M3500B/M5500B Service Manual
2
-
5
Test and Inspection Matrix
2
Defibrillator Test -
AC or DC Power
Module (DM)
(Required only
when AC or DC
power module
used.)
Using only the AC or DC
power module, enter Diag-
nostic Mode and run the
Defibrillator Test (at
200Joules) (page 2-27):
If the measurements are as
follows:
Example: DM:p
Displayed by
M3500B/M5500B:
•
Available Energy after
Shock
•
0
•
Msec to charge
•
< 15000 msec
•
Delivered energy
•
Actual delivered energy
200 +7%
•
Impedance
•
42 to 57
Ω
•
Defib errors
•
None (0)
then the Defibrillator test
(Power Module) passes
Defibrillator Test -
Battery Power
(DB)
Using only battery power,
enter Diagnostic Mode and
run the Defibrillator Test (at
200Joules) (page 2-28).
If the measurements are as
follows:
Example: DB:p
Displayed by
M3500B/M5500B
•
Available Energy after
Shock
•
0
•
Msec to charge
•
< 3000 msec
•
Delivered energy
•
Actual delivered energy
200 +7%
•
Impedance
•
42 to 57
Ω
•
Defib errors
•
None (0)
then the Defibrillator test
(Battery Power) passes
Defibrillator
Disarm Test (D)
Enter Diagnostic Mode and
run the Defibrillator Disarm
Test (page 2-29)
If all readings as expected
Available energy - failure >0
No errors reported
then the Defibrillator Disarm
test passes
Example: D:p
Table 2-1 Performance Verification and Safety Tests (Continued)
Test Group Name
Test or Inspection to
Perform
Expected Test Results
Data to Record
P (pass) or F (fail)
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...