs earlier in this chapter Display and LCD Adapter Board related problems Monitor screen is blank Instrument appears to boot as it should alarm LEDs switch on off successful test sound and yellow LED on the System Board blinks Display and other cables not seated correctly See Troubleshooting the System Board LEDs ear lier in this chapter Refit cables correctly If this does not work replace the...

Страница 205: ...TouchStrip dis colored cracked or surface is deformed Cleaning agents abrasive material or strong sol vents were used to clean the Instrument Replace the Monitor Bezel and clean the Instru ment in the future with only isopropyl alcohol Wipe cleaning agents off surfaces immediately Lack of user control of dis play Cable from the System Board to the Monitor Bezel not connected prop erly Defective To...

Страница 206: ... software incom patible prompt message Ask Biomed department to check software revi sions Central can not identify this bed prompt message The location of this bed is not clear Check the bed assign ments at the Information Center Measurement Server related problems No waves Measurement Server defective LEDs on display unit blink but Server LED does not Check green LED on Server Exchange the Server...

Страница 207: ...s appear unstable or unreliable Low light transmission of SpO2 sensor LEDs to the sensor photo diode Patient has low peripheral perfusion Check Transmission and Perfusion Index as fol lows 1 Fit SpO2 sensor on your finger 2 Enter Config Mode 3 Select SpO2 4 Switch SpO2 On Lightly press the Touch strip until text appears at the bottom of the screen Perfusion Index xx x Transm red yyyy Transm infra ...

Страница 208: ...t extension is sup ported by the system configuration hardware and software Check that no UPDATE FW mes sage is displayed 2 Check CO2 Data Acquisition hardware revision If the Measure ment Server Extension is incompatible with wire less network either use this Measurement Server Extension on a monitor without wireless network or upgrade the Measure ment Server Extension contact your service rep re...

Страница 209: ...n calstick and calibrate if necessary No wave dis played and no channel present 1 Measurement Extension is used with a monitor and or Measurement Server with Release A software A xx xx 2 CO2 measurement switched off 1 Check if the Measure ment Extension is sup ported by the system configuration hardware and software 2 Check that CO2 meas urement is switched on Audio Speaker related problems Buzzin...

Страница 210: ... 2 Isolating and Solving Instrument Problems 194 Troubleshooting the Instrument Buzzing noise in Instrument Defective Audio Circuit Defective Power Supply Replace the System Board Replace the Power Sup ply ...

Страница 211: ...order hardware fault detected Replace the Recorder LocalRecorder check cable prompt message Communication error Intermittent connection to Recorder or faulty cable Check cable connection is secure between the Monitor and the Recorder If connection is sound but the problem persists replace the cable Frame error in RS232 protocol due to a connection problem or effects of interference from external e...

Страница 212: ...ts replace the Recorder The printhead is too cold 0 C For instance the Recorder may have been tranferred from storage where the temperature was below 0 C The ambient temperature in the room where the Recorder is used should be above 0 C Allow the Recorder to warm up for a few minutes then try again If the problem persists replace the Recorder LocalRecorder disconnected status message The Recorder ...

Страница 213: ...ays fatal error codes In the Service Mode the Status Log displays both fatal and non fatal error codes with a blank line separating the two Errors that occur during start up or regular monitoring are logged in the Status Log The Status Log can be printed and cleared If fatal errors occur repeatedly send a printout of the Status Log Service Mode to Philips There are two categories of error Fatal Er...

Страница 214: ...C or N H and C These denote fatal errors which have caused a Hot start like switching the Instrument off on or a Cold start like per forming a System Test N This denotes a non fatal error Device ID Code This number refers to a software module and is used to indicate the location of a fault n This number is the internal software module number Error Code This number describes the type of fault ident...

Страница 215: ...20005 Software Condition N 17205 20009 Software Condition N 17205 20012 Software Condition N 17205 20058 Software Condition N 17205 20500 Software Condition N 17205 20553 Software Condition N 17205 20601 Software Condition N 17216 20050 Software Condition N 17250 20642 Software Condition N 17315 22043 Software Condition N 17316 12000 Software Condition N 17316 12100 Software Condition H 16400 610 ...

Страница 216: ...ilure replace Server H 16400 133 Hardware Failure replace Server H 16400 591 Hardware Failure replace Server H 16400 592 Hardware Failure replace Server H 17203 20102 Hardware Failure replace Server H 17206 20015 Hardware Failure replace Server H 17300 20107 Software failure upgrade to revision B H 32753 28672 Hardware Failure replace Server Table 2 Error Codes for Release B Severity Device ID Err...

Страница 217: ...d C 32749 20002 Measurement Extension has been exchanged no action required M3046A Monitor N 16510 20337 Software Condition N 17205 20050 Software Condition N 17205 20058 Software Condition N 17205 20650 Software Condition N 17205 20664 Software Condition N 17206 20010 Software Condition N 17215 21480 Software Condition N 17215 21580 Software Condition N 17250 20502 Software Condition N 17250 2052...

Страница 218: ...ormation Required Action M3000A Measurement Server H 17300 20096 Software Condition H 17300 20307 Can occur during or after upgrades no action required M3046A Monitor N 16510 20337 Software Condition N 17205 20014 Software Condition N 17205 20505 Software Condition N 17205 20604 Software Condition N 17303 20000 Software Condition N 17303 20006 Software Condition N 17303 20703 Software Condition N ...

Страница 219: ...ondition H 17305 20008 Software Condition H 17305 30056 Software Condition H 17215 22002 Software Condition H 17215 22060 Software Condition H 32750 21400 Software Condition H 32750 21224 Software Condition Table 4 Error Codes for Release D Severity Device ID Error Code Information Required Action M3046A Monitor N 17312 20064 Software Condition N 17315 20016 Software Condition H 16400 611 Software...

Страница 220: ...igure the Wireless LAN assembly of the M3 M4 and the Site Survey Configura tion Tool for the same Domain and Security ID Configure using the Wireless Configura tion Tool available on the Philips Information Center CD ROM 3 Check that the RangeLAN2 Site Survey and Configuration tool is running if not double click on the RangeLAN2 Configuration Icon to start it The main window appears as shown below...

Страница 221: ... MAC address of the radio board 0020a6338fdb with a Received Signal Strength of 90 The number to the right of the bar graph is the indicator Acceptance Criteria The RSSI should be greater than 80 If it is less than 80 check the antenna cable connection to the Radio board inside the M3046A to see that it is mated properly When the test is completed click on Done in the Site Survey window then on Ex...

Страница 222: ...t to Monitoring Mode To execute a coldstart test do the following Step 1 Switch the Monitor into Service Mode Step 2 Press the Setup key Step 3 Select Status Log Step 4 Clear the Status Log and Confirm for both Monitor and Server Step 5 Press Exit Step 6 Select Hardware Test Step 7 Select Coldstart Step 8 Press Confirm The Monitor and Server if connected perform a reboot The coldstart reboot is th...

Страница 223: ...ver the whole screen once Be aware that numerics are updated also during a running display test This should not be considered a fault Failure The test is failed if the display has not reverted to the pre test state once the test has passed over the whole screen once If the test fails repeat the test to make sure that it is not a single sporadic fault If it fails again exchange the System Board The...

Страница 224: ...s If the flow cannot be adjusted to the desired value If the Pump makes considerably more noise than new Pumps After replacing the pump perform a CO2 calibration check as described in CO2 Gas Measurement Calibration Check on page 152 When replacing the pump in the M3015A Measurement Server Extension sidestream CO2 you should also replace the CO2 scrubber at the same time Recorder Status Check You ...

Страница 225: ...on is used to switch between 50 Hz and 60 Hz for the ECG filtering frequency CAUTION Make sure that the Line Frequency is set correctly for the country option ordered The factory default is 60 Hz for all units regardless of country option How To Access the Monitor and Server Revision Screen 1 Press Setup 2 Move the highlight to Revision 3 Depending on which screen you want to display toggle betwee...

Страница 226: ...he Monitor should be investigated technically however as soon as possible ECG EQUIP MALF Measurement defective Exchange the Measurement Server The Measurement Server should be investigated technically as soon as possible NBP EQUIP MALF Measurement defective Exchange the Measurement Server The Measurement Server should be investigated technically as soon as possible Pressure Zero Check Cal Pressure...

Страница 227: ...AD SERVER LINK You cannot use this combination of Monitor Measurement Server and cable Switch off the monitor and contact your biomedical department BAD SERVER LINK plus Measurement Server Revision not supported status message in red An M3000A Measurement Server with revision A software is connected to an M3046A Monitor with a software revision of B or later This combination does not allow monitor...

Страница 228: ...nts for example CO2 2nd Press Temp from the Measurement Extension are available Exchange the Measurement Server for an M3001A with option C06 or C18 Some measurements in MMS Extension are not supported prompt message The M3046A Monitor does not support some of the measurements in a measurement extension Consider using a monitor that supports these measurements Measurement Server not supported stat...

Страница 229: ...ers Defective transducer Replace the suspect transducer Measurement Server defective Exchange the Measurement Server The Measurement Server should be investigated technically as soon as possible Monitor screen dim Brightness controls not properly adjusted Adjust brightness controls The screen may not be as bright when the Monitor is operating from the battery Monitor screen blank Power is not conn...

Страница 230: ...nect the standard link bar before continuing Error codes 20210 20307 20002 20102 206nn 20nnn Several error codes will occur as a result of an upgrade and do not require any action In order to assess which codes result from an upgrade you should always check and clear the status log before starting an upgrade When the upgrade is complete recheck the status log and clear before normal operation is r...

Страница 231: ...g for the page This can happen if the JetEye is switched on but the printer is switched off when you start printing Make sure both the printer and if necessary the JetEye are switched on then print again A REMOTE PRINTER NOT AVAILABLE mes sage appears on the screen Check that the Monitor is connected to the Philips LAN or the M3 Print Servera network If it is not connect it see Connecting a Remote...

Страница 232: ...k if there are any paper scraps caught under the print head Paper moves but Recorder does not print Paper is incorrectly loaded Local Recorder discon nected message appears on the screen The Recorder is disconnected from the Monitor or switched off Ensure that the Recorder cable is securely connected to the sockets on the Monitor and on the Recorder and that the Recorder is switched on Local Recor...

Страница 233: ...t functions display information only either operating time or number of cycles for parts or features of the Instrument and cannot be modified a The Self Test Cycles OpTimeMon OpTimeMeasS and NBP Cycles are selections which display this type of information and cannot be modified b The Coldstart Test Display Test Reset PumpOpTime Reset IRSourceTime and Backlight Test display this type of information...

Страница 234: ...strument 218 Troubleshooting the Instrument Answers to the Tutorial for Troubleshooting the Instrument 1 c See page 197 for more details 2 a See page 208 for more details 3 e See pages 5 2 to page 175 for more details 4 c See pages 197 for more details ...

Страница 235: ...ment procedures for the M3015A Measurement Server Extension are easy to complete when you follow the step by step procedures in this chapter The main replaceable parts of the M8043A Battery Charger and Conditioner are The Air Filters left hand and right hand The Cooling Fans left hand and right hand Follow the step by step removal and replacement procedures starting on page 239 The M3000A and M300...

Страница 236: ...Failure to adhere to this warning could cause serious injury or death Residual High Voltage There may be hazardous voltages stored in capacitors on the Power Supply board even when the Instrument is not connected to the mains Power Supply High voltages are present on the Power Supply board for up to 20 seconds after disconnecting the mains power cord so always disconnect the Power Supply cord and ...

Страница 237: ...r You must remove the Power Supply to access the fuse To remove the Power Supply do the following Step 1 Remove the Measurement Server and where present the Measurement Server Extension Step 2 Remove the battery Step 3 Switch on the Instrument and disconnect the AC power cord from both the Instrument and the mains power source after 20 seconds the capacitors on the Power Supply board are drained S...

Страница 238: ...attery Step 3 Switch the Instrument on and disconnect the AC power cord from the Instrument and the mains power source after 20 seconds the capacitors on the Power Supply board are drained Step 4 Push a thin bladed screwdriver into the hole next to the cover Step 5 Release the cover Step 6 Prise open the clip on the Server to Monitor link bar and slide it off On newer monitors just turn the catch ...

Страница 239: ... Monitor up again and gently lift the Monitor Bezel up and out Gently let it rest in front of the Monitor Be careful It still has a cable attached Step 9 Remove the white plastic clamp securing the ribbon cable connections To remove push the latch at the top left corner to the left Step 10 Use a small screwdriver to remove the Monitor Bezel assembly cable NOTE Do not touch the LCD display ...

Страница 240: ...sis on the Monitor Bezel For Monitors with a Wireless LAN Assembly To avoid antenna damage slide the chassis very carefully out from the front of the housing disconnect the antenna cable and lay the chassis on the Monitor Bezel Removing the System Board To remove the System Board do the following Step 1 Remove the Power Supply as described previously in this chapter Step 2 Remove the chassis as de...

Страница 241: ... to disconnect the Display Assembly cable Step 5 Lay the chassis face down on a clean scratch free surface Step 6 Disconnect the loudspeaker cable 1 For Monitors with Wireless LAN assembly Disconnect the ventilator wires 2 from the system board Unclip the fan assembly from the metal sheet 3 and pull forwards to remove Step 7 Remove the 4 System Board screws 4 1 4 2 3 4 2 4 4 3 1 ...

Страница 242: ... connector then remove the assembly from the system board Removing the LCD Assembly and Backlight Tubes This procedure describes how to remove the LCD Assembly which includes the LCD screen and rubber anti vibration cushion Once the assembly has been removed the backlight tubes can be removed and replaced Although you need to remove the chassis you do not need to remove the System Board or the Pow...

Страница 243: ...ouch the front of the display assembly the screen or the backlight tubes with your fingers Doing so can reduce the lifetime of the parts Step 4 Then push from the sides Step 5 Lay the LCD display on a flat scratch free surface Step 6 Unplug the backlight tube cables ...

Страница 244: ... Step 1 Remove the chassis as described previously in this chapter Step 2 Remove the Power Supply as described previously in this chapter Step 3 The Connector Board VGA is fixed with a screw and can no longer be taken out by removing the Connector Board cover and pulling the Connector Board out The Plastic Housing must be completely removed to allow the Connector Board to be unscrewed from the sid...

Страница 245: ...r do the following Step 1 Remove the chassis as described previously in this chapter Step 2 Remove the Power Supply Step 3 Remove the Connector Board Step 4 Disconnect the speaker cable running to the System Board Step 5 Rotate the two plastic holders retaining the speaker Step 6 Remove the speaker ...

Страница 246: ...tain ing clips on the chassis Make sure the bottom edge of the System Board sits correctly in the clips to give a good electrical contact Refitting the LCD Assembly and Backlight Tubes To refit the LCD Assembly and backlight tubes complete the steps for removing the LCD Assembly and backlight tubes in reverse NOTE When you push the Display Assembly back into the chassis make sure that the 6 clips ...

Страница 247: ...gh the middle hole 1 below in the right hand column of three holes as viewed from the back side of the sheet metal Connect the free end of the antenna cable to the connector on the corner of the wireless LAN board 2 above Refitting the Battery To refit the battery do the following Step 1 Slide the battery compartment door toward the rear of the Monitor and open it down Step 2 While holding the bat...

Страница 248: ...ver A pair of large tweezers In addition for removing the Pump you will need a large bladed screwdriver WARNING There is high voltage inside the Instrument 800V Do not connect the Measurement Server Extension to a Monitor while the Extension housing is open As well parts inside the Instrument may be contaminated with bacteria Protect yourself from possible infection by wearing examination gloves d...

Страница 249: ...h to be removed by hand Step 3 Remove the two pins and set them aside for refitting NOTE Do not lose these long mounting pins since the Extension will not function unless they are in place Step 4 Using your hands gently pry the bottom cover away from the Extension at the link bar end first The bottom cover is press latched at the link bar end Remove it gently making sure not to bang or touch the i...

Страница 250: ... M3015A Measurement Server Extension 234 Repairing the Instrument The following illustration shows the location of the replaceable parts in the M3015A Measurement Server Extension Pump CO2 Scrubber Brackets holding the CO2 Scrubber ...

Страница 251: ... Step 2 Being careful not to touch anything else in the Extension use tweezers to pull the body of the CO2 Scrubber out of the bracket Step 3 Holding the body of the CO2 Scrubber with your fingers carefully disconnect the Extension intake tube from the scrubber end and remove the CO2 Scrubber from the Extension Step 4 Dispose of the CO2 Scrubber according to local legal requirements for low volume...

Страница 252: ...isconnect the flow tubing attached to the Extension from the Pump NOTE Be sure to note which tube attaches to the inlet and which tube attaches to the outlet Step 4 Gently disconnect the power lead which attaches the Pump to the Extension Step 5 Remove the Pump NOTE After replacing the Pump reset the value displayed on the Reset PumpOpTime selection to zero Service Mode CO2 Setup When the PumpOpTi...

Страница 253: ...teria Protect yourself from possible infection by wearing examination gloves during these procedures Refitting the CO2 Scrubber WARNING The CO2 Scrubber contains lithium hydroxide monohydrate This is a strong base Do not open or damage the CO2 Scrubber If you come into contact with the CO2 Scrubber material flush the area immediately with water and consult a doctor To refit the CO2 Scrubber do the...

Страница 254: ...ension if the pump touches the PC board Step 4 Replace the top part of the bracket and screw firmly into position NOTE After replacing the Pump reset the value displayed on the Reset PumpOpTime selection to zero Service Mode CO2 Setup When the PumpOpTime has been reset an INOP will be generated CO2 OCCLUSION To clear this INOP you must perform a flow check and store the flow in Service Mode select...

Страница 255: ... 1 Disconnect the Battery Charger and Conditioner from AC mains power Remove any batteries from the charger Step 2 Place the device upside down and remove the five cross head screws holding the chassis baseplate Step 3 Lift the front edge of the chassis baseplate then carefully pull it forwards slightly Note that some connected cables prevent you from completely removing the baseplate Screws ...

Страница 256: ...see Opening the Housing on page 239 Step 2 Carefully manouvre the baseplate to gain internal access to the left hand cooling fan Step 3 Remove the four screws holding the cooling fan and grill Step 4 Disconnect the two pin plug A connecting the cooling fan cable to the printed circuit board Free the cable from the cable guide B Left hand Cooling Fan ...

Страница 257: ...hen reconnecting the two pin plug ensure that the red wire faces the left side of the housing Step 6 Remove the four screws holding the right hand cooling fan and grill Step 7 Disconnect the two pin plug D connecting the cooling fan cable to the printed circuit board Free the cable from the cable guide C A B Right hand Cooling Fan C D Heat Sink ...

Страница 258: ...circuit board This allows better access to the inside of the charger Step 3 Remove the two screws holding the air filter Step 4 Remove the air filter cover Step 5 Remove the filter mat and clean the dust out or exchange the filter mat if required Step 6 Insert the mat and re fit the cover and screws Reassembling the Battery Charger and Conditioner Reassembly is the reverse of the disassembly proce...

Страница 259: ... the warnings cautions and safety procedures in this chapter a Do not perform any disassembly procedures other than Server and Extension removal with power applied to the Instrument b High voltages are present on the Power Supply board for up to 20 seconds after disconnecting the mains power cord c Never insert a battery without the Power Supply board being present d All of the above are correct a...

Страница 260: ...Repairing the Instrument 244 Repairing the Instrument Answers to Tutorial for Repairing the Instrument 1 c See pages 222 231 and 245 for more details 2 d See pages 219 and 220 for more details 3 b See page 221 for more details ...

Страница 261: ... do not contain any serviceable parts If defective these pieces must be returned to the factory Compatibility Matrix Release A to Release B New hardware starts with serial date code DE850XXXX as well as with the serial date codes in the range DE71305620 to DE73105823 NEW OLD Main System Board M3046 66502 exchange M3046 68502 Connector Board VGA cover M3046 66522 M3046 44109 Monitor Power Supply M3...

Страница 262: ...ew Main System Board and a new Chassis and Housing are also required NOTE The new Connector Board VGA is fixed with a screw and can no longer be taken out by removing the Connector Board cover and pulling the Connector Board out The Plastic Housing must be completely removed see Repairing the Instrument to allow the Connector Board to be unscrewed from the side of the Monitor Old Power Supply An o...

Страница 263: ...rts from your Philips representative The Item numbers where applicable correspond to the illustration which follows Exchange parts are parts that have been returned to Philips and reconditioned for further use Parts offered as exchange parts are in excellent service order according to rigorous Philips standards but offer a considerable price advantage to the user ...

Страница 264: ...onitor Bezel IrDA Interface Danish 1 M3046 62242 453563333021 Monitor Bezel IrDA Interface Traditional Chinese 1 M3046 62243 453563333031 Monitor Bezel IrDA Interface Simplified Chinese 1 M3046 62244 453563333041 Monitor Bezel IrDA Interface Portuguese 1 M3046 62245 453563333051 Monitor Bezel IrDA Interface Greek 1 M3046 62247 453563333061 Monitor Bezel IrDA Interface Russian 1 M3046 62250 4535633...

Страница 265: ...tor Bezel Serial Recorder Interface Traditional Chinese M3046 62343 453563485061 Monitor Bezel Serial Recorder Interface Simplified Chinese M3046 62344 453563485071 Monitor Bezel Serial Recorder Interface Portuguese M3046 62345 453563485081 Monitor Bezel Serial Recorder Interface Greek M3046 62347 453563485091 Monitor Bezel Serial Recorder Interface Russian M3046 62349 453563485101 Monitor Bezel S...

Страница 266: ...046 60202 453563332721 Display assembly 4 2090 0577 N A Monitor Display Backlight tube for display ass M3046 60201 5 2090 0380 N A Monitor Display Backlight tube for display ass M3046 60202 5 M3046 66521 453563333151 Connector Board old 7 M3046 66522 453563333161 Connector Board VGA M3046 44107 453563332631 Connector Board cover old M3046 44109 453563332641 Connector Board cover VGA M3046 64105 45...

Страница 267: ...adapter plate for rollstand M3080A A10 M3046 44110 453563484821 Removable Frame for the serial connector M3046 64102 453563333091 Small Parts Kit See separate table on page 253 a Use the M3 M4 Support Tool to configure the software for a specific release language if necessary b Order only through the Sales channel stating the serial number Exchange Part Number 12 Digit Reordering Number Descriptio...

Страница 268: ...Monitor Parts 252 Replacement Parts Monitor Parts 5 4 3 7 2 1 6 ...

Страница 269: ... Fits all Measurement Servers and Server Extensions 8 4 M3046 41001 Foot Bumper 9 1 5041 8116 Rubber key 10 1 5041 8118 Hook 11 4 RFI Metal Clip 12 3 0515 0886 Screw 13 2 Pins for Battery Compartment Door 14 1 M3046 48102 IR window 15 2 White Pull Bezel Tabs 16 1 M3046 42305 White Plastic Cable Keeper 17 1 M3015 16 White Tab with Spring 18 1 M3046 44104 Grey plastic cover for upgrade port 19 8 9 1...

Страница 270: ...Option C12 Conventinal 12 Lead ECG with IntelliVue onlya English M3001 64101 453563462481 French M3001 64102 453563462491 German M3001 64103 453563462501 Dutch M3001 64104 453563462511 Spanish M3001 64105 453563462521 Italian M3001 64106 453563462381 Norwegian M3001 64107 453563462391 Swedish M3001 64108 453563462401 Finnish M3001 64109 453563462411 Japanese M3001 64101 453563462481 Danish M3001 6...

Страница 271: ...01 453563462661 Portuguese M3001 44114 453563462601 Greek M3001 44115 453563462611 Turkish M3001 44101 453563462661 Russian M3001 44117 453563462621 Hungarian M3001 44101 453563462661 Czech M3001 44119 453563462631 Polish M3001 44120 453563462641 Slovak M3001 44101 453563462661 Korean M3001 44101 453563462661 a All versions of the M3001A Measurement Server offer EASI lead placement using an EASI 5...

Страница 272: ...ed Chinese M3001 60501 453563460691 M3001 68501 453563460631 Portuguese M3001 60501 453563460691 M3001 68501 453563460631 Greek M3001 60501 453563460691 M3001 68501 453563460631 Turkish M3001 60501 453563460691 M3001 68501 453563460631 Russian M3001 60501 453563460691 M3001 68501 453563460631 Hungarian M3001 60501 453563460691 M3001 68501 453563460631 Czech M3001 60501 453563460691 M3001 68501 453...

Страница 273: ...0651 Dutch M3001 60701 453563460711 M3001 68701 453563460651 Spanish M3001 60701 453563460711 M3001 68701 453563460651 Italian M3001 60701 453563460711 M3001 68701 453563460651 Norwegian M3001 60701 453563460711 M3001 68701 453563460651 Swedish M3001 60701 453563460711 M3001 68701 453563460651 Finnish M3001 60701 453563460711 M3001 68701 453563460651 Japanese M3001 60701 453563460711 M3001 68701 4...

Страница 274: ...01 68801 453563460591 Portuguese M3001 60801 453563460721 M3001 68801 453563460591 Greek M3001 60801 453563460721 M3001 68801 453563460591 Turkish M3001 60801 453563460721 M3001 68801 453563460591 Russian M3001 60801 453563460721 M3001 68801 453563460591 Hungarian M3001 60801 453563460721 M3001 68801 453563460591 Czech M3001 60801 453563460721 M3001 68801 453563460591 Polish M3001 60801 4535634607...

Страница 275: ...ied Chinese M3001 60502 453563487021 M3001 68502 453563486951 Portuguese M3001 60502 453563487021 M3001 68502 453563486951 Greek M3001 60502 453563487021 M3001 68502 453563486951 Turkish M3001 60502 453563487021 M3001 68502 453563486951 Russian M3001 60502 453563487021 M3001 68502 453563486951 Hungarian M3001 60502 453563487021 M3001 68502 453563486951 Czech M3001 60502 453563487021 M3001 68502 45...

Страница 276: ...6971 Dutch M3001 60702 453563487041 M3001 68702 453563486971 Spanish M3001 60702 453563487041 M3001 68702 453563486971 Italian M3001 60702 453563487041 M3001 68702 453563486971 Norwegian M3001 60702 453563487041 M3001 68702 453563486971 Swedish M3001 60702 453563487041 M3001 68702 453563486971 Finnish M3001 60702 453563487041 M3001 68702 453563486971 Japanese M3001 60702 453563487041 M3001 68702 4...

Страница 277: ...01 68802 453563486981 Portuguese M3001 60802 453563487051 M3001 68802 453563486981 Greek M3001 60802 453563487051 M3001 68802 453563486981 Turkish M3001 60802 453563487051 M3001 68802 453563486981 Russian M3001 60802 453563487051 M3001 68802 453563486981 Hungarian M3001 60802 453563487051 M3001 68802 453563486981 Czech M3001 60802 453563487051 M3001 68802 453563486981 Polish M3001 60802 4535634870...

Страница 278: ...3563331881 Exchange Measurement Server Norwegian Rel A M3000 6007A 453563485521 M3000 6807A 453563331901 Exchange Measurement Server Swedish Rel A M3000 6008A 453563485531 M3000 6808A 453563331921 Exchange Measurement Server Finnish Rel A M3000 6009A 453563485541 M3000 6809A 453563331941 Exchange Measurement Server Japanese Rel A N A N A M3000 6810A 453563331961 Exchange Measurement Server Danish ...

Страница 279: ...3563331891 Exchange Measurement Server ECG NBP SpO2 Press Temp Norwegian Rel D 0 M3000 6007B 453563485641 M3000 6807B 453563331911 Exchange Measurement Server ECG NBP SpO2 Press Temp Swedish Rel D 0 M3000 6008B 453563485651 M3000 6808B 453563331931 Exchange Measurement Server ECG NBP SpO2 Press Temp Finnish Rel D 0 M3000 6009B 453563485661 M3000 6809B 453563331951 Exchange Measurement Server ECG N...

Страница 280: ...0 M3000 6031B 453563331771 M3000 6831B 453563332091 Exchange Measurement Server ECG NBP SpO2 French Rel D 0 M3000 6032B 453563485741 M3000 6832B 453563332101 Exchange Measurement Server ECG NBP SpO2 German Rel D 0 M3000 6033B 453563485751 M3000 6833B 453563332111 Exchange Measurement Server ECG NBP SpO2 Dutch Rel D 0 M3000 6034B 453563485761 M3000 6834B 453563332121 Exchange Measurement Server ECG...

Страница 281: ... ECG NBP SpO2 Simplified Chinese Rel D 0 M3000 6043B 453563485841 M3000 6843B 453563332211 Exchange Measurement Server ECG NBP SpO2 Portuguese Rel D 0 M3000 6044B 453563485851 M3000 6844B 453563332221 Exchange Measurement Server ECG NBP SpO2 Greek Rel D 0 M3000 6045B 453563485861 M3000 6845B 453563332231 Exchange Measurement Server ECG NBP SpO2 Russian Rel D 0 M3000 6047B 453563377811 M3000 6847B ...

Страница 282: ...Temperature Swedish M3015 44139 453563332361 M3015A Front Bezel with Pressure and Temperature Finnish M3015 44140 453563332371 M3015A Front Bezel with Pressure and Temperature Japanese M3015 44144 453563332381 M3015A Front Bezel with Pressure and Temperature Portuguese M3015 44145 453563332391 M3015A Front Bezel with Pressure and Temperature Greek M3015 44147 453563332401 M3015A Front Bezel with P...

Страница 283: ...453563402681 M3015A Front Bezel without Pressure and Temperature Portuguese M3015 44175 453563402691 M3015A Front Bezel without Pressure and Temperature Greek M3015 44177 453563402701 M3015A Front Bezel without Pressure and Temperature Russian M3015 44180 453563402711 M3015A Front Bezel without Pressure and Temperature Polish M3015 29303 453563332261 M3015A Pump Kit including CO2 scrubber Can only...

Страница 284: ...nd Chinese order also the local language bezel as shown under M3015A Parts on page 266 Both hardware revisions A and B use the same bezels M3015 6001A 453563332421 M3015 6801A 453563332431 M3015A unit old hardware without Pressure and Temperaturea M3015 6031A 453563477861 M3015 6831A 453563477871 Hardware Revision B with Serial Number Prefix DE435xxxxx M3015A unit new hardware with Pressure and Te...

Страница 285: ...M3015A Measurement Server Extension Parts Replacement Parts 269 M3015A Showing Front Bezel with Press Temp and Mounting Pin 20 21 ...

Страница 286: ...Spanish M3016 44136 453563332481 M3016A Front Bezel Italian M3016 44137 453563332491 M3016A Front Bezel Norwegian M3016 44138 453563332501 M3016A Front Bezel Swedish M3016 44139 453563332511 M3016A Front Bezel Finnish M3016 44140 453563332521 M3016A Front Bezel Japanese M3016 44144 453563332531 M3016A Front Bezel Portuguese M3016 44145 453563332541 M3016A Front Bezel Greek M3016 44147 453563332551...

Страница 287: ...ront Bezel Dutch Product Option Number ExchangePart Number 12 Digit Reordering Number Description M3016A A01 M3016 6801A 453563332581 Exchange M3016A Measurement Server Extension Pressure and Temperature with Mainstream CO2 Englisha a For all languages apart from French Danish And Chinese order also the local language bezel as shown under M3016A Parts on page 270 M3016A A02 M3016 6831A 45356348309...

Страница 288: ...telliVue Information Center CD ROM M3080 10801 N A VPS Recovery CD Product Option Number Part Number 12 Digit Reordering Number Description Cables with both ends terminated with straight MSL connectors M3081A A11 M3081 61602 453563377851 2m Measurement Server to Monitor M3081A A12 M3081 61603 453563402731 4m Measurement Server to Monitor M3081 61626 453563474781 0 75m Measurement Server to Monitor...

Страница 289: ...ly with US country code for FCC countries Argentina Brazil Canada Chile China Columbia Dominican Republic Hong Kong Malaysia New Zealand Panama Taiwan Thailand US M3046 69564 453563333491 Exchange Wireless Assembly for Singapore country code France Mexico Singapore M3046 69566 453563333501 Exchange Wireless Assembly for Spain M3046 69567 453563333511 Exchange Wireless Assembly for ETSI countries w...

Страница 290: ...ge Part Number 12 Digit Reordering Number Description Item M3046 68302 453563333321 Exchange Battery 6 Part Number 12 Digit Reordering Number Description M8043 60000 453563498901 Smart Battery Charger and Conditioner New M8043 60010 451261001281 Replacement Kit 2 Air Fans 2 Filter Mats M8043 60011 451261001291 Replacement Kit 2 Filter Mats M8043 68000 453563498911 Smart Battery Charger and Conditi...

Страница 291: ...AL INOP description 180 CO2 EQUIP MALF INOP description 179 CO2 Gas Measurement Calibration Check 152 CO2 NO TRANSDUCER INOP description 179 CO2 NO TUBING INOP description 181 CO2 OCCLUSION INOP description 181 CO2 OVERRANGE INOP description 181 CO2 PURGING INOP description 182 CO2 Scrubber replacement 125 CO2 SENSOR WARM UP INOP description 180 CO2 UPDATE FW INOP description 180 CO2 WAIT CAL 2 IN...

Страница 292: ...on 32 numeric 10 numeric label 10 nurse call connector 6 92 Nurse Call Relay connecting 76 modification 76 Performance Test 157 O On Off Standby button 5 P Parts listing 247 parts and accessories iv Passwords 142 patient leakage current 64 name 10 safety 64 Patient safety checks 159 performance iv Performance Assurance Test 143 Power On Test 142 Power Supply related problems 188 PRESS zero button ...

Страница 293: ...eration 37 temperature extremes operating environment 66 Tests checklist 141 reporting in service record 138 time 10 Time counters resetting 153 touchstrip 5 TouchStrip problems 189 TouchStrip related problems 189 transducer connector 7 trickle charging 132 Troubleshooting Checklists 166 U unpacking 70 UNSUPPORTED LAN INOP description 179 V vibration 66 Visual Test 142 W wave 10 label 10 Z Zero bu...

Отзывы: