Draft - 1 Aug 08
Understanding SpO
2
Alarms
8-20
SpO
2
Monitoring
SpO
2
Enable/
Disable at
Information
Center
You can enable/disable SpO
2
monitoring at the Information Center. See
“Patient Configurable Settings in Telemetry Setup” on page 9-6.
When SpO
2
is enabled, the Patient Sector and Patient Window of the
Information Center display a “T” next to the SpO
2
numeric (for example,
“SpO
2
T 90%”) to indicate that the measurement was made via telemetry.
SpO
2
Auto
ON at
Information
Center
The SpO
2
parameter is automatically enabled at the IntelliVue Information
Center if a manual SpO
2
measurement is initiated at the transceiver while in
Spot Check mode or if the SpO
2
sensor is inserted into the transceiver while the
transceiver is in Continuous SpO
2
mode.
When a patient is discharged and the transceiver is in Continuous mode, the
SpO
2
parameter is disabled. To reactivate the SpO
2
parameter Auto ON feature
from the transceiver, remember to do one of the following when a patient is
discharged:
–
remove the SpO
2
cable from the transceiver, wait 15 seconds, then
reinsert the cable
or
–
if using TeleMon, reset the transceiver to Manual mode.
Note—
The SpO
2
parameter Auto ON feature only needs to be reactivated when
the transceiver is in Continuous mode at discharge.
Note—
SpO
2
can always be enabled/disabled at the IntelliVue Information
Center.
Understanding SpO
2
Alarms
Physiologic SpO
2
alarms are generated and displayed at the Information Center.
SpO
2
offers high and low limit alarms, and a high priority (red level) oxygen
desaturation alarm. The SpO
2
low limit can be set between 50 and 99%
inclusive, in 1% increments. The desaturation limit is set automatically at 10
Содержание IntelliVue TRx
Страница 6: ...Draft 1 Aug 08 About this Book vi ...
Страница 40: ...Draft 1 Aug 08 Accessories 2 18 Product Safety ...
Страница 54: ...Draft 1 Aug 08 Audible Tones 3 14 Transceiver Controls ...
Страница 74: ...Draft 1 Aug 08 Battery Information 4 20 Basic Operation ...
Страница 94: ...Draft 1 Aug 08 Technical Alarms INOPs 5 20 Alarms ...
Страница 152: ...Draft 1 Aug 08 How the QT Analysis Algorithm Works 7 24 ST AR Arrhythmia Monitoring ...
Страница 176: ...Draft 1 Aug 08 Optimizing SpO2 Measurement Performance 8 24 SpO2 Monitoring ...
Страница 236: ...Draft 1 Aug 08 Troubleshooting 11 16 Maintenance Cleaning Troubleshooting ...
Страница 268: ...Draft 1 Aug 08 SpO2 Accessories A 10 Accessories ...
Страница 270: ...Draft 1 Aug 08 B 2 Sales and Support Offices ...
Страница 274: ...Draft 1 Aug 08 Index 4 ...