Configuration
6-44
Software Installation & Configuration
•
Type in the desired
label
prefix
(e.g.
Bed, Tel, Mon
) in the
Prefix:
box using the
keyboard.
•
Type in the
first number
(typically 1) of the
Label
sequence in the
Start with Index:
box.
•
Type in the desired
number of labels
with the
Prefix
entered above in the
Number of
labels:
field. Typically this is the number of
Beds, Telemetry Equipment
, or M3/
M4/IntelliVue
Monitors
in the clinical
Unit
.
•
For
Bed Labels
, click on the clinical
Unit
where these beds reside from the
Associated with Unit
field. This field is grayed out for
Telemetry Equipment
and
Monitor
labels.
•
Click
OK
to enter these
Labels
in the
Labels
list.
Step 5.
Select an icon --
SCC, Switch
or
Monitor Labels
-- in the left column of the
Equipment Setup for ...
field.
If an
SCC
icon is selected, follow the
Setup LBN Procedure
that follows:
If a
Monitor Labels
icon is selected, skip to the
Setup Monitor Procedure
.
If a
Switch
icon is selected, skip to the
Setup Switch Procedure
.
Setup LBN
Procedure
Step 1. Assign
SDN
monitoring
Equipment
(hardwired and telemetry) on SCC
Logical
Branch Numbers
(
LBN
s) to
Bed Labels
as follows:
•
Click on (highlight) the first
SCC
in the left column of the
Equipment Setup for
Selected SCC
field.
•
Click on (highlight) the first
LBN #
in the
LBN
column to activate it for assigning a
patient monitor.
•
Click
Setup
to bring up the
Map Monitor to LBN #X
window
for that
LBN #
.
Select the type of patient monitor attached to that
LBN #
--
Hardwired
or
Telemetry
-- by
clicking in the circle preceding it.
–
If
Telemetry
was selected, select a
Label
for that telemetry monitor (e.g.
Tel1
) in
the
Equipment Label
box.
If
Hardwired
was selected, proceed to the following step.
–
If this patient monitor is to be assigned a
Bed Label
, click on (highlight) the
Bed
Label
to be assigned to that patient monitor from the
Located at bed
list.
If this patient monitor is not to have a bed label, click a check
in the
Setup
equipment only
box.
–
Click
OK
to enter that monitor and
Bed Label
assignment in the
Equipment Setup
for Selected SCC.
Step 2. Repeat Step 6 for each
LBN #
and
SCC
in the list.
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...