Service Menu
26
Chapter 4 - Power-up Default Settings and the Service Menu
Service Menu
The purpose of the Service Menu is to allow factory, field-service, and hospital
biomedical technicians access to a series of test and system-related information
screens for the purpose of verifying monitor performance or troubleshooting
problems.
To access the Service Menu, press the navigation knob when the Power Status
Frame is selected. The following can be viewed from the Service Menu:
Data transfer
Disabled
Yes (if real-time transfer)
Data Transfer Destination
Recorder/
Printer
Yes
Recorder: Paper speed
3 cm/minute
Yes
Recorder: Paper style
US
Yes
Time format
12 hour
Yes
Date
“today”
Yes
Date format
mm/dd/yy
Yes
Patient ID
time/date encoded
No
Language
Configured language
Yes
Table 4
Power-Up Default Settings
Parameter
Factory Default
Saved Through Power Cycle
Содержание FM-2
Страница 4: ......
Страница 10: ......
Страница 12: ......
Страница 24: ...Routine Maintenance and Cleaning 12 Chapter 1 General Information ...
Страница 42: ...Service Menu 30 Chapter 4 Power up Default Settings and the Service Menu ...
Страница 74: ...Main PCB Disassembly Procedures 62 Chapter 6 Disassembly Reassembly Guide ...
Страница 84: ...Assembly Drawings 72 Chapter 7 Spare Parts Figure 8 Assembly Drawing Sheet 2 Front Case Assembly ...
Страница 85: ...Assembly Drawings Chapter 7 Spare Parts 73 Figure 9 Assembly Drawing Sheet 3 Rear Case Assembly ...
Страница 86: ...Assembly Drawings 74 Chapter 7 Spare Parts Figure 10 Assembly Drawing Sheet 4 Interface PCB Rear Connector PCB ...
Страница 87: ...Assembly Drawings Chapter 7 Spare Parts 75 Figure 11 Assembly Drawing Sheet 5 Side Connector PCB Panel Label ...
Страница 88: ...Assembly Drawings 76 Chapter 7 Spare Parts ...