When to Perform Tests Blocks
Chapter 8 - Test and Inspection Matrix
81
Table 21
Test and Inspection Matrix
Test Block
Name
Test
Expected Results
What to
Record
Visual Test
(See 20)
Inspect exterior of fetal
monitor for damage.
No Visual Damage
V:P or V:F
Power-On Self-
Test
(See 20)
Power on the monitor
Displays Normal Monitoring
Screen Configuration and
emits tones
PO:P or PO:F
Safety Tests:
•
Safety Test 1
(See 86)
Enclosure Leakage
Current - S.F.C. Open
Supply
All Safety Test results are in
range of Table 22 on page 87
S:P or S:F
•
Safety Test 2
(See 88)
Patient Leakage Cur-
rent - S.F.C. Open
Supply
All Safety Test results are in
range of Table 23 on page 89
S:P or S:F
Key to Table
:
P = Pass
F = Fail
Содержание FM-2
Страница 4: ......
Страница 10: ......
Страница 12: ......
Страница 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Страница 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Страница 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...
Страница 84: ...Assembly Drawings 72 Chapter 7 Spare Parts Figure 8 Assembly Drawing Sheet 2 Front Case Assembly ...
Страница 85: ...Assembly Drawings Chapter 7 Spare Parts 73 Figure 9 Assembly Drawing Sheet 3 Rear Case Assembly ...
Страница 86: ...Assembly Drawings 74 Chapter 7 Spare Parts Figure 10 Assembly Drawing Sheet 4 Interface PCB Rear Connector PCB ...
Страница 87: ...Assembly Drawings Chapter 7 Spare Parts 75 Figure 11 Assembly Drawing Sheet 5 Side Connector PCB Panel Label ...
Страница 88: ...Assembly Drawings 76 Chapter 7 Spare Parts ...