Operational Problems
3
-
30
Troubleshooting
Figure 3-3 Power Module Indicators
Power Indicator
Charge Done
Indicator
Charging Indicator
Table 3-16 Power Module Indicator Matrix
INDICATORS
SYSTEM STATE
MEANING
POWER
CHARG-
ING
DONE
Con-
nected
to Mains
(AC/DC) ?
Con-
nected
M3500B/
M5500B ?
Battery in
Place?
Unit Power
ON ?
R = RED G = GREEN
OR = ORANGE/AMBER
Y = YES N = NO --- = don’t care
OFF
OFF
OFF
N
---
---
---
Normal
G
OFF
OFF
Y
---
---
---
Normal
G
OFF
OFF
Y
Y
N
---
Normal
G
OR
OFF
Y
Y
Y
---
Battery is charging.
G
G
OFF
Y
Y
Y
---
Battery 90% charged.
G
G
G
Y
Y
Y
---
Battery fully charged.
R
OFF
OFF
Y
Y
---
---
If Power light was green before
connecting to unit, fault in
M3500B/M5500B.
R
OFF
OFF
Y
N
---
---
Mains voltage low or
Fault in Power Module.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...