Troubleshooting
7-76
Maintenance, Troubleshooting, and Repair
If transfer time is determined to be too long, adjust the Bandwidth Utilization setting to try
to shorten the time period. This change takes 3-5 seconds and can be done while monitoring.
Decreasing the time period of the data transfer will increase the bandwidth utilization of the
hospital LAN, which may be undesirable.
Note
Patient Data Transfer durations are estimates. Actual time on a hospital network will vary
depending on the design of the network and the level of network bandwidth that is available.
Change Units
Password
Changes the Unit settings password. This is available on Database servers and M3150 Local
Database Information Centers only.
Configuration
Opens the Printer dialog box to delete a print jobs in the queue or a stuck print job while the
devices are in monitoring mode. This is not to be used to add or delete printers! Adding and
deleting printers must be done via the
page of the Config Wizard.
Table 7-10. Patient Data Transfer - Bandwidth Utilization
Hospital Infrastructure
Topology
Peak Bandwidth
Utilization (Average)
Patient Data Transfer
Duration
Bandwidth Utilization
Setting
Pure 100 Mbit Hospital LAN
72 Mbps
32 Mbps (avg)
1-2 minutes
High
56 Mbps
30 Mbps (avg)
3-5 minutes
Medium
32 Mbps
8 Mbps (avg)
10-15 minutes
Low
Mixed 10/100 Mbit Hospital LAN
8 Mbps
8 Mbps (avg)
6-8 minutes
High
8 Mbps
6 Mbps (avg)
10-12 minutes
Medium
8 Mbps
4 Mbps (avg)
15-20 minutes
Low
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...