Positioning ECG Electrodes
6-14
ECG Monitoring
5-Wire
Placement
(Standard
Mode)
Lead
Placement
RA
directly below the clavicle and near the
right shoulder
LA
directly below the clavicle and near the
left shoulder
LL
on the left lower abdomen
RL
on the right lower abdomen
V
on the chest, the position depends on your
required lead selection. The typical
position is V1, although this may vary
according based on your hospital’s
protocol.
V2
V3
V4
V5
V6
V1
RA
LA
RL
LL