Top Case Assemblies
4
-
50
Removal and Replacement
Interface PCA
The Interface PCA is located in the top case, under the Control Stack.
NOTE
When replacing the Interface PCA, the manual keypad MUST also be
replaced. The screws which attach the Interface PCA to the case are UNDER
the manual key overlay.
CAUTION
See cautions regarding handling and connection of flex circuits on page 4-1.
Removal
1. Disconnect and remove the Control Stack. See "Control Stack" on page
2. Remove the plastic insulating shield over the SpO
2
and Interface PCAs.
Remove the tape that secures the shield to the metal clips around the SpO
2
and ECG connectors and pull the shield straight up.
3. Disconnect the Interface PCA as follows:
a.
Referring to Figure 4-29, disconnect the cable from the ECG connec-
tor to the Interface PCA at the Interface PCA. Press the latch on top
of the connector and pull straight out. Swing the cable up out of way.
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...