M3500B/M5500B Service Manual
2
-
13
Performance Verification
2
Functional Checks
The following functional checks exercise the basic functions of the defibrilla-
tor/monitor. They are intended as a broad check of the unit’s performance, and
are designed to complement (not replace) the Diagnostic Tests described later.
If all elements of a test pass, record that test as a PASS and return to the main
diagnostic menu by pressing
. If there is any failure, begin trouble-
shooting and repairing the unit as needed. See "Troubleshooting"
on page 3-1.
ECG Functional Checks
This section describes how to check the operation of the ECG functions. Each
of the ECG checks assumes the unit and the simulator are still set up as they
were at the end of the previous ECG check.
To check ECG display and Heart Rate (HR) functions:
1. Connect the ECG simulator to both the Pads input and the 3- or
5-lead ECG cable. Set the simulator for normal sinus rhythm (NSR), 1mV
amplitude, at some nominal rate (e.g., 60 bpm).
2. Set the M3500B/M5500B to Manual operating mode (not Diagnostic
Mode).
3. Using the
softkey, verify that the display shows a normal
ECG with a clean baseline for both Pads and Lead II.
4. Verify the Heart Rate (HR) displayed is correct.
5. Disconnect the ECG simulator from the pads cable and verify that the dis-
play shows a dashed line in place of the waveform and that the unit both
alarms and gives the
Pads Off
message.
6. If using a 5-lead ECG cable, set the defibrillator to monitor from the V
lead.
7. Disconnect each of the ECG leads from the simulator one at a time, and
verify that the display shows a dashed line in place of the waveform and
that the unit both alarms and gives the
Leads Off
message.
To check ECG printing functions:
1. Reconnect the simulator.
2. Print a strip and verify that the strip shows a normal ECG with a clean
baseline.
3. Verify that the date, time, and configuration information printed at the top
of the strip is correct.
MAIN
LEAD SELECT
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...