Configuring the System
8-4
Start-up and Configuration
•
Move the cursor across the display to see that it tracks normally.
•
Click on a
Patient Window
button to verify control and normal application and
display(s) response. For single display configurations, the
Main Screen
will occupy
the top half of the screen and the
Patient Window
, the bottom half. For dual displays,
the
Patient Window
will appear on the second display.
Step 3.
Verify
normal system operation
by checking the following:
•
Patient data appear for all configured Patient Sectors
•
Patient numerics are displayed normally
•
Waveforms are updated smoothly
•
Arrhythmia softkeys appear on bedside monitors
•
For telemetry monitors, telemetry setup can be changed successfully
•
no error messages are displayed
Step 4.
Test the
Philips Recorder(s)
by requesting a recording from a Patient Sector, a
bedside monitor, or use the
Recording/Printing Diagnostics
application in the
Diagnostics
menu of the
Service
mode. Verify that annotated data and waveforms are
recorded correctly.
Step 5.
Verify
alarm
annunciation
, sound and recording, using an alarm limit adjustment or
other means that does not interfere with patient care. Verify that the alarm sound and
alarm recording activate correctly.
Step 6.
Test the
printer
by printing a text-based report or a graphical report with waveforms.
Or use the
Recording/Printing Diagnostics
application in the
Diagnostics
menu of
the
Service
mode. Check that the data print correctly.
Step 7.
Test the
keyboard
by accessing the
Purchased Options and Support Information
application of the
Service
mode. Type characters from all keys, (upper and lower case)
in the
Contact
field and verify that the correct entries appear.
Step 8.
Select each Information Center using the buttons on the
KVM Switch
front panel and
verify that the keyboard and mouse correctly control that Information Center’s
Main
Screen
applications.
If the KVM Switch is being used to control a 2nd display:
– Verify the following for each Information Center connected to the Switch that has
dual display capability
– the proper applications are activated on the 2nd display when selected on the
Information Center’s
Main Screen
– the mouse cursor moves freely between the
Main Screen
and the 2nd Display for
the selected Information Center
– the keyboard and mouse correctly control the applications in the 2nd Display
– the connected 2nd display is not a touch display
Database
Storage Test
After at least 10 minutes of monitoring,
Patient Data Review applications
shall be reviewed
to verify that stored data are correctly displayed for all patient monitors -- SDN, M3/M4
(wired and wireless) and the IntelliVue Patient Monitors.
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...