Top Case Assemblies
4
-
38
Removal and Replacement
SpO
2
PCA
The SpO
2
PCA is located next to the ECG and SpO
2
connectors.
Removal
1. Disconnect and remove the ECG connector. See "ECG Connector" on
page 4-40.
2. Disconnect and remove the SpO
2
connector. See "SpO
page 4-42.
3. Note the orientation and placement of the flex circuit from the SpO
2
PCA
to the Interface PCA. Disconnect this flex circuit at both ends. These con-
nectors have a sliding bar latch - slide out toward the edge of the PCA to
release, then pull out the flex.
4. Remove the SpO
2
PCA and discard the flex circuit to the Interface PCA.
Figure 4-22 Removing the Screws - SpO
2
PCA
ECG Connector
SpO
2
Connector
Interface PCA
Interface PCA Connectors:
SpO
2
ECG
Flex Circuit - SpO
2
PCA to Interface
PCA
Ferrite
SpO
2
PCA
Содержание Heartstart XLT
Страница 1: ...H E A R T S TA R T X LT S e r v i c e M a n u a l M 3 5 0 0 B M 5 5 0 0 B ...
Страница 2: ......
Страница 3: ...Service Manual M3500B HeartStart XLT M5500B Heartstart 4000 Defibrillator Monitor ...
Страница 6: ......
Страница 16: ......
Страница 20: ......
Страница 52: ......
Страница 110: ...Opening the Sealed Case 4 24 Removal and Replacement Figure 4 13 Battery Connector Location 4 Pin Battery Connector ...
Страница 195: ...M3500B M5500B Service Manual 6 17 System Functional Descriptions 6 Figure 6 3 Power Distribution and Fusing ...
Страница 205: ...M3500B M5500B Service Manual 7 9 Waveforms 150J continued 7 Waveforms 150J continued 150J 75 ohms 150J 100 ohms ...
Страница 206: ...Waveforms 150J continued 7 10 Specifications Waveforms 150J continued 150J 125 ohms 2ms div 150J 125 ohms 5 ms div ...
Страница 208: ...Waveforms 200J continued 7 12 Specifications Waveforms 200J continued 200J 75ohms 200J 100ohms ...
Страница 216: ......
Страница 223: ......
Страница 224: ...M3500 90900 Printed in USA February 2003 Second Edition M3500 90900 2 ...