Configuration
6-28
Software Installation & Configuration
Buttons
at the bottom of the window do the following:
OK
stores the information entered. It then appears in the
List of network devices
in
the
Network Configuration
window.
Cancel
closes the
Add to network domain
window with no Device being added.
The remaining buttons to the right of the
List of network centrals and hallways
have the
same use as for Clinical Units.
List of network devices
lists the network devices connected to this system. This field is
accessed by
Network Statistics
to provide status information on network switches and
access points.
Columns
in the
List of network devices
field are:
Note
IP Addresses for devices
must be entered
in this window in order to identify their
performance in the
Network Statistics
application.
If no switches are entered, then the
Equipment Setup
page will not allow any of the Mx/
IntelliVue Patient Monitor installation options.
Adding Network
Devices
In addition to adding these network devices to the Database Server/Information Center
system, the network devices must also be configured to communicate with the system. The
following table provides what their additional configuration requirements are:
Field
Description
Device Name
Names given to a
12-Lead ECG Management,
24 Port
-
Switch, Application Server, Holter,
Networked Printer, SNTP Time Source, Alert Data Export Client, Wireless Access Point
(with an identifying icon) on the Network of this Server.
Device Names
for devices are specified
in the
Add new network device to this domain window
when the device is added to the
network
Host Name
Host name of device being added
Device Type
Type of device:
12-Lead ECG Management,
24 Port
-
Switch, Alert Data Export Client,
Application Server, Holter, Networked Printer, SNTP Time Source, Wireless Access Point
IP Address
Internet Protocol Address of the device that identifies its location on the Network.
Printer MAC Address
(Media Access Control Address) is the 12 digit HEX number that uniquely identifies the printer.
This number must be obtained from the printer’s configuration printout.
Device Type
Systems Supported on
Requirements
12-Lead ECG Management
Database Server
ECG Management system must be configured to
receive Information Center data (PMDExport folder
must be created). The Database Server must be
configured with the 12-Lead ECG Management
system’s hostname
24 Port Switch
all
Must be configured using the Information Center
ConfigTool
Содержание IntelliVue Series
Страница 152: ...IntelliVue Information Center Application Software 3 22 Software Description ...
Страница 442: ...Troubleshooting 7 156 Maintenance Troubleshooting and Repair ...
Страница 488: ...Procedure B 12 Web Installation on the Database Server ...
Страница 492: ...Procedure C 4 External Modem Installation ...
Страница 500: ...Using RAS D 8 Dial In Procedure for Remote Access to Information Center Systems ...