Telemetry Functions at the Information Center
9-2
Telemetry Functions at the Information Center & TeleMon
Telemetry Functions at the Information Center
1
Telemetry
Controls in
the Patient
Window
The Patient Window at the Information Center (accessed from the Patient
Window control in the Patient Sector) includes controls for a number of
telemetry operations. For detailed instructions on these operations, see the
IntelliVue Information Center Instructions for Use
or the
Online Help
.
To View ECG or SpO
2
Alarm Limits
Move the cursor over the HR or SpO
2
label to display the current high and low
alarm limits.
To Change ECG or SpO
2
Alarm Limits
Move the cursor over the High or Low numeric to display up/down arrow
controls for adjusting the limit. After adjusting the limit, move the cursor away
from the area to dismiss the limit controls.
To Change ECG Waveform Size
Move the cursor over the ECG waveform to display the ECG Waveform Size
control. Select the desired size from the list.
To Select Lead
Move the cursor over the ECG waveform to display the Lead Selection control.
Select the desired lead from the list.
Important—
Do not set the primary and secondary channels to the same lead.
To Change Va and Vb Default Lead Settings (6-lead only)
Move the cursor over the ECG waveform to display the Lead Selection popup.
Select the label from the label list. For Va or Vb, select Va or Vb, then select the
lead to be assigned. Assignment of the same V lead to both Va and Vb is not
allowed.
Important—
Do not set the primary and secondary channels to the same lead.
1. For information on ST functions performed at the Information Center, see “ST/AR ST
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...